EMERALD MEDICAL, INC. NUCLEUS PULPOSIS EVACUATOR SYSTEM
Device Facts
| Record ID | K964919 |
|---|---|
| Device Name | EMERALD MEDICAL, INC. NUCLEUS PULPOSIS EVACUATOR SYSTEM |
| Applicant | Emerald Medical, Inc. |
| Product Code | HRX · Orthopedic |
| Decision Date | Jan 28, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.1100 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The EMERALD MEDICAL, INC. Nucleus Pulposis Evacuator System is an assortment of surgical instruments utilized in minimally invasive percutaneous lumbar discectomy surgical procedures.
Device Story
Surgical instrument kit for minimally invasive percutaneous lumbar discectomy; used by surgeons in clinical/OR settings; facilitates disc removal; components include introduction tools and evacuation instruments; manual operation; provides mechanical assistance for tissue removal; benefits include reduced invasiveness compared to traditional open discectomy.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Manual surgical instruments; materials consistent with predicate arthroscopic discectomy systems; form factor designed for percutaneous access; non-powered; non-sterile (intended for sterilization).
Indications for Use
Indicated for patients undergoing minimally invasive percutaneous lumbar discectomy surgical procedures.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
Predicate Devices
- Acufex Microsurgical Interceptre Arthroscopic Discectomy System (K941709)
- Smith & Nephew Arthroscopic MicroDiscectomy System (AMD)
Reference Devices
- Surgical Dynamics instrument introduction systems
- Sofamor Danek instrument introduction systems
Related Devices
- K960222 — PERCUTANEOUS ENDOSCOPIC DISCECTOMY · Richard Wolf Medical Instruments Corp. · Mar 8, 1996
- K961713 — INSTRUMENT SET FOR ARTHROSCOPIC PERCUTANEOUS DISCECTOMY · KARL STORZ Endoscopy-America, Inc. · May 31, 1996
- K082194 — LAURIMED PERCUTANEOUS DISCECTOMY SYSTEM, MODEL FG-000002 · Laurimed, LLC · Aug 28, 2008
- K972768 — BLACKSTONE MEDICAL, INC. LAPAROSCOPIC DISC REMOVAL SYSTEM · Blackstone Medical, Inc. · Sep 25, 1997
- K973405 — YEUNG ENDOSCOPIC SPINE SYSTEM · Richard Wolf Medical Instruments Corp. · Mar 13, 1998