K970357 · Epitope, Inc. · PJD · Mar 10, 1997 · Clinical Chemistry
Device Facts
Record ID
K970357
Device Name
EPISCREEN
Applicant
Epitope, Inc.
Product Code
PJD · Clinical Chemistry
Decision Date
Mar 10, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1675
Device Class
Class 2
Indications for Use
EpiScreen is intended to collect oral fluid specimens, contain those specimens, and preserve the specimens after collection and during transport from the collection area to the laboratory.
Device Story
EpiScreen device consists of treated absorbent cotton fiber pad on nylon stick and preservative solution in plastic vial. Pad impregnated with salts and gelatin creates hypertonic environment; induces osmotic gradient across buccal/gingival mucosa; enhances flow of mucosal transudate onto cotton fibers. Pad placed between lower gum and cheek for collection period; removed and placed into vial containing preservative to inhibit oral microorganism growth. Vial sealed and transported to laboratory for processing. Used by clinicians/technicians to facilitate oral fluid specimen collection for diagnostic testing.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Components: treated absorbent cotton fiber pad, nylon stick, plastic specimen vial, preservative solution. Principle: osmotic gradient induction via hypertonic pad impregnation. Form factor: manual collection applicator and transport container. No electronic or software components.
Indications for Use
Indicated for the collection, preservation, and transport of oral fluid specimens for laboratory analysis. No specific patient population, age, or gender restrictions are defined.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
Predicate Devices
Saliva Sampler (Saliva Diagnostics Systems)
Salivette (Sarstedt, Inc.)
Saliva Sac (BioQuant, Inc.)
Submission Summary (Full Text)
{0}
Epitope, Inc. 8505 SW Creekside Place
Beaverton, OR 97008 503-641-6115
EpiScreen® Oral Specimen Collection Device
January 29, 1997
# 510(k) Summary
**Trade Name:** EpiScreen™
**Common Name:** Oral Specimen Collection Device
**MAR 10 1997** 14970357
## Device Description
The EpiScreen device consists of a treated absorbent cotton fiber pad affixed to a nylon stick (Collection Pad) and a preservative solution in a plastic container (Specimen Vial). The Collection Pad is impregnated with a mixture of common salts and gelatin, creating a hypertonic environment which produces an osmotic gradient across the buccal and gingival mucosa. The pad is placed in contact with the gingival mucosa (between the lower gum and cheek) which enhances the flow of mucosal transudate onto the absorptive cotton fibers of the pad. Following the collection period, the Collection Pad is removed from the mouth and placed into a Specimen Vial. The vial contains a preservative solution which serves to inhibit the growth of oral microorganisms recovered on the Collection Pad. The vial is sealed with a plastic cap and transported to a laboratory for processing and testing.
## Intended Use
EpiScreen is intended to collect oral fluid specimens, contain those specimens, and preserve the specimens after collection and during transport from the collection area to the laboratory.
## Classification
EpiScreen is classified as, or is substantially equivalent to, one or more of the following:
- Specimen Container (Class I, 21 CFR §880.6175, General Hospital and Personal Use Devices, General Hospital Panel, Product Code 80FMH);
- Specimen Transport and Storage Container (Class I, 21 CFR §880.3250, Hematology and Pathology Devices, Pathology Panel, Product Code 88KDT);
- Absorbent Tipped Applicator (Class I, 21 CFR §880.6025, General Hospital and Personal Use Devices, General Hospital Panel, Product Code 80FOR);
- Microbiological Specimen Collection and Transport Device (Class I, 21 CFR §866.2900), Immunology and Microbiology Devices, Microbiology Panel, Product Code 83LIO).
Like EpiScreen, these devices are used for taking specimens, and for containing those specimens, and preserving (with or without a fixative solution) the viability and integrity of the specimens during transport from the collecting area to the laboratory.
EpiScreen 510(k) Summary
Epitope, Inc.
{1}
Epitope, Inc. 8505 SW Creekside Place
Beaverton, OR 97008 503-641-6115
EpiScreen® Oral Specimen Collection Device
January 29, 1997
# Comparison of Technological Characteristics
EpiScreen is substantially equivalent to the following devices:
- Saliva Sampler manufactured by Saliva Diagnostics Systems of Vancouver, WA.
- Salivette manufactured by Sarstedt, Inc. of Newton, NC.
- Saliva Sac manufactured by BioQuant, Inc. of Ann Arbor, MI.
Like EpiScreen, the Saliva Sampler, Salivette, and Saliva Sac are intended for collecting an oral fluid specimen, and for containing and transporting that specimen. The following table lists the indications and components for EpiScreen and the three equivalent devices. Note that all four devices share the same indication as well as major components (collection apparatus and transport container). While EpiScreen has different technological characteristics than the predicate devices discussed here, EpiScreen does not raise different questions of safety and efficacy.
## Comparison Table: EpiScreen, Saliva Sampler, Saliva Sac, and Salivette
| | EpiScreen | Saliva Sampler | Saliva Sac | Salivette |
| --- | --- | --- | --- | --- |
| Indication | Collection/transport of oral fluid specimens | Collection/transport of oral fluid specimens | Collection/transport of oral fluid specimens | Collection/transport of oral fluid specimens |
| Collection Apparatus | ☑ | ☑ | ☑ | ☑ |
| Transport Container | ☑ | ☑ | ☑ | ☑ |
| Fixative Solution | ☑ | ☑ | N/A | N/A |
| Suspension Insert | N/A | N/A | N/A | ☑ |
| Volume Indicator | N/A | ☑ | N/A | N/A |
EpiScreen 510(k) Summary
Epitope, Inc.
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
10903 New Hampshire Avenue
Document Control Center – WO66-G609
Silver Spring, MD 20993-0002
Epitope, Inc.
Caroline R. Sayre, RAC
Regulatory Affairs Supervisor
8505 SW CREEKSIDE PLACE
BEAVERTON, OREGON 97008
FEB 06 2015
Re: K970357
Trade/Device Name: EpiScreen™ Oral Specimen Collection Device
Regulation Number: 21 CFR 862.1675
Regulation Name: Blood specimen collection device
Regulatory Class: II
Product Code: PJD
Dated: January 29, 1997
Received: January 30, 1997
Dear Ms. Sayre:
This letter corrects our previous Substantially Equivalent (SE) letter of March 10, 1997.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}
Page 2 – Ms. Sayre
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH’s Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,

Courtney H. Lias, Ph.D.
Director
Division of Chemistry and Toxicology Devices
Office of In Vitro Diagnostics and Radiological Health
Center for Devices and Radiological Health
Enclosure
{4}
510(k) Number (if known): ___________________________
Device Name: EpiScreen™ Oral Specimen Collection Device
Indications for Use:
EpiScreen is intended for use in the collection, preservation, and transport of oral fluid specimens.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Davon Hansen
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K920357
Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.