K970395 · Drial Consultants, Inc. · DKZ · Jun 26, 1997 · Clinical Toxicology
Device Facts
Record ID
K970395
Device Name
QUIKSTRIP ONE STEP METHAMPHETAMINE
Applicant
Drial Consultants, Inc.
Product Code
DKZ · Clinical Toxicology
Decision Date
Jun 26, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3100
Device Class
Class 2
Intended Use
Syntron's QuikStrip One Step Methamphetamine assay is a rapid, qualitative, competitive binding immunoassay for the determination of methamphetamine in urine. The test provides only preliminary data which should be confirmed by other methods such as gas chromatography/mass spectrophotometry (GC/MS). Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated. Syntron's QuikStrip One Step Methamphetamine Test is not intended to monitor drug levels, but only to screen urines for the presence of methamphetamine and its metabolites.
Device Story
Chromatographic absorbent immunoassay; detects methamphetamine in urine. Principle: competitive binding; drug/metabolites in sample compete with immobilized drug conjugate for limited antibody sites. Sample flows through membrane; labeled antibody-dye conjugate binds free drug forming antibody-antigen complex. Complex competes with immobilized antigen in positive reaction zone; absence of magenta band indicates drug concentration >500 ng/ml. Control zone binds unbound dye conjugate to produce magenta band, confirming device function. Used in clinical/forensic settings; operated by healthcare professionals. Output is visual qualitative result (presence/absence of color band). Results are preliminary; require GC/MS confirmation for clinical decision-making.
Clinical Evidence
Clinical trial of 305 urine samples. Compared against GC/MS confirmation. Results: relative sensitivity 100%, relative specificity 97.7%, overall accuracy 98.68%. In-house testing against Syva EMIT® II showed 100% sensitivity, 100% specificity, and 100% accuracy.
Technological Characteristics
Chromatographic absorbent device; porous membrane support; antibody-dye conjugate; competitive binding immunoassay. Qualitative visual readout. No external energy source. Standalone test strip format.
Indications for Use
Indicated for the qualitative screening of human urine for the presence of methamphetamine and its metabolites. Intended for professional use to provide preliminary analytical test results. Requires confirmation by GC/MS.
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Reference Devices
Syva EMIT® II
Related Devices
K970265 — ONE STEP URINE DRUG OF ABUSE METHAMPHETAMINE TEST · Drial Consultants, Inc. · Apr 8, 1997
K970724 — QUICKPAC II ONE STEP METHAMPHETAMINE TEST · Drial Consultants, Inc. · May 13, 1997
K973269 — QUIKSTRIP ONE STEP BENZODIAZEPINE TEST · Drial Consultants, Inc. · Feb 25, 1998
K971109 — QUIKPAC II ONE STEP AMPHETAMINE TEST · Drial Consultants, Inc. · May 14, 1997
Submission Summary (Full Text)
{0}
510k Submission for
QuikStrip One Step Methamphetamine Test K970395
Syntron Bioresearch, Inc.
Revision C 5/06/97 Printed on 5/15/97
JUN 26 1997
Page 24 of 57
## 8. Summary of Safety and Effectiveness
The sponsor, Syntron Bioresearch, Inc. (2774 Loker Ave. West, Carlsbad, California, 92008), has developed, manufactured and tested under GMP/GLP guidelines a device for the qualitative testing of urine for the presence of methamphetamine and its metabolites in a screening format.
The trade name of the device is QuikStrip One Step Methamphetamine Test having a designated common name of Methamphetamine Test System and a classification as a class II device per 21 CFR ¶ 862.3610. This device is intended for the medical/forensic screening of urine.
Syntron's QuikStrip One Step Methamphetamine Test consists of a chromatographic absorbent device in which the drug or drug metabolites in the sample compete with a drug conjugate immobilized on a porous membrane support for the limited antibody sites. As the test sample flows up through the absorbent device, the labeled antibody-dye conjugate binds to the free drug in the specimen forming an antibody:antigen complex. This complex competes with immobilized antigen conjugate in the positive reaction zone and will not produce a magenta color band when the drug is above the detection level of 500 ng/ml. Unbound dye conjugate binds to the reagent in the control zone, producing a magenta color band, demonstrating that the reagents and device are functioning correctly.
In-house testing of Syntron's QuikStrip One Step Methamphetamine Test yielded a relative sensitivity or agreement within positives and relative specificity or agreement within negatives of 1.00 and an accuracy of 100% when tested against Syva EMIT® II on samples documented to be positive by GC/MS. A clinical trial consisting of 305 samples was run and the combined data yielded a relative sensitivity or agreement within positives of 100%, a relative specificity or agreement within negatives of 97.7% with an accuracy of 98.68%.
All positive samples by either screening method were confirmed by GC/MS. Four of the 6 samples were negative for methamphetamine, but positive for Amphetamine by GC/MS. The other two were negative for methamphetamine but positive for Pseudoephedrine and Phenyl Propanolamine (PPA) respectively.
Additional information on this submission may be obtained by contacting Dr. Cleve W. Laird, President, Drial Consultants, Inc. at 805-522-6223(Ca) or by fax at 805-522-1526.
Proprietary Information
Confidential
Do Not Copy
Proprietary Information
Confidential
{1}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
Dr. Cleve W. Laird
President and CEO
Drial Consultants, Inc.
1420 Los Angeles Avenue, Suite 201
Simi Valley, California 93065
Re: K970395
QuikStrip One Step Methamphetamine Test
Regulatory Class: II
Product Code: DKZ, LAF
Dated: May 13, 1997
Received: May 16, 1997
Dear Dr. Laird:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
JUN 26 1997
{2}
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical Laboratory Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{3}
510(k) Number (if Known): Not yet Assigned
Device Name: QuikStrip One Step Methamphetamine Test
Indications For Use:
## INTENDED USE
Syntron's QuikStrip One Step Methamphetamine assay is a rapid, qualitative, competitive binding immunoassay for the determination of methamphetamine in urine. The test provides only preliminary data which should be confirmed by other methods such as gas chromatography/mass spectrophotometry (GC/MS). Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated⁶. Syntron's QuikStrip One Step Methamphetamine Test is not intended to monitor drug levels, but only to screen urines for the presence of methamphetamine and its metabolites.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANDOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

Perscription Use: ☑ (Per 21 CFR 801.109 or Over The Counter Use: ☐ (Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.