K970277 · Astoria-Pacific, Inc. · KQP · Dec 11, 1997 · Clinical Chemistry
Device Facts
Record ID
K970277
Device Name
URIDYLTRANSFERASE KIT, 50 HOUR (80-4000-13K)
Applicant
Astoria-Pacific, Inc.
Product Code
KQP · Clinical Chemistry
Decision Date
Dec 11, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1315
Device Class
Class 2
Attributes
Pediatric
Indications for Use
This Astoria-Pacific SPOTCHECK® Uridyltransferase 50-Hour Reagent Kit is for the qualitative determination of galactose-1-phosphate uridylyltransferase, EC 2.7.7.12 (GALT) activity in whole blood saturated filter paper disks (S&S 903 filter paper or equivalent), using the API™ 300 SPOTCHECK® Analyzer or the RFA-300 System. Measurements of galactose-1-phosphate uridyltransferase are used in the diagnosis and treatment of the hereditary disease galactosemia. This method is intended for in vitro diagnostic use as an aid in screening for decreased levels of GALT activity in infants. This method is not for monitoring purposes.
Device Story
The Uridyltransferase 50-Hour Reagent Kit provides reagents for the quantitative determination of GALT enzyme activity in whole blood samples collected on filter paper. The device is used with the API 300 SPOTCHECK Analyzer or RFA-300 System. The principle of operation involves a biochemical reaction where GALT reacts with galactose and UDP Glucose, followed by NADP reduction at 37°C. The analyzer measures the resulting fluorescent compound at 450 nm excitation and 550 nm emission. The system is designed for mass screening in clinical laboratory settings. The output is a fluorescence measurement used by healthcare providers to identify infants with decreased GALT activity, aiding in the diagnosis of galactosemia. The kit includes pre-weighed or pre-made reagents to streamline preparation and includes a NADH standard for curve calibration.
Clinical Evidence
Clinical testing performed on 27 whole blood saturated filter paper disk specimens, including 20 normal neonatal samples and 7 deficient samples from juveniles and adults. Results showed 100% correlation with no false positives or false negatives reported.
Technological Characteristics
Reagent kit for fluorometric enzyme assay. Components include succinate and Tris buffers, NADP, Dithiothreitol, Magnesium, EDTA, Triton X-100, Glu-1-Phosphate, Gal-1-Phosphate, UDPG, G6PDH (yeast), PGluM (rabbit muscle), and NADH standard. Operates via automated fluorescence detection at 450 nm excitation/550 nm emission. Designed for use with API 300 SPOTCHECK or RFA-300 systems.
Indications for Use
Indicated for the qualitative screening of decreased galactose-1-phosphate uridylyltransferase (GALT) enzyme activity in infants to aid in the diagnosis and treatment of galactosemia, using whole blood saturated filter paper disks.
Regulatory Classification
Identification
A galactose-1-phosphate uridyl transferase test system is a device intended to measure the activity of the enzyme galactose-1-phosphate uridyl transferase in erythrocytes (red blood cells). Measurements of galactose-1-phosphate uridyl transferase are used in the diagnosis and treatment of the hereditary disease galactosemia (disorder of galactose metabolism) in infants.
{0}
K970277
# 510(k) SUMMARY
Astoria-Pacific
Uridyltransferase 50-Hour Reagent Kit
# 510(K) SUMMARY
DEC 11 1997
# IDENTIFICATION
Submitter's Name: Raymond L. Pavitt, President Astoria-Pacific, Inc.
Address: 14600 SE 82nd Drive
Clackamas, OR 97015-0830, USA
Telephone: 503-657-3010
FAX: 503-655-7367
Contact Person: Lester B. Garrison, Diagnostics Manager
Date Summary Prepared: August 4, 1997
# NAME OF DEVICE
Proprietary Name: Uridyltransferase (GALT) 50-Hour Reagent Kit
Common Name: Uridyltransferase (GALT) Test System
Classification Name: 862.1315 Galactose-1-Phosphate Uridyltransferase Test System
# LEGALLY MARKETED DEVICE EQUIVALENT TO PROPOSED DEVICE
The Uridyltransferase 50-Hour Reagent Kit is substantially equivalent to a legally marketed predicate device, Sigma Galactose-1-Phosphate Uridyltransferase (Gal-PUT) Deficiency Screen.
# DESCRIPTION OF PROPOSED DEVICE
The proposed device, Uridyltransferase 50-Hour Reagent Kit, is a set of reagents to be used with the API™ 300 SPOTCHECK® Analyzer or the RFA-300 System for the quantitative determination of the enzyme galactose-1-phosphate uridyltransferase (UT) in whole blood saturated filter paper disks. The amount of uridyltransferase activity is determined by measuring the fluorescent compound produced in the reaction of UT with galactose and UDP Glucose, followed by NADP reduction at 37°C. The excitation wavelength of the reaction product is 450 nm, and its emission is measured at 550 nm. The method is specific for uridyltransferase.
The method is designed for mass screening, with enough reagents in each 50-Hour Reagent Kit for 1 week plus start-up (50 hours total) of run time. It is packaged to reduce storage space and to require a minimum of time to prepare. Each component is packaged with the correct weight to prepare the required volume of reagent. The standard is in a concentrated form, to permit easy dilution to prepare a standard curve. The materials provided and their components are listed below.
{1}
510(k) SUMMARY
Astoria-Pacific
Uridyltransferase 50-Hour Reagent Kit
REAGENTS
| DESCRIPTION | Component | Concentration | Quantity |
| --- | --- | --- | --- |
| GALT Extraction Buffer | Succinate Buffer pH 5.2 | 0.8 mM | 1 Liter |
| Recipient Diluent | Tris Buffer, pH 7.8 | 100 mM | 1 Liter |
| | Triton X-100 | 0.5 mL/L | |
| GALT Buffer Part A. | Tris Buffer, pH 7.8 | 100 mM | 5 x 150 mL |
| | NADP | 1.3 mM | |
| | Dithiothreitol | 1.0 mM | |
| GALT Buffer Part B. | Magnesium | 0.6 mM | |
| | EDTA | 0.1 mM | |
| | Triton X-100 | 0.5 mL/L | |
| PGluM Substrate Reagent | Glu-1-Phosphate | 10. mM | 10 mL |
| GALT Substrate Reagent | Gal-1-Phosphate | 4.3 mM | 5 x 60 mL |
| | UDPG | 0.5 mM | |
| | G6PDH Yeast | 1.6 KU/L | |
| | PGluM Rabbit Muscle | 0.8 KU/L | |
| NADH Stock Standard | TEA Buffer, pH 9.0 | 50 mM | 2 x 50 mL |
| | NADH | 50 μM | |
## INTENDED USE
This Astoria-Pacific SPOTCHECK® Uridyltransferase 50-Hour Reagent Kit is for the qualitative determination of galactose-1-phosphate uridylyltransferase, EC 2.7.7.12 (GALT) activity in whole blood saturated filter paper disks (S&S 903 filter paper or equivalent), using the API™ 300 SPOTCHECK® Analyzer or the RFA-300 System. Measurements of galactose-1-phosphate uridylyltransferase are used in the diagnosis and treatment of the hereditary disease galactosemia. This method is intended for in vitro diagnostic use as an aid in screening for decreased levels of GALT activity in infants. This method is not for monitoring purposes.
## SUMMARY OF TECHNOLOGICAL CHARACTERISTICS OF DEVICE COMPARED TO PREDICATE DEVICE
## Similarities with Predicate Device
The Astoria-Pacific SPOTCHECK® Uridyltransferase 50-Hour Reagent Kit is similar to a predicate device, Sigma Galactose-1-Phosphate Uridyltransferase (Gal-PUT) Deficiency Screen, having the same technological characteristics and intended use. Both are diagnostic reagent kits to be used for the qualitative, fluorometric determination of the enzyme uridylyltransferase in blood; both have the same chemical principles and reaction mechanism.
{2}
510(k) SUMMARY
Astoria-Pacific
Uridyltransferase 50-Hour Reagent Kit
The reactants and reactions are the same. Both devices use the same incubation temperature and a similar incubation time.
## Differences from Predicate Device
While the method and reagents are similar for the proposed and predicate devices, there are minor differences. The Sigma device uses G6PD and PGluM from the sample for the reactions. The proposed device adds these enzymes to the reagent. The Sigma device requires manual reading of fluorescence; the API™ 300 SPOTCHECK® Analyzer reads the fluorescence automatically.
There are also differences in the wavelengths of the excitation and emission filters. The proposed device designates 450 nm and 550 nm for the excitation and emission wavelengths, respectively; the Sigma device uses a broad band UV lamp for excitation and visual fluorometry for reading the emitted fluorescence. Because there is a broad bandpass of about 40 nm for the filters on the API™ 300 SPOTCHECK® Analyzer and the emitted fluorescent light is broad spectrum, this difference is not consequential.
The API Uridyltransferase 50-Hour Reagent Kit contains a NADH standard for comparision of the fluorescence readings.
The Sigma Gal-PUT Deficiency Screen test requires the user to prepare all reagents; in the Astoria-Pacific SPOTCHECK® Uridyltransferase 50-Hour Reagent Kit, the reagents are either pre-made or pre-weighed.
The minor differences between the Astoria-Pacific SPOTCHECK® Uridyltransferase 50-Hour Reagent Kit and the Sigma Gal-PUT Deficiency Screen raise no questions of safety or efficacy.
## BRIEF DISCUSSION OF NONCLINICAL TESTS
Because there is such a strong similarity between the Astoria-Pacific SPOTCHECK® Uridyltransferase 50-Hour Reagent Kit and the Sigma Gal-PUT Deficiency Screen, no nonclinical testing has been performed to demonstrate substantial equivalence.
## BRIEF DISCUSSION OF CLINICAL TESTS
Clinical testing was performed on whole blood saturated filter paper disks. These tests include 27 specimens analyzed, of which 20 were normal neonatal blood samples and 7 were deficient samples from juveniles and adults. Results correlated well, as there were no false positives and no false negatives among any of the samples tested.
## CONCLUSIONS DRAWN FROM THE NONCLINICAL AND CLINICAL TESTS
The clinical tests demonstrate that the proposed device, the Astoria-Pacific SPOTCHECK® Uridyltransferase 50-Hour Reagent Kit, is safe, effective and performs as well as a predicate device, as evidenced by the correlation of results obtained from clinical studies.
## OTHER INFORMATION
No other information has been requested by FDA at this time.
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
DEC 11 1997
Lester B. Garrison
Submission Correspondent
Astoria-Pacific, Inc.
14600 S.E. 82nd Drive
P.O. Box 830
Clackamas, Oregon 97015
Re: K970277
Uridyltransferase Test System
Regulatory Class: II
Product Code: KQP
Dated: November 5, 1997
Received: November 6, 1997
Dear Mr. Garrison:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical Laboratory Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}
Page ___ of ___
510(k) Number (if known): K970277
Device Name: Undyltransferase 50-Hour Reagent Kit
Indications For Use:
October 8, 1997
Astoria-Pacific, Inc.
Statement of Indications for use:
This Astoria-Pacific SPOTCHECK® Uridyltransferase 50-Hour Reagent Kit is for the qualitative determination of galactose-1-phosphate uridyltransferase (EC 2.7.7.12) (GALT) activity in whole blood saturated filter paper disks, S&S 903 filter paper or equivalent, using the API™ 300 SPOTCHECK® Analyzer or the AFA-300 System. Measurements of galactose-1-phosphate uridyltransferase are used in the diagnosis and treatment of the hereditary disease galactosemia. This method is intended for in vitro diagnostic use as an aid in screening for decreased levels of GALT enzyme activity in infants. This method is not for monitoring purposes.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number
OR
Over-The-Counter Use
(Optional Format 1-2-96)
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number
Prescription Use ☑
(Per 21 CFR 801.109)
TOTAL P.08
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.