DIFFUSER PANEL

K970240 · Focused Viewer, Inc. · IXC · Mar 11, 1997 · Radiology

Device Facts

Record IDK970240
Device NameDIFFUSER PANEL
ApplicantFocused Viewer, Inc.
Product CodeIXC · Radiology
Decision DateMar 11, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1890
Device ClassClass 1

Indications for Use

The Diffuser Panel is designed to diffuse the light from the light source in Radiographic Film Illuminators.

Device Story

Diffuser Panel functions as a component for radiographic film illuminators; consists of light diffusing layers, brightness enhancer films, and a scratch-resistant polycarbonate cover. Device installed in illuminator housing; operates by diffusing light from internal source to provide uniform illumination across x-ray film. Used in clinical settings by radiologists or clinicians to facilitate diagnostic review of radiographic images. Provides consistent light distribution to aid visual interpretation of films.

Clinical Evidence

No clinical data; bench testing only. Equivalence based on functional comparison of light diffusion and distribution characteristics.

Technological Characteristics

Components: light diffusing films, brightness enhancer films, scratch-resistant polycarbonate panel. Function: light diffusion and uniform distribution. Form factor: panel insert for radiographic film illuminators.

Indications for Use

Indicated for use as a component in radiographic film illuminators to diffuse and evenly distribute light across x-ray film for clinical viewing.

Regulatory Classification

Identification

A radiographic film illuminator is a device containing a visible light source covered with a translucent front that is intended to be used to view medical radiographs.

Predicate Devices

Submission Summary (Full Text)

{0} K970240 MAR 11 1997 Focused Viewer, Inc. Radiographic Film Illuminator 510(k) Notification # 510(k) SUMMARY 1. Name of submitter, address, phone, fax, contact person and date of summary preparation. Focused Viewer, Inc. 2420 Pasadena Blvd. Wauwatosa, Wisconsin 53226 The contact person is: William Brauer, M.D. Owner Phone: (800)843-9269 Fax: (414)334-8293 Summary was prepared on January 14, 1997. 2. Name of device Trade/Proprietary Name: Diffuser Panel Common/Usual Names: Translucent front of a Radiographic Film Illuminator Classification Name: Radiographic Film Illuminator (21 CFR 892.1890) 3. Legally Marketed Equivalent Device Name We are claiming substantial equivalence to the panel marketed by Diversified Diagnostic Products. Refer to K875233. Original 510(k). January, 1997. 10 {1} Focused Viewer, Inc. Radiographic Film Illuminator 510(k) Notification # 510(k) SUMMARY CONTINUED ## 4. Description of the device The Diffuser Panel consists of layers of light diffusing and brightness enhancer films, plus a clear polycarbonate panel that is coated with a scratch guard to protect the surface of the brightness enhancer from scratching. ## 5. Intended use of the device The Diffuser Panel is designed to diffuse the light from the light source in Radiographic Film Illuminators. ## 6. Comparison of technological characteristics with predicate device The function of glass or plastic translucent panels is to diffuse light, and distribute bright light evenly over the x-ray film. The Focused Viewer Diffuser Panel is composed of plastic and is designed to diffuse light and distribute light evenly over the x-ray film in a manner equivalent to marketed devices and to provide at least equivalent brightness. William Brauer, M.D. Owner Focused Viewer, Inc. JAN 21, 1997 Date Original 510(k). January, 1997.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...