← Product Code [LHL](/productcode/LHL) · K970149

# LEGIONELLA ELISA TEST SYSTEM (K970149)

_Zeus Scientific, Inc. · LHL · Jun 18, 1997 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K970149

## Device Facts

- **Applicant:** Zeus Scientific, Inc.
- **Product Code:** [LHL](/productcode/LHL.md)
- **Decision Date:** Jun 18, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3300
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

The Zeus Scientific, Inc. Legionella ELISA test system is intended for the quantitative and/or quantitative detection of IgG/A/M-class antibody to Legionella pneumophila Groups 1 - 6 in human serum. The test system is intended to be used as an aid in the diagnosis of Legionella-related infections. This test is for in vitro diagnostic use.

## Device Story

The Legionella ELISA Test System is an in vitro diagnostic assay used to detect IgG, IgA, and IgM antibodies against Legionella pneumophila (serogroups 1-6) in human serum samples. The device utilizes enzyme-linked immunosorbent assay (ELISA) technology to identify specific antibodies. It is intended for use in clinical laboratory settings by trained laboratory personnel. The test results provide qualitative or quantitative data to assist healthcare providers in diagnosing Legionella-related infections. By identifying the presence of these antibodies, the test aids in clinical decision-making regarding patient management and treatment for suspected Legionellosis.

## Clinical Evidence

No clinical data provided in the document.

## Technological Characteristics

ELISA-based immunoassay for the detection of IgG/A/M-class antibodies to Legionella pneumophila. In vitro diagnostic device.

## Regulatory Identification

Haemophilus spp. serological reagents are devices that consist of antigens and antisera, including antisera conjugated with a fluorescent dye, that are used in serological tests to identify Haemophilus spp. directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genus Haemophilus and provides epidemiological information on diseases cause by these microorganisms. Diseases most often caused by Haemophilus spp. include pneumonia, pharyngitis, sinusitis, vaginitis, chancroid venereal disease, and a contagious form of conjunctivitis (inflammation of eyelid membranes).

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

DEPARTMENT OF HEALTH &amp; HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850

JUN 18 1997

Mark J. Kopnitsky
Vice President
Research &amp; Development
Zeus Scientific, Inc.
P.O. Box 38
Raritan, NJ 08869

Re: K970149
Trade Name: Legionella ELISA Test System
Regulatory Class: II
Product Code: LHL
Dated: June 4, 1997
Received: June 9, 1997

Dear Mr. Kopnitsky:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

{1}

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical Laboratory Devices
Office of Device Evaluation
Center for Devices and Radiological Health

Enclosure

{2}

Page 1 of 1

510(k) Number (if known): _________________________

Device Name: *Legionella ELISA Test System*

Indications For Use:

The Zeus Scientific, Inc. Legionella ELISA test system is intended for the quantitative and/or quantitative detection of IgG/A/M-class antibody to *Legionella pneumophila* Groups 1 - 6 in human serum. The test system is intended to be used as an aid in the diagnosis of Legionella-related infections. This test is for *in vitro* diagnostic use.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number: _________________________

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use
(Optional Format 1-2-96)

2

---

**Source:** [https://fda.innolitics.com/device/K970149](https://fda.innolitics.com/device/K970149)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
