IMMAGE IMMUNOCHEMISTRY SYSTEM COMPLEMENTC3 (C3) AND COMPLEMENTC4 (C4) REAGENT
K964842 · Beckman Instruments, Inc. · CZW · Jun 11, 1997 · Immunology
Device Facts
Record ID
K964842
Device Name
IMMAGE IMMUNOCHEMISTRY SYSTEM COMPLEMENTC3 (C3) AND COMPLEMENTC4 (C4) REAGENT
Applicant
Beckman Instruments, Inc.
Product Code
CZW · Immunology
Decision Date
Jun 11, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5240
Device Class
Class 2
Indications for Use
The IMMAGE Immunochemistry System Complement C3 (C3) Reagent, when used in conjunction with Beckman IMMAGE™ Immunochemistry Systems and Beckman Calibrator 1, is intended for the quantitative determination of Complement C3 by rate nephelometry. The IMMAGE Immunochemistry System Complement C4 (C4) Reagent, when used in conjunction with Beckman IMMAGE™ Immunochemistry Systems and Beckman Calibrator 1, is intended for the quantitative determination of Complement C4 by rate nephelometry.
Device Story
Reagents for quantitative measurement of Complement C3 and C4 in human serum; used on Beckman IMMAGE Immunochemistry System. System utilizes rate nephelometry; measures light scatter intensity changes over time as antigen-antibody complexes form. Reagents stored in plastic cartridges; assay performed at 37°C. Operated by laboratory personnel in clinical settings. Output provides concentration values (mg/dL) to clinicians for diagnosing immunologic disorders and complement deficiencies. Benefits include automated, precise quantification of complement proteins compared to manual or older methods.
Clinical Evidence
Bench testing only. Method comparison study (n=100 for C3, n=124 for C4) against predicate Array System reagents showed high correlation (r=0.994 for C3; r=0.984 for C4). Precision studies (n=80 per level) demonstrated total C.V.s of 3.0-3.6% for C3 and 3.4-5.2% for C4. Stability testing supports 24-month shelf-life and 14-day open-container/calibration stability.
Technological Characteristics
Rate nephelometry assay; 37°C reaction temperature. Reagents supplied in plastic cartridges. Quantitative measurement of C3 and C4 proteins in human serum. System-based automated analysis.
Indications for Use
Indicated for quantitative determination of human Complement C3 and C4 in serum via rate nephelometry to aid in the diagnosis of immunologic disorders associated with complement component deficiencies.
Regulatory Classification
Identification
A complement components immunological test system is a device that consists of the reagents used to measure by immunochemical techniques complement components C1q , C1r , C1s , C2 , C3 , C4 , C5 , C6 , C7 , C8 , and C9 , in serum, other body fluids, and tissues. Complement is a group of serum proteins which destroy infectious agents. Measurements of these proteins aids in the diagnosis of immunologic disorders, especially those associated with deficiencies of complement components.
Predicate Devices
Immunochemistry Systems Complement C3 Reagent (K771603)
Immunochemistry Systems Complement C4 Reagent (K780913)
Submission Summary (Full Text)
{1}
200 South Kraemer Boulevard, Box 8000, Brea, CA 92822-8000 • (714) 993-5321
K964842
# BECKMAN
Summary of Safety & Effectiveness
IMMAGE™ Immunochemistry System Complement C3 (C3) and Complement C4 (C4) Reagents
# 1.0 Submitted By:
Margie George
Project Manager
Beckman Instruments, Inc.
200 S. Kraemer Blvd., W-343
Brea, California 92822-8000
Telephone: (714) 961-3765
FAX: (714) 961-3759
# 2.0 Date Submitted:
26 November 1996
# 3.0 Device Name(s):
# 3.1 Proprietary Names
IMMAGE™ Immunochemistry System Complement C3 (C3) Reagent
IMMAGE™ Immunochemistry System Complement C4 (C4) Reagent
# 3.2 Classification Name
Complement components) immunological test system (21 CFR § 866.5240)
# 4.0 Predicate Device(s):
| IMMAGE System Reagent | Predicate | Manufacturer | Docket Number |
| --- | --- | --- | --- |
| Complement C3 (C3) Reagent | Immunochemistry Systems Complement C3 Reagent | Beckman Instruments Inc. * | K771603 |
| Complement C4 (C4) Reagent | Immunochemistry Systems Complement C4 Reagent | Beckman Instruments Inc.* | K780913 |
Beckman Instruments, Inc.
twx: 910-592-1260 • telex: 06-78413
{2}
Beckman Instruments, Inc., Section 510(k) Notification
IMMAGE™ Immunochemistry System Complement C3 (C3) and Complement C4 (C4) Reagents
Summary of Safety & Effectiveness
2
# 5.0 Description:
The IMMAGE Immunochemistry System C3 and C4 Reagents in conjunction with Beckman Calibrator 1, are intended for use in the quantitative determination of Complement C3 and Complement C4 concentrations respectively in human serum samples on Beckman's IMMAGE Immunochemistry System.
# 6.0 Intended Use:
The IMMAGE Immunochemistry System Complement C3 (C3) Reagent, when used in conjunction with Beckman IMMAGE™ Immunochemistry Systems and Beckman Calibrator 1, is intended for the quantitative determination of Complement C3 by rate nephelometry.
The IMMAGE Immunochemistry System Complement C4 (C4) Reagent, when used in conjunction with Beckman IMMAGE™ Immunochemistry Systems and Beckman Calibrator 1, is intended for the quantitative determination of Complement C4 by rate nephelometry.
file: c3-4510k.sse
{3}
Beckman Instruments, Inc., Section 510(k) Notification
IMMAGE™ Immunochemistry System Complement C3 (C3) and Complement C4 (C4) Reagents
Summary of Safety & Effectiveness
3
# 7.0 Comparison to Predicate(s):
The following tables show similarities and differences between the predicates identified in Section 4.0 of this summary.
C3 Similarities With The Predicate
| Reagent | Aspect/Characteristic | Comments |
| --- | --- | --- |
| IMMAGE System C3 Reagent | Assay method - rate nephelometry | Same as Beckman's Immunochemistry Systems Complement C3 reagent on the Array System |
| | Sample / reagent ratios | |
| | Antibody | |
C4 Similarities With The Predicate
| Reagent | Aspect/Characteristic | Comments |
| --- | --- | --- |
| IMMAGE System C4 Reagent | Assay method - rate nephelometry | Same as Beckman's Immunochemistry Systems Complement C4 reagent on the Array System |
| | Sample / reagent ratios | |
| | Measuring Range: Initial 10 - 130 mg/dL Extended 10 - 4,680 mg/dL | |
| | Antibody | |
C3 Differences From The Predicate
| Reagent | Aspect/Characteristic | Comments |
| --- | --- | --- |
| IMMAGE System C3 Reagent | Reaction Temperature | IMMAGE System C3 assays run at 37°C and the predicate runs at 26.7°C. |
| IMMAGE System C3 Reagent | Reagent Container | IMMAGE reagent is stored in a plastic cartridge and predicate reagent is stored in glass vials. |
C4 Differences From The Predicate
| Reagent | Aspect/Characteristic | Comments |
| --- | --- | --- |
| IMMAGE System C4 Reagent | Reaction Temperature | IMMAGE System C4 assays run at 37°C and the predicate runs at 26.7°C. |
| IMMAGE System C4 Reagent | Reagent Container | IMMAGE reagent is stored in a plastic cartridge and predicate reagent is stored in glass vials. |
file: c3-4510k.sse
{4}
Beckman Instruments, Inc., Section 510(k) Notification
IMMAGE™ Immunochemistry System Complement C3 (C3) and Complement C4 (C4) Reagents
Summary of Safety & Effectiveness
4
### 8.0 Summary of Performance Data:
The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. Equivalence is demonstrated through method comparison, stability, and imprecision experiments that relate results obtained from the Beckman Reagents to the IMMAGE System Reagents.
Method Comparison Study Results
IMMAGE Complement C3 (C3) and Complement C4 (C4) Reagents
| Analyte | Sample Type | Slope | Intercept | r | n | Predicate Method |
| --- | --- | --- | --- | --- | --- | --- |
| Complement C3 | serum | 0.969 | 6.18 | 0.994 | 100 | Complement C3 Reagent on the Array |
| Complement C4 | serum | 0.983 | -1.48 | 0.984 | 124 | Complement C4 Reagent on the Array |
Stability Study Results
| Reagent | Product Claim |
| --- | --- |
| IMMAGE C3 & C4 | 24 month shelf-life 14 day open container stability 14 day calibration stability |
file: c3-4510k.sse
{5}
Beckman Instruments, Inc., Section 510(k) Notification
IMMAGE™ Immunochemistry System Complement C3 (C3) and Complement C4 (C4) Reagents
Summary of Safety & Effectiveness
5
### Estimated Imprecision
IMMAGE System C3 Reagent Precision Study Results
| Sample | Mean (mg/dL) | S.D. (mg/dL) | %C.V. | N |
| --- | --- | --- | --- | --- |
| Within-Run Imprecision | | | | |
| Level 1 | 70.3 | 1.79 | 2.5 | 80 |
| Level 2 | 106 | 3.1 | 2.9 | 80 |
| Total Precision | | | | |
| Level 1 | 70.3 | 2.09 | 3.0 | 80 |
| Level 2 | 106 | 3.9 | 3.6 | 80 |
IMMAGE System C4 Reagent Precision Study Results
| Sample | Mean (mg/dL) | S.D. (mg/dL) | %C.V. | N |
| --- | --- | --- | --- | --- |
| Within-Run Imprecision | | | | |
| Level 1 | 18.8 | 0.60 | 3.2 | 80 |
| Level 2 | 42.9 | 1.45 | 3.4 | 80 |
| Level 3 | 65.5 | 1.65 | 2.5 | 80 |
| Total Precision | | | | |
| Level 1 | 18.8 | 0.65 | 3.4 | 80 |
| Level 2 | 42.9 | 2.21 | 5.2 | 80 |
| Level 3 | 65.5 | 2.46 | 3.7 | 80 |
This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.
file: c3-4510k.sse
{6}
[LOGO]
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
JUN 11 1997
Ms. Margie George
Project Manager
Beckman Instruments, Inc.
200 S. Kraemer Boulevard, W-343
Brea, California 92822-8000
Re: K964842/S1
* Trade Name: IMMAGE™ Immunochemistry System Complement C3 and C4 Reagents
Regulatory Class: II
Product Code: CZW
II
DBI
Dated: March 27, 1997
Received: March 28, 1997
Dear Ms. George:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
{7}
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Gutman
Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical
Laboratory Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{8}
page __ of __
510(k) Number (if known): K964842
Device Name: IMMAGE™ Immunochemistry System
Complement C4 (C4) Reagent
Indications for Use:
The IMMAGE Immunochemistry System Complement C4 (C4) Reagent, when used in conjunction with Beckman IMMAGE™ Immunochemistry Systems and Beckman Calibrator 1, is intended for the quantitative determination of human complement C4 by rate nephelometry.
21 CFR 866.5240 Complement components immunological test system
(a) Identification. A Complement components immunological test system is a device that consists of the reagents used to measure by immunochemical techniques complement components C₁q, C₁r, C₁s, C₂, C₃, C₄, C₅, C₆, C₇, C₈, and C₉ in serum, other body fluids, and tissues. Complement is a group of serum proteins which destroy infectious agents. Measurements of these proteins aids in the diagnosis of immunologic disorders, especially those associated with deficiencies of complement components.
(b) Classification. Class II (performance standards).
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☑
(per 21 CFR 801.109)
OR
Over-the-Counter Use ____
Optional Format 1-2-96
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.