TANDEMRX

K970054 · Boston Scientific Corp · GCJ · Feb 10, 1997 · Gastroenterology, Urology

Device Facts

Record IDK970054
Device NameTANDEMRX
ApplicantBoston Scientific Corp
Product CodeGCJ · Gastroenterology, Urology
Decision DateFeb 10, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The TandemRX is indicated for use to cannulate and inject contrast media to obtain a cholangiogram of the biliary duct system. The contrast media is injected through the cannula and fluoroscopy or x-ray is performed to obtain the cholangiogram.

Device Story

Triple lumen ERCP cannula; facilitates biliary duct cannulation and contrast media injection. Features one open channel for .035” guidewire passage and two lumens for contrast media injection (joined at proximal end for single-port access). Allows guidewire exchange isolated from injection agents; eliminates need for stylet during scope passage. Used in clinical settings by gastroenterologists/endoscopists during ERCP procedures. Enables visualization of biliary duct system via fluoroscopy/x-ray; assists in diagnostic or therapeutic biliary interventions.

Clinical Evidence

Bench testing only. Biocompatibility assessment performed on patient- and fluid-contact materials with satisfactory results. No clinical data presented.

Technological Characteristics

Triple lumen cannula; components include shaft, hub, guidewire introducer, and outer sheath. Compatible with .035” guidewire. Materials are biocompatible. Non-powered, manual surgical instrument.

Indications for Use

Indicated for patients requiring cannulation of the biliary duct system for cholangiogram procedures via contrast media injection and fluoroscopy/x-ray.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1} K 97 0054 SECTION 9 510(K) SUMMARY FEB 10 1997 ## FOI RELEASABLE Pursuant to §513(i)(3)(A) of the Food, Drug, and Cosmetic Act, Boston Scientific Corporation is required to submit with this Premarket Notification “...adequate summary of any information respecting safety and effectiveness or state that such information will be made available upon request of any person.” Boston Scientific Corporation chooses to submit a summary of information respecting safety and effectiveness. &gt; **DATE:** January 6, 1997 &gt; **COMMON/USUAL NAMES:** Injection Cannula &gt; **TRADE/PROPRIETARY NAME:** Unknown at this Time &gt; **CLASSIFICATION NAME &amp; DEVICE CLASSIFICATION:** Class I | Name | Number | 21 CFR Ref. | | --- | --- | --- | | Cannula, Injection | 79 FGY | KNS | &gt; **DEVICE PANEL/BRANCH:** Gastroenterology-Urology (GU)<br/> Gastro-Renal (GRDB) &gt; **OWNER/OPERATOR:** Boston Scientific Corporation<br/> One Boston Scientific Place<br/> Natick, MA 01760 &gt; **CONTACT PERSON:** Lisa M. Quaglia, Senior Regulatory Affairs Specialist ## DESCRIPTION The Microvasive TandemRX is a triple lumen ERCP cannula capable of accepting an .035” guidewire in one open channel while simultaneously injecting contrast media through the other two lumens. The two lumens are joined at the proximal end so that injection of contrast medium can be accomplished through a single injection port. The open channel allows for the quick exchange of a guidewire completely isolated from injection agents. No stylet is necessary for scope passage. The TandemRX may be placed with or without the aid of a guidewire. Premarket Notification, TandemRX, January 6, 1997 Proprietary and Confidential Information of Boston Scientific Corporation {2} # INDICATIONS FOR USE The TandemRX is indicated for use to cannulate and inject contrast media to obtain a cholangiogram of the biliary duct system. The contrast media is injected through the cannula and fluoroscopy or x-ray is performed to obtain the cholangiogram. # DESCRIPTIVE AND TECHNOLOGICAL CHARACTERISTICS OF PROPOSED AND PREDICATE DEVICES Boston Scientific Corporation believes that the TandemRX is substantially equivalent to the currently-marketed Microvasive Tandem XL Cannula. The major components of the TandemRX are the shaft, hub, guidewire introducer, and the outer sheath. A thorough comparison of the descriptive characteristics between the TandemRX and the predicate device shows equivalence. # PERFORMANCE CHARACTERISTICS Laboratory testing regarding characteristics was performed on TandemRX to verify its safety and performance. A biocompatibility assessment was performed on the patient- and fluid-contact materials of the TandemRX with satisfactory results. # CONCLUSION Boston Scientific Corporation believes that TandemRX is substantially equivalent to the currently-marketed Microvasive Tandem XL. The Modified TandemRX is equivalent in its indications for use, while being very similar in design and materials. In addition, Boston Scientific Corporation has presented laboratory testing and biocompatibility information. The information presented provides assurance that the TandemRX will meet the minimum requirements that are considered acceptable for its intended use. Premarket Notification, TandemRX, January 6, 1997 Proprietary and Confidential Information of Boston Scientific Corporation
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...