The Borrelia burgdorferi IgG/IgM ELISA kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative presumptive detection of total (IgG/IgM) antibodies to Borrelia burgdorferi in human serum. This ELISA should only be used for patients with signs and patients with symptoms that are consistent with Lyme disease. Equivocal or positive results should be supplemented by testing with a standardized Western blot procedure. Positive supplemental results are supportive evidence of exposure to B. burgdorferi and can be used to support a clinical diagnosis of Lyme disease.
Device Story
ELISA test kit for qualitative presumptive detection of total IgG/IgM antibodies to Borrelia burgdorferi in human serum; utilizes purified B. burgdorferi antigen attached to solid-phase microtiter wells. Patient serum added to wells; specific antibodies bind to antigen; unbound antibodies washed away. Enzyme-labeled anti-human IgG/IgM conjugate added; binds to captured antibodies. Substrate solution added; enzyme presence triggers color change. Reaction stopped; color intensity measured photometrically. Indirect measurement of specific antibody concentration. Used in clinical laboratory settings by trained technicians. Results support clinical diagnosis of Lyme disease when combined with standardized Western blot procedure for equivocal or positive findings.
Clinical Evidence
Evaluated using CDC Lyme disease serum panel (n=47) and fresh clinical sera (n=176). CDC panel showed 71% agreement with clinical diagnosis. Fresh sera study compared device to Lyme STAT and Western Blot; device showed 4% (7/176) positivity in 2-step testing (ELISA + WB). Precision testing (n=810 determinations) performed per NCCLS EP5; CVs generally <15% for most samples. Cross-reactivity assessed against lipemic, bilirubinemic, RPR+, dsDNA+, RF+, EBV+, CMV+, RMS+, elevated ESR, and CRP samples; minimal cross-reactivity observed.
Technological Characteristics
Enzyme-Linked Immunosorbent Assay (ELISA) using purified B. burgdorferi antigen on solid-phase microtiter wells. Photometric detection of colorimetric substrate reaction. Manual or semi-automated laboratory procedure. No electronic connectivity or software algorithms described.
Indications for Use
Indicated for patients with signs and symptoms consistent with Lyme disease to detect total IgG/IgM antibodies to Borrelia burgdorferi in human serum.
Regulatory Classification
Identification
Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), the Treponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies to Treponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genus Treponema and provides epidemiological information on syphilis.
Predicate Devices
BioWhittaker’s Lyme STAT test
Reference Devices
Mardx Diagnostics Western Blot
Related Devices
K965129 — BORRELIA BURGDORFERI IGM ELISA TEST SYSTEM · Wampole Laboratories · Mar 26, 1997
K983605 — IS BORRELIA BURGDORFERI IGG/IGM ELISA TEST SYSTEM · Columbia Bioscience, Inc. · Dec 16, 1998
K033083 — BORRELIA BURGDORFERI IGG/IGM ELISA TEST SYSTEM · Trinity Biotech USA · Nov 26, 2003
Submission Summary (Full Text)
{0}
Summary of Safety and Effectiveness Information Borrelia burgdorferi IgG/IgM ELISA Test Kit
K965131
I. Immuno Probe Inc.
1306 Bailes Lane, Suite F
Frederick, Maryland 21701
Contact person: William Boteler
Telephone: 301-695-7920
Date of preparation: March 4, 1997
MAR 26 1997
## II. Description of Device
The Borrelia burgdorferi IgG/IgM ELISA kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative presumptive detection of total (IgG/IgM) antibodies to Borrelia burgdorferi in human serum. This ELISA should only be used for patients with signs and patients with symptoms that are consistent with Lyme disease. Equivocal or positive results should be supplemented by testing with a standardized Western blot procedure. Positive supplemental results are supportive evidence of exposure to B. burgdorferi and can be used to support a clinical diagnosis of Lyme disease.
The Borrelia burgdorferi IgG/IgM ELISA test is an enzyme linked immunosorbent assay to detect IgG/IgM antibodies to Borrelia burgdorferi. Purified Borrelia burgdorferi antigen is attached to a solid phase microtiter well. Diluted test sera is added to each well. If the antibodies are present that recognize the antigen, they will bind to the antigen in the well. After incubation the wells are washed to remove unbound antibody. An enzyme labeled anti-human IgG/IgM is added to each well. If antibody is present it will bind to the antibody attached to the antigen on the well. After incubation the wells are washed to remove unbound conjugate. A substrate solution is added to each well. If enzyme is present the substrate will undergo a color change. After an incubation period the reaction is stopped and the color intensity is measured photometrically, producing an indirect measurement of specific antibody in the patient specimen.
## III. Predicate Device
The Borrelia burgdorferi IgG/IgM ELISA test is substantially equivalent to BioWhittaker’s Lyme STAT test. Equivalence is demonstrated by the following comparative results:
{1}
# Performance Characteristics
## Table 1 The CDC Lyme Disease Serum Panel Stratified by Time After Onset.
The following information is from a serum panel obtained from the CDC and tested by Wampole. The results are presented as a means to convey further information on the performance of this assay with a masked, characterized serum panel. This does not imply an endorsement of the assay by the CDC.
**Wampole B. burgdorferi IgG/IgM ELISA Result**
| Time From Onset | + | Equivocal | - | Total | % Agreement with Clinical Diagnosis |
| --- | --- | --- | --- | --- | --- |
| normals | 0 | 0 | 5 | 5 | 100% |
| <1 month | 2 | 0 | 3 | 5 | 40.0% |
| 1-2 months | 6 | 2 | 2 | 10 | 80.0% |
| 3-12 months | 9 | 3 | 7 | 19 | 63.3% |
| > 1yr | 8 | 0 | 0 | 8 | 100.0% |
| Total | 25 | 5 | 17 | 47 | |
Because the equivocals would have been tested by immunoblotting in a 2-step test system, they were considered 1-step positive for purposes of calculating % agreement. The Wampole Borrelia burgdorferi IgG/IgM ELISA demonstrated 71% (30/42) agreement (sensitivity) with clinical diagnosis of Lyme disease.
{2}
Study 2: 176 fresh sera from patients of various ages and genders that were submitted to a large clinical lab for B. burgdorferi antibody testing were tested on the Wampole B. burgdorferi IgG/IgM ELISA and Lyme Stat. Any serum found positive or equivocal was tested by Mardx Diagnostics Western Blot on both IgG and IgM.
Table 2: Western blot results
| | | Western Blot | |
| --- | --- | --- | --- |
| | | + | - |
| Wampole | + | 6 | 3 |
| | eq. | 1 | 6 |
| Lyme Stat | + | 5 | 3 |
| --- | --- | --- | --- |
| | eq. | 0 | 5 |
| Type of Result | Wampole (95%CI) | Lyme Stat (95%CI) |
| --- | --- | --- |
| 1-step (ELISA) pos. or eq. | 9.1% (4.8%-13.4%)
(16/176) | 7.4% (3.4%-11.3%)
(13/176) |
| 1-step pos. or eq. &
2-step (WB) pos. | 4% (1.0%-6.9%)
(7/176) | 2.8% (0.3%-5.3%)
(5/176) |
| 2-step pos. among
1-step pos. or eq. | 44% (18.9%-68.6%)
(7/16) | 39% (11.5%-65.4%)
(5/13) |
{3}
2. Precision. Seven sera were assayed ten times each on three different plates at three different sites. An additional three serum were assayed ten times each on three different plates at two different sites. The intersite precision is shown in Table 3. With appropriate technique the user should obtain precision of <15% CV.
Table 3
*Borrelia burgdorferi* IgG/IgM ELISA Inter Assay Precision Between Sites
| Inter-Assay (n=90) | | | | |
| --- | --- | --- | --- | --- |
| # | X | SD | CV | n |
| 1. | 1.18 | 0.113 | 9.58% | 90 |
| 2. | 2.26 | 0.226 | 10.02% | 90 |
| 3. | 3.85 | 0.278 | 7.21% | 90 |
| 4. | 5.36 | 0.402 | 7.50% | 90 |
| 5. | 2.04 | 0.186 | 9.11% | 90 |
| 6. | 0.33 | 0.132 | 40.57% | 90 |
| 7. | 0.34 | 0.234 | 68.93% | 90 |
| 8. | 1.47 | 0.152 | 10.34% | 60 |
| 9. | 1.47 | 0.104 | 7.10% | 60 |
| 10. | 1.65 | 0.131 | 7.95% | 60 |
| HPC | 5.31 | 0.314 | 5.91% | 15 |
| LPC | 2.70 | 0.107 | 3.96% | 15 |
| NC | 0.20 | 0.043 | 21.97% | 15 |
| CAL | 3.34 | 0.103 | 3.08% | 45 |
A total of 810 determinations were made at the three sites. In all 810 determinations there was not a cas a positive result for a negative serum or a negative result for a positive serum.
X = Mean ISR
SD = standard deviation
CV = coefficient of variation = SD/X x 100
The methods in NCCLS EP5 were utilized for precision parameters.
{4}
3. Cross-Reactivity. The following potentially cross-reactive sera were run on the Borrelia burgdorferi IgG/IgM ELISA assay to assess cross-reactivity with the assay: lipemic, bilirubinemic, RPR +, dsDNA +, RF +, EBV +, CMV +, RMS +, elevated ESR, and CRP. The data in Table 4 illustrates the amount of reactivity with the sera.
TABLE 4
Cross Reactivity
| Laboratory result (Titer) | # of samples | # of positives |
| --- | --- | --- |
| Lipemic (+++) | 5 | 0 |
| Bilirubinemic (1.9-16.9) | 5 | 1 |
| RPR + (1:2 - 1:64) | 10 | 0 |
| Rheumatoid Factor + (1:40-1:320) | 3 | 0 |
| Epstein Barr Virus Antibody + (1:40-1:2560) | 7 | 0 |
| Cytomegalovirus Antibody + (O.D. 0.718-2.308) | 6 | 1 |
| Rocky Mt Spotted Fever Antibody + (1:256-1:16.384) | 4 | 0 |
| CRP + (2.79-8.61 mg/dl) | 5 | 0 |
| Elevated ESR (40-115) | 10 | 0 |
| dsDNA + (52.3-1072 IU) | 16 | 0 |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.