ACCESSORIES TO THE STEREOTACTIC BODY FRAM

K970291 · Elekta Instrument AB · IYE · Apr 24, 1997 · Radiology

Device Facts

Record IDK970291
Device NameACCESSORIES TO THE STEREOTACTIC BODY FRAM
ApplicantElekta Instrument AB
Product CodeIYE · Radiology
Decision DateApr 24, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2

Indications for Use

The Accessories to the Stereotactic Body Frame is a device designed for stereotactic diagnostic localization and stereotactic radiotherapy of extra cranial targets.

Device Story

Accessories to Stereotactic Body Frame; includes laser chest pointer mounted on arc-ruler for patient alignment; includes pillow holder (shell) for reproducible patient positioning; pillow options include vacuum bag with polystyrene spheres or polyurethane foam-filled bag; includes quality check instrument for coordinate verification. Used in clinical radiotherapy settings by medical staff. Device facilitates accurate patient fixation and refixation for target localization during diagnostic exams and radiotherapy treatment. Enhances positioning reproducibility compared to base frame.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Mechanical accessories for stereotactic body frame. Includes laser chest pointer, arc-ruler, longitudinal ruler, and pillow holder (shell). Pillow materials: vacuum bag with polystyrene spheres or polyurethane foam-filled bag. Quality check instrument included for coordinate verification. No software or electronic processing.

Indications for Use

Indicated for patients requiring stereotactic diagnostic localization and stereotactic radiotherapy of extra cranial targets.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0} APR 24 1997 K970291 # SECTION 17: SUMMARY OF SAFETY AND EFFECTIVENESS This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92. ## 17.1 SUBMITTER INFORMATION a. Company Name: Elekta Instrument AB b. Company Address: St. Larsgatan 2, S-582 24 Linkoping, Sweden c. Company Phone: (011) 4613248300 (011) 4613112312 d. Contact Person: Sverker Glans Vice President Quality and Regulatory Affairs Elekta Instrument AB e. Date Summary Prepared: January 17, 1997 ## 17.2 DEVICE IDENTIFICATION a. Trade/Proprietary Name: Accessories to the Stereotactic Body Frame b. Classification Name: Medical charged-particle radiation therapy system ## 17.3 IDENTIFICATION OF PREDICATE DEVICE | Company | Device | 510(k) No. | Date Cleared | | --- | --- | --- | --- | | Elekta Instrument AB | Stereotactic Body Frame | K960338 | April 23, 1996 | ## 17.4 DEVICE DESCRIPTION The Stereotactic Body Frame utilizes a reference system of coordinates that can be reproducibly localized during diagnostic examination and treatment. The Accessories to 11 {1} the Stereotactic Body Frame include a laser chest pointer which is mounted on an arc-ruler and is used for alignment of the patient. The coordinates used for patient positioning can be easily read on the arc-ruler holding the chest pointer and on the longitudinal ruler, along which the arc-ruler can be moved. An additional accessory to the Stereotactic Body Frame is the pillow holder, or shell. This framework is affixed to the body frame and allows for reproducible positioning of the patient within the frame. The pillow holder also allows for removal and insertion of the pillow. The pillow is available as either a vacuum bag filled with polystyrene spheres, or as a polyurethane foam-filled bag. The Stereotactic Body Frame comes equipped with a quality check instrument for verifying the accuracy of the coordinate set-up. ## 17.5 SUBSTANTIAL EQUIVALENCE The Elekta Accessories to the Stereotactic Body Frame is substantially equivalent to other stereotactic frames currently in commercial distribution by Elekta Instruments AB in terms of the intended use of achieving a safe and accurate fixation and refixation of a patient’s body for localization of targets for radiotherapy treatment. The fundamental technical characteristics are similar to those of the predicate device and are listed on the comparison chart provided in this 510(k) submission. Differences that exist between these systems is the addition of the pillow holder which aids in the accuracy of the positioning of the patient within the frame and the conversion of the chest marker to a laser pointer. The addition of these accessories does not affect the relative safety or effectiveness of the device. ## 17.6 INTENDED USE The Accessories to the Stereotactic Body Frame is a device designed for stereotactic diagnostic localization and stereotactic radiotherapy of extra cranial targets. {2} 117 ## 17.7 TECHNOLOGICAL CHARACTERISTICS A comparison of the technological characteristics of the Accessories to the Stereotactic Body Frame and the predicate device is provided in this 510(k) submission. ## 17.8 510(K) CHECKLIST This notification contains all information required by 21 CFR 807.87. A completed copy of the Premarket Notification 510(k) Reviewer’s Checklist is provided in this submission.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...