K970233 · Gary A. Gelbfish · MCW · Oct 8, 1997 · Cardiovascular
Device Facts
Record ID
K970233
Device Name
GELBFISH ENDO-VAC SYSTEM
Applicant
Gary A. Gelbfish
Product Code
MCW · Cardiovascular
Decision Date
Oct 8, 1997
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 870.4875
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Gelbfish Endo-Vac System is intended for the percutaneous aspiration of thrombus from acutely clotted AV grafts as a part of a percutaneous thrombectomy procedure. The device incorporates an anti-clogging/self-cleaning mechanism to maintain its uninterrupted function. The device is not (by itself) intended to restore function to clotted AV grafts. It is only used in conjunction with standard percutaneous thrombectomy procedures to facilitate the aspiration of clot from the graft. The arterial and venous anastomotic plug/clot and pathology must still be addressed by standard Fogarty and/or balloon catheter procedures. Expertise in traditional pharmoco-mechanical techniques is necessary, and may be used in conjunction with the Endo-Vac System. The enclosed device is indicated for use to remove thrombus in clotted synthetic ePTFE grafts only. Its short length makes it suitable for aspiration of clot from the “body” of the graft. It is NOT intended to cross arterial or venous anastomoses or to be used on clot contained in native artery or vein.
Device Story
Endovascular suction catheter system for removing thrombus from synthetic ePTFE AV grafts. Components: introducer sheath with side-port/stopcock, dilator, guidewire, and stainless steel anti-clogging/self-cleaning unit. Operation: sheath inserted into graft; suction applied via hand-held valve. Self-cleaning mechanism uses a reciprocating blunt tip and saline irrigation to clear clogs; tip extended for suction, withdrawn into sheath for saline flushing. Used by clinicians in interventional radiology/vascular surgery settings. Facilitates clot removal during standard thrombectomy; does not replace Fogarty/balloon catheter procedures for anastomotic pathology. Benefits: maintains uninterrupted suction function during procedure.
Clinical Evidence
Clinical study of 47/47 thrombectomy procedures demonstrated successful thrombus clearance from synthetic ePTFE grafts. Study confirmed device does not interfere with standard interventional radiology instruments or techniques.
Technological Characteristics
System includes introducer sheath, dilator, guidewire, and stainless steel self-cleaning unit. Self-cleaning mechanism: blunt tip, half-cylindrical tube, 22-gauge saline injection tube. Energy source: external wall suction. Materials: medical grade stainless steel. Mechanical operation: reciprocating tip movement for clog clearance. Non-electronic, manual operation.
Indications for Use
Indicated for percutaneous aspiration of thrombus from acutely clotted synthetic ePTFE AV grafts in patients undergoing percutaneous thrombectomy. Not for use in native arteries/veins or crossing anastomoses. Requires use with standard thrombectomy procedures.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
Predicate Devices
Endovascular suction catheter (A. S. Aloe Company, 1960s)
{0}
K970233
# 510(k) Summary
OCT - 8 1997
## GELBFISH ENDOVAC SYSTEM
**Common Name:** Endovascular Suction Catheter
**Sponsor:** Gary A. Gelbfish, M.D.
2502 Avenue I
New York, NY 10035
Phone: 800-563-1463
Fax: 718-421-0628
**Contact:** Gary A. Gelbfish, M.D.
**Prepared:** October 1, 1997
## A. LEGALLY MARKETED PREDICATE DEVICES
The Gelbfish EndoVac System is substantially equivalent to two pre-amendment devices, both endovascular suction catheters, marketed in the 1960s by the A. S. Aloe Company, St. Louis, MO, and by the Klista Company, Nashville, TN. These devices were used in general vascular surgery applications where it was necessary to remove thrombus. It is also similar to another pre-amendments suction thrombectomy device that is still marketed today by Medi-tech, the Greenfield Pulmonary Embolectomy Catheter which uses a suction mechanism for the extraction of clot from the pulmonary vasculature.
## B. DEVICE DESCRIPTION
The Gelbfish EndoVac System consists of a set of three main groups of components: (1) an introducer sheath with side-port for suction connection (with three-way stopcock), plus dilator and guidewire (used for initial sheath insertion), (2) a stainless steel anti-clogging/self-cleaning unit with connector for saline injection, dual check valve, tubing and syringe for saline irrigation, and (3) suction components--a fail-safe trumpet suction valve, and suction tubing. All but the stainless part are components that are presently marketed as medical devices or components. Only the stainless steel self-cleaning unit is new, and it is fabricated from medical grade stainless steel components used for blood contact applications.
{1}
C. INDICATIONS FOR USE
The Gelbfish EndoVac System is intended for the percutaneous aspiration of thrombus from acutely clotted AV grafts as a part of a percutaneous thrombectomy procedure. The device incorporates an anti-clogging/self-cleaning mechanism to maintain its uninterrupted function.
The device is not (by itself) intended to restore function to clotted AV grafts. It is only used in conjunction with standard percutaneous thrombectomy procedures to facilitate the aspiration of clot from the graft. The arterial and venous anastomotic plug/clot and pathology must still be addressed by standard Fogarty and/or balloon catheter procedures. Expertise in traditional pharmoco-mechanical techniques is necessary, and may be used in conjunction with the EndoVac System.
The enclosed device is indicated for use to remove thrombus in clotted synthetic ePTFE grafts only. Its short length makes it suitable for aspiration of clot from the body of the graft. It is NOT intended to cross arterial or venous anastomoses or to be used on clot contained in native artery or vein.
D. TECHNOLOGICAL CHARACTERISTICS
The percutaneous thrombectomy procedure in which the EndoVac device is used involves a series of steps, almost all of which are already part of both the radiological and surgical procedures. The only enhancement is the aspiration of clot from the graft via the introducer sheath in conjunction with the anti-clogging/self-cleaning mechanism.
The anti-clogging/self-cleaning mechanism consists of a blunt tip of the same diameter as the tip of the sheath, a short half-cylindrical tube of the same diameter as the tip, and a 22-gauge tube to inject saline just behind the tip and to provide the mechanical movement of the tip. When the tip is extended just beyond the end of the sheath, the half-cylindrical tube allows thrombus to be sucked into the sheath through its open half. When thrombus clogs the sheath during suction, the tip is withdrawn into the sheath, cutting off the interior of the sheath from the clotted graft. Saline is introduced inside the tip behind the clog and allows the sheath to be cleared, after which the tip is again extended into the open position and aspiration continues.
Clinical procedure: An introducer sheath is inserted in the venous limb of the graft pointing towards the venous anastomosis. The sideport is connected to wall suction controlled via a hand held valve. The sheath with self-cleaning mechanism is inserted into the graft, suction is
{2}
activated and the sheath is slowly pulled back, reciprocating the self-cleaning mechanism and with active irrigation as necessary.
## E. TESTING
Mechanical testing has demonstrated that the stainless steel cleaning mechanism, which is laser welded together, can withstand a pull test of 20 lbs.
A clinical study of the device showed that in 47/47 thrombectomy procedures, the EndoVac device successfully cleared thrombus from the graft. The study also demonstrated that the EndoVac device does not interfere in any way with the use of the other instruments and techniques which are required to complete the interventional radiology thrombectomy procedure.
## F. CONCLUSIONS
The Gelbfish EndoVac System has the same intended use, notably the aspiration of clot, as the predicate devices. Dr. Gelbfish has demonstrated through his clinical study on the device and his comparison of the device's characteristics with those of the predicate devices that it is substantially equivalent to the predicate devices.
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20856
Gary A. Gelbfish, M.D.
c/o Robert L. Sheridan
Vice President for Device Evaluation
C.L. McIntosh Associates Inc.
12300 Twinbrook Parkway, Suite 625
Rockville, Maryland 20852
OCT - 8 1997
Re: K970233
Gelbfish EndoVac System
Regulatory Class: II
Product Code: MCW
Dated: July 8, 1997
Received: July 9, 1997
Dear Dr. Gelbfish:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal
{4}
Page 2 - Gary A. Gelbfish, M.D.
Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the *Federal Register*. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
{5}
# STATEMENT OF INDICATIONS FOR USE
510(k) Number (if known): ___________________________
Device Name: **Gelbfish Endo-Vac System**
Indications For Use:
The **Gelbfish Endo-Vac System** is intended for the percutaneous aspiration of thrombus from acutely clotted AV grafts as a part of a percutaneous thrombectomy procedure. The device incorporates an anti-clogging/self-cleaning mechanism to maintain its uninterrupted function.
The device is not (by itself) intended to restore function to clotted AV grafts. It is only used in conjunction with standard percutaneous thrombectomy procedures to facilitate the aspiration of clot from the graft. The arterial and venous anastomotic plug/clot and pathology must still be addressed by standard Fogarty and/or balloon catheter procedures. Expertise in traditional pharmoco-mechanical techniques is necessary, and may be used in conjunction with the Endo-Vac System.
The enclosed device is indicated for use to remove thrombus in clotted synthetic ePTFE grafts only. Its short length makes it suitable for aspiration of clot from the “body” of the graft. It is NOT intended to cross arterial or venous anastomoses or to be used on clot contained in native artery or vein.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Judith Danielson for. J. Ryan 10/7/97
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
000048
q
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.