CAVITRON JET WITH SPS SYSTEM

K970342 · Dentsply Intl. · ELC · Apr 4, 1997 · Dental

Device Facts

Record IDK970342
Device NameCAVITRON JET WITH SPS SYSTEM
ApplicantDentsply Intl.
Product CodeELC · Dental
Decision DateApr 4, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4850
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The CAVITRON™ JET WITH SPS SYSTEM is designed for use in general prophylaxis and periodontal treatments and other areas of operative dentistry. This CAVITRON™ JET WITH SPS SYSTEM is used to debride light to heavy calculus and plaque deposits from tooth and root surfaces.

Device Story

Portable, stand-alone prophylaxis unit for dental offices; operated by dental professionals. Device features two modes: ultrasonic scaling and air polishing. Scaling mode uses ultrasonic inserts to produce 30,000 strokes/second with water irrigation to remove calculus. JET mode uses air/powder/water spray with sodium bicarbonate for stain/plaque removal. System includes base unit, foot control, handpiece, and inserts. Automatic mode selection occurs based on inserted tip. Electronic system includes automatic frequency control (nominally 30 kHz) and automatic stroke control (gain control) to maintain performance under load. Foot pedal controls activation and boost power mode. Benefits include improved patient comfort, effective calculus removal, and restoration of tooth enamel esthetics.

Clinical Evidence

Bench testing only. In-house testing confirmed performance within specifications. SEM analysis showed no significant qualitative difference in tooth surface effects compared to the predicate device. EMC testing conducted by Underwriters' Laboratory with positive results.

Technological Characteristics

Ultrasonic scaler and air polisher. Electromechanical transducer converts magnetic field to mechanical motion. Nominal frequency 30 kHz. Features automatic frequency control and automatic gain control. Base unit material resistant to wipe disinfectants. Connectivity: standard electrical outlet. Components: electronic generator, foot control, handpiece, inserts. Enhancements: boost power mode, improved pinch tube, accessible air filters.

Indications for Use

Indicated for general prophylaxis, periodontal treatments, and operative dentistry to debride light to heavy calculus and plaque from tooth and root surfaces.

Regulatory Classification

Identification

An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.

Predicate Devices

Submission Summary (Full Text)

{1} 510(k) SUMMARY Page 1 of 2 DENSITY X970342 NAME & ADDRESS: DENTSPLY International 570 West College Avenue P.O. Box 872 York, PA 17405-0872 (717) 845-7511 Fax (717) 854-2343 P. J. Lehn Telefax (717) 849-4343 APR - 4 1997 CONTACT: P. Jeffery Lehn DATE PREPARED: January 28, 1997 TRADE NAME: CAVITRON® JET WITH SPS SYSTEM CLASSIFICATION NAME: Ultrasonic Scaler 21 CFR, 872.4850 PREDICATE DEVICE: CAVITRON® JET SCALER K864996 DESCRIPTION: The CAVITRON® JET WITH SPS SYSTEM is a portable stand alone multifunctional prophylaxis unit. It has two operating modes: an ultrasonic scaler mode and an air polishing (JET) mode. The ultrasonic scaler mode operates using a Cavitron® ultrasonic scaling insert with a fine, warm water spray, easily and effectively removing stubborn calculus and stains both supragingivally and subgingivally. The JET mode operates using a Cavitron® JET Air Polishing Insert with an air/powder/water spray, easily removing gross stain and plaque from all exposed tooth surfaces. The CAVITRON® JET WITH SPS SYSTEM has six components: electronic generator (contained in a base unit, foot control, handpiece and cable assembly, detachable line cord, DENTSPLY® Cavitron® Ultrasonic Sterilizable Inserts, and Cavitron® JET Air Polishing Inserts. The CAVITRON® JET WITH SPS SYSTEM plugs into an ordinary electrical outlet and converts the 60 cycle AC house current into either high frequency electromechanical vibrations in the scaling mode, or a spray mixture of air, water and powder in the JET mode. Mode selection occurs automatically when either a Cavitron® ultrasonic sterilizable insert or a Cavitron® JET air polishing insert is placed into the handpiece. The device consists of three parts: an electromechanical transducer, an electronic driving system, and an air/powder module. The transducer converts electrical energy in the form of a magnetic field into mechanical motion. The electronic system, in addition to providing the magnetic field, incorporates two control loops, one providing automatic frequency control driving each transducer at it's optimum operating frequency (nominally 30 kHz), while the second loop provides automatic stroke control to compensate for different transducer loading conditions. The air/powder module consists of the automatic selection and control electronics, the pneumatic controls and a powder reservoir. In the scaling mode, the CAVITRON® JET WITH SPS SYSTEM produces 30,000 microscopically small strokes each second at the transducer's tip. This, combined with acoustic effects of the coolant water, produces a synergistic action that sweeps away the heaviest calculus deposits while providing exceptional operator and patient comfort. In the JET mode, the CAVITRON® JET WITH SPS SYSTEM produces a spray mixture of air, water and specially processed sodium bicarbonate required for cleaning, polishing and restoring the natural esthetics of tooth enamel. 000031 {2} 510(k) SUMMARY (cont'd.) Page 2 of 2 # CAVITRON™ JET WITH SPS SYSTEM DESCRIPTION (cont'd.): While the controls for the power and water flow of the CAVITRON™ JET WITH SPS SYSTEM are located on the base unit, the on/off control used during operation is located on the foot pedal. When the foot pedal is pressed to the first position, the ultrasonic tip is activated and irrigation starts when in the scaling mode; only water flows when in the JET mode. The water from the irrigation system acts as both a lavage and a coolant. When the foot pedal is pressed to the second position and the unit is in the scaling mode, power to the handpiece is increased slightly for the length of time the second position is engaged. Irrigation continues to flow. When the unit is in the JET mode, a spray mixture of water, air and powder is generated. INTENDED USE: The CAVITRON™ JET WITH SPS SYSTEM is designed for use in general prophylaxis and periodontal treatments and other areas of operative dentistry. This CAVITRON™ JET WITH SPS SYSTEM is used to debride light to heavy calculus and plaque deposits from tooth and root surfaces. TECHNOLOGICAL CHARACTERISTICS: The CAVITRON™ JET WITH SPS SYSTEM is substantially equivalent to the predicate device, DENTSPLY's Cavitron® JET Scaler (K864996), in intended use, as well as in design and construction. The basic technology, primary energy source, maximum electrical input power, maximum input air and water pressure, maximum mechanical output energy, and base unit material are identical to the predicate device. The CAVITRON™ JET WITH SPS SYSTEM has the following enhancements: 1) a low power capability providing lower tip strokes than previously obtainable for periodontal applications and greater patient comfort; 2) a foot switch activated boost power mode which, when enabled, applies increased power to the handpiece; 3) a limited Automatic Gain Control circuit which automatically increases insert drive power, maintaining tip stroke under load; 4) a smooth or appropriately textured, white base unit capable of withstanding common wipe disinfectants without discoloring or deteriorating; 5) easier access to air filters for maintenance; and 6) an enhanced pinch tube design. The CAVITRON™ JET WITH SPS SYSTEM meets DENTSPLY Specifications. Underwriters' Laboratory conducted EMC testing on the SYSTEM with positive results. Results of in-house testing indicated that the CAVITRON™ JET WITH SPS SYSTEM performed within stated specifications and was found to be as effective as the predicate device. Results of SEM Analysis indicated that no significant qualitative difference was observed on the teeth following the use of the CAVITRON™ JET WITH SPS SYSTEM compared to the predicate device. We believe that the performance data and the testing reported above demonstrate the safety and effectiveness of the CAVITRON™ JET WITH SPS SYSTEM for its intended use. 000032
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...