← Product Code [FDX](/productcode/FDX) · K965037

# CELL RECOVERY SYSTEM (K965037)

_Medical Device Technologies, Inc. · FDX · Jan 8, 1997 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K965037

## Device Facts

- **Applicant:** Medical Device Technologies, Inc.
- **Product Code:** [FDX](/productcode/FDX.md)
- **Decision Date:** Jan 8, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The MDT Cell Recovery System™ (CRS) is an automated cytology brush for use with a rigid or flexible cystoscope in the collection and retrieval of cells for cytological evaluation.

## Device Story

Automated cytology brush system for cell collection; used with rigid/flexible cystoscopes (6 French working channel). Components: nylon bristles, stainless steel wires, nickel-titanium shaft, Teflon sheath. Battery-operated console with foot pedal control; automates brush rotation and aspiration of dislodged cells into 60 cc syringe. Used in clinical settings by physicians; enables hands-free operation during procedure. Facilitates cytological evaluation by improving sample retrieval efficiency.

## Clinical Evidence

Bench testing only. No clinical data presented.

## Technological Characteristics

Materials: nylon bristles, stainless steel wires, nickel-titanium shaft, Teflon sheath. Dimensions: 10 mm brush diameter, 6.4 mm brush length, 156 cm sheath length. Energy source: battery-operated console. Operation: automated rotation and aspiration via foot pedal. Single-use device.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- CRS (cleared March 20, 1996)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K965037

JAN 8 1997

# SUMMARY OF SAFETY AND EFFECTIVENESS

## MEDICAL DEVICE TECHNOLOGIES, INC.

## CELL RECOVERY SYSTEM™ (CRS) 510(k)

|  Company: | Medical Device Technologies, Inc.
9171 Towne Centre Drive, Suite 355
San Diego, CA 92122
Tel: (619) 455-7127
Fax: (916) 455-7295  |
| --- | --- |
|  Company Representative: | Dick Sloan
VP New Business Development  |
|  Date 510(k) Prepared: | December 16, 1996  |
|  Device Name: | Cell Recovery System™  |
|  Classification Name: | Endoscope and Accessories  |
|  Classification: | Class II  |

The MDT Cell Recovery System™ (CRS) is an automated cytology brush for use with a rigid or flexible cystoscope in the collection and retrieval of cells for cytological evaluation. The brush, consisting of nylon bristles secured by twisted stainless steel wires fused and blunted at the tip, is attached to a nickel-titanium shaft and housed in a Teflon sheath. The brush and sheath are designed for single use with cystoscopes having a 6 French working channel. The brush diameter is 10 mm, the brush length is 6.4 mm, and the sheath is 156 cm long. This cytology brush is similar in size, features and components to other commercially available cytology brushes, including the previous CRS version, and has the same intended use.

The CRS facilitates brushing of cells by automating the rotation of the cytology brush and aspiration of dislodged cells using a battery operated console coupled to the brush. The console is operated using a foot pedal for hands free operation. The operator can maneuver the brush to the desired location and initiate the brushing procedure as well as the sample retrieval into a 60 cc syringe.

Results from bench-top testing showed the modified CRS is substantially equivalent to the previous version of the CRS device cleared for marketing on March 20, 1996. These studies have demonstrated that the modified CRS features do not pose new safety questions when compared to the currently marketed CRS device.

---

**Source:** [https://fda.innolitics.com/device/K965037](https://fda.innolitics.com/device/K965037)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
