VISTA ILLUMINATION SYSTEM

K970217 · Vista Medical Technologies, Inc. · FFS · Apr 2, 1997 · Gastroenterology, Urology

Device Facts

Record IDK970217
Device NameVISTA ILLUMINATION SYSTEM
ApplicantVista Medical Technologies, Inc.
Product CodeFFS · Gastroenterology, Urology
Decision DateApr 2, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Vista Illumination System is intended to provide visible light for illumination of various rigid and flexible endoscopes, laparoscopes, cameras, and headlamps,etc. It is intended for use in all types of endoscopic and endoscopic assisted procedures, as well as visualization of the evacuated cardiac chamber.

Device Story

Vista Illumination System; Xenon light source unit; fiberoptic cable. Provides visible light for rigid/flexible endoscopes, laparoscopes, video cameras, headlamps. Features dual ports with independent attenuator controls; intensity level indicators; system/lamp fault LEDs; LCD lamp life display. Fiberoptic cable produces uniform beam profile; includes adapters for Storz, ACMI, Wolfe; core diameters 0.1mm-1mm; 10-foot length. Used in clinical settings for endoscopic procedures; operated by physicians/clinical staff. Benefits patient by providing necessary illumination for visualization during minimally invasive surgery.

Clinical Evidence

Bench testing only. Compliance with IEC 601-1 (General Safety Requirements for Medical Electrical Equipment), IEC 601-1-2 (Electrical Magnetic Compatibility), and UL 544 (Standards for Safety, Medical and Dental Equipment).

Technological Characteristics

Xenon light source; fiberoptic cable delivery. Dual ports with independent attenuators. LED indicators for system/lamp status; LCD for lamp life. Fiber core diameters 0.1mm-1mm. Adapters for Storz, ACMI, Wolfe. Standards: IEC 601-1, IEC 601-1-2, UL 544.

Indications for Use

Indicated for illumination of rigid and flexible endoscopes, laparoscopes, cameras, and headlamps during endoscopic and endoscopic-assisted procedures, including visualization of the evacuated cardiac chamber.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1} Vista Medical Technologies Westborough, MA K 970 217 510(k) Notification Illumination System January 1997 APR - 2 1997 510(k) SUMMARY January, 1997 ## COMPANY NAME AND ADDRESS Vista Medical Technologies 134 Flanders Road Westborough, MA 01581 ## CONTACT PERSON Martin Newman Director of Regulatory Affairs and Quality Assurance Telephone: (508) 366-3668 Fax: (508) 366-1543 ## DEVICE TRADE NAME Vista Illumination System ## COMMON NAME Xenon Light Source for Endoscopic Procedures ## PREDICATE DEVICE Device Name: Cogent Light Illuminator 10010 Xenon Light Source and Fiberoptic Cable Classification: Illuminator, Fiberoptic for Endoscope Manufacturer: Cogent Light Technologies, Inc. 650 S. San Vicente Blvd. Los Angeles, CA 90048-4620 510(k)#: K933405 510(k) 28 {2} Vista Medical Technologies Westborough, MA 510(k) Notification Illumination System January 1997 ## DEVICE DESCRIPTION The Vista Illumination System, is a Xenon Light Source illumination system consisting of a light source unit which houses the lamp and a fiberoptic cable. The system provides visible light for illumination of rigid or flexible endoscopes, laparoscopes, video cameras, headlamps,etc. The light source has 2 ports which are provided with independent attenuator controls and separate light output intensity level indicators. The front panel contains dual intensity level indicators as well as system fault and lamp fault LED's. The LED's will light if the lamp fails to start or fails during use for any reason. There is a LCD on the front panel that provides an indication of lamp life. The fiberoptic cable produces a uniform beam profile from a single fiber. There are three different adapters to allow connection to various endoscopic devices currently available. The thread types are for Storz, ACMI and Wolfe. Custom adapters are available. The fiber is available in core diameter of 0.1mm to 1mm with a standard cable length of 10 feet. ## INTENDED USE The Vista Illumination System is intended to provide visible light for illumination of various rigid and flexible endoscopes, laparoscopes, cameras, and headlamps,etc. It is intended for use in all types of endoscopic and endoscopic assisted procedures, as well as visualization of the evacuated cardiac chamber. ## PERFORMANCE DATA IEC 601-1 "General Safety Requirements for Medical Electrical Equipment" IEC 601-1-2 "Electrical Magnetic Compatibility" UL 544 "Standards for Safety, Medical and Dental Equipment" 510(k) 29
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...