K970182 · Johnson & Johnson Medical, Inc. · MHX · Aug 18, 1997 · Cardiovascular
Device Facts
Record ID
K970182
Device Name
DINAMAP COMPACT MONITOR
Applicant
Johnson & Johnson Medical, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Aug 18, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The DINAMAP Compact Monitor is intended to monitor a single adult, pediatric or neonatal patient’s vital signs at the bedside. Vital signs parameters include noninvasive blood pressure (systolic, diastolic and mean arterial pressure), pulse rate, temperature and/or oxygen saturation (pulse oximetry). The portable device is designed for use in numerous clinical settings, primarily in various hospital departments such as emergency, radiology, recovery, medical/surgical, labor & delivery, endoscopy, cardiac step-down, etc. and can also be used during many specialized procedures in satellite areas, or in ambulatory surgery centers, physicians’ offices or alternate care settings.
Device Story
Portable bedside monitor for adult, pediatric, neonatal patients; used in hospitals, surgery centers, physician offices. Inputs: oscillometric NIBP/heart rate; red/infrared spectroscopy (Nellcor pulse oximetry); thermistor-based temperature. Device processes signals to display vital signs; optional printer/recorder; connectivity to central stations/host systems. Operated by healthcare professionals. Benefits: continuous bedside monitoring for clinical decision-making.
Clinical Evidence
Clinical studies conducted for predictive thermometry accuracy and NIBP accuracy across adult, pediatric, and neonatal populations. Bench testing performed for pulse oximetry, environmental, and electromagnetic compatibility. No specific numerical results provided in summary.
Technological Characteristics
Software-driven electronic monitor. NIBP: oscillometry. Pulse Oximetry: red/infrared spectroscopy (Nellcor). Temperature: thermistor with predictive algorithm. Portable, battery/AC powered. Connectivity: central station/host system interface. No new technological characteristics compared to predicates.
Indications for Use
Indicated for monitoring vital signs (NIBP, pulse rate, temperature, SpO2) in adult, pediatric, and neonatal patients in hospital, outpatient, and clinical settings.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}
Johnson & Johnson MEDICAL INC.
K970182
AUG 18 1997
# 510(k) Summary
| Date | January 16, 1997 |
| --- | --- |
| Contact | Annette M. Hillring
Director, Regulatory Affairs
Johnson & Johnson Medical, Inc.
4110 George Road
Tampa, Florida 33634
Telephone: (813) 887-2256
Telefax: (813) 887-2263 |
| Device Name | DINAMAP* Compact Monitor |
| Common Names | Physiological or Vital Signs Monitor, Patient Monitor
Includes the following monitoring parameters:
• Noninvasive Blood Pressure & Heart Rate
• Pulse Oximetry & Heart Rate
• Predictive Temperature
An optional recorder is also available. |
| Classification | The classification names, 21 Code of Federal Regulations (CFR) Part and Paragraph numbers, and classification of the DINAMAP Compact Monitor and its monitoring parameters follow. The tier categorization based on the list (January 27, 1994) distributed by the Office of Device Evaluation is also included. |
Continued on next page
4110 GEORGE ROAD, TAMPA, FL 33634 • 813-887-2000
{1}
510(k) Summary, Continued
## Classification (continued)
| Classification Name | 21 CFR § & Class | Tier |
| --- | --- | --- |
| System, Measurement, Blood Pressure, Noninvasive | 870.1130 II | 2 |
| Computer, Blood Pressure | 870.1110 II | 2 |
| Alarm, Blood Pressure | 870.1100 II | 2 |
| Oximeter | 870.2700 II | 2 |
| Oximeter, Ear | 870.2710 II | 2 |
| Thermometer, Clinical Electronic | 880.2910 II | 2 |
| Recorder, Paper Chart | 870.2810 II | 1 |
| Display, Cathode-Ray Tube, Medical | 870.2450 II | 1 |
## Predicate Devices
The following table summarizes the predicate devices for the Compact Monitor and its monitoring parameters and 510(k) numbers:
| Compact | Predicate Device & Model | 510(k) Number(s) |
| --- | --- | --- |
| System | DINAMAP PLUS Monitor Model 9700 | K943709
K912188 |
| System | DINAMAP XL Monitor | K942700 |
| Temperature | DINAMAP XL Monitor | K942700 |
| NIBP | DINAMAP PLUS Monitor Model 9700 | K943709
K912188
K942700 |
| Oximetry | Nellcor® Model N-180 Pulse Oximeter
DINAMAP PLUS Monitor Model 9700 | K913695
K943709
K912188 |
Continued on next page
{2}
510(k) Summary, Continued
## Device Description
The DINAMAP Compact Monitor is a prescription device intended for use only by health care professionals. Three configurations of the monitor with the following vital signs parameters will be available:
- Noninvasive Blood Pressure and Heart Rate,
- Noninvasive Blood Pressure and Heart Rate, and Predictive Oral/Rectal Thermometry
- Noninvasive Blood Pressure and Heart Rate, Predictive Thermometry, and Nellcor® Pulse Oximetry and Heart Rate
In addition, the currently-marketed Sherwood Medical Inc. FirstTemp™ Genius™ Infrared Tympanic Thermometry Unit (K920713) may be physically attached to the side of (but not electrically integrated with) the DINAMAP Compact Monitor.
The device is designed for monitoring adult, pediatric and neonatal patients in hospital, outpatient surgery center, physician office and/or alternate healthcare settings. It is portable and capable of operation from an external power source or an internal lead-acid rechargeable battery. An optional printer is also available.
The Compact Monitor provides connection for the currently marketed Johnson & Johnson Medical, Inc. (JJMI, formerly Critikon, Inc.) OBSERVER® Central Station (K933404), other monitoring devices, a remote display, data collection system, remote alarm and/or host information system.
## Indications
The DINAMAP Compact Monitor is intended to monitor a single patient’s vital signs at the bedside. The patient populations include adult, pediatric and neonatal.
Continued on next page
iv - 3
{3}
iv - 4
# 510(k) Summary, Continued
## Technological Characteristics
The DINAMAP Compact Monitor and its monitoring parameters have the same technological characteristics as the predicate devices. There are no new technological characteristics. The Compact Monitor and the predicate devices are all software-driven electronic devices. The Compact monitoring parameters and predicate devices monitoring parameters utilize the following technologies:
- Noninvasive Blood Pressure & Heart Rate: Oscillometry
- Pulse Oximetry & Heart Rate: Nellcor®, Inc., red & infrared spectroscopy
- Thermometry: Thermistor with predictive algorithm
## Nonclinical Tests
Several bench studies were conducted which demonstrate safety and effectiveness of the Compact Monitor and monitoring parameters:
- Pulse Oximetry
- Environmental
- Electromagnetic Compatibility
## Clinical Tests
Several clinical studies were conducted which demonstrate safety and effectiveness of the Compact Monitor and monitoring parameters:
- Predictive Thermometry Accuracy
- Adult Noninvasive Blood Pressure Accuracy
- Pediatric Noninvasive Blood Pressure Accuracy
- Neonatal Noninvasive Blood Pressure Accuracy
## Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR Part 807, and based on the information provided in this premarket notification, Johnson & Johnson Medical concludes that the new device, the DINAMAP Compact Monitor is safe, effective and substantially equivalent to the predicate devices as described herein.
## Other Information
Johnson & Johnson Medical will update and include in this summary any other information deemed reasonably necessary by the FDA
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20856
AUG 18 1997
Ms. Annette M. Hillring
Johnson & Johnson Medical Inc.
4110 George Road
Tampa, Florida 33634
Re: K970182
DINAMAP* Compact Monitor
Regulatory Class: II (two)
Product Code: 74 MHX
Dated: May 19, 1997
Received: May 20, 1997
Dear Ms. Hillring:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}
Page 2 - Ms. Annette M. Hillring
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular,
Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{6}
# Indications for Use
510(k) Number: K970182
Device Name: DINAMAP* Compact Monitor
Indications for Use:
The DINAMAP Compact Monitor is intended to monitor a single adult, pediatric or neonatal patient’s vital signs at the bedside. Vital signs parameters include noninvasive blood pressure (systolic, diastolic and mean arterial pressure), pulse rate, temperature and/or oxygen saturation (pulse oximetry). The portable device is designed for use in numerous clinical settings, primarily in various hospital departments such as emergency, radiology, recovery, medical/surgical, labor & delivery, endoscopy, cardiac step-down, etc. and can also be used during many specialized procedures in satellite areas, or in ambulatory surgery centers, physicians’ offices or alternate care settings.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
Prescription Use ☑ (Per 21 CFR 801.109)
510(k) Number K970182
OR
Over-The-Counter Use
(Optional Format 1-2-96)
*Trademark
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.