← Product Code [LMB](/productcode/LMB) · K964803

# UNIVISION ECHOCARDIOLOGY SYSTEM (K964803)

_Network Concepts, Inc. · LMB · Feb 11, 1997 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K964803

## Device Facts

- **Applicant:** Network Concepts, Inc.
- **Product Code:** [LMB](/productcode/LMB.md)
- **Decision Date:** Feb 11, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2010
- **Device Class:** Class 1
- **Review Panel:** Radiology

## Indications for Use

The UNIVISION™ Echocardiology System is a network system intended to use digital technology to acquire images from an echocardiography ultrasound system for viewing outside the exam room and long term storage and retrieval on digital media.

## Device Story

System acquires echocardiography ultrasound images via Image Acquisition Unit (IAU); transmits data over network to workstations for review; archives exams on optical disc media (CD-R jukebox). Used in hospital/clinical environments by cardiology staff. Employs MPEG-1 (ISO/IEC 11172) compression (30:1 ratio) for efficient transmission and storage. Enables remote viewing of exams in progress or historical studies. Facilitates clinical workflow by providing centralized access to echocardiographic data for diagnostic review.

## Clinical Evidence

Bench testing only. Comparison of original ultrasound images versus MPEG-1 compressed images demonstrated no significant difference in image quality.

## Technological Characteristics

Networked image management system; comprises Image Acquisition Unit (IAU), Workstation, and Archive Subsystem (jukebox server/CD-R). Uses MPEG-1 (ISO/IEC 11172) standard for video/audio compression at 30:1 ratio. Connectivity via network interface.

## Regulatory Identification

A medical image storage device is a hardware device that provides electronic storage and retrieval functions for medical images. Examples include electronic hardware devices employing magnetic and optical discs, magnetic tapes, and digital memory.

## Predicate Devices

- TomTec P90 system ([K950279](/device/K950279.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FEB 11 1997
K964803

# PREMARKET NOTIFICATION SUMMARY
## UNIVISION™ ECHOCARDIOLOGY SYSTEM
### Network Concepts, Inc.

## SUBMITTER:
Larry Sieb
Division Manager, Cardiology,
NetworkConcepts, Inc.
2135 W. Greenview Drive
Middleton, WI 53562

Tel: (608) 836-9334
Fax: (608) 831-8333

## DEVICE NAME:

Classification: Class II

Common/Usual Name: Mini-Pacs (Picture Archiving and Communications System): Image Management System

Proprietary Name: UNIVISION™ Echocardiology System

## INTENDED USE:

The UNIVISION™ Echocardiology System is a network system intended to use digital technology to acquire images from an echocardiography ultrasound system for viewing outside the exam room and long term storage and retrieval on digital media. The substantial equivalent device is TomTec Imaging Systems’ TomTec P90 system, FDA 510K Number K950279.

## DEVICE DESCRIPTION:

UNIVISION™ Echocardiology System is a real-time network system that provides review of echocardiograms either inside or outside the hospital environment. The exams are then archived on optical disc media. The UNIVISION™ Echocardiology System is comprised of three major subsystems: the Image Acquisition Unit (IAU); the Workstation; and the Archive Subsystem.

- The Image Acquisition Unit (IAU) interfaces to a variety of ultrasound systems and obtains the patient’s information and image data during exams.

{1}

This information is stored locally on the IAU for access anywhere on the network.

- Workstations provide the review of either exams in progress or previous exams that may be stored anywhere on the network.
- The Archive Subsystem consists of a jukebox server and CD-R (Compact Disc - Recordable) for archive of the echocardiography exams.

All image transfers over the network and on-and-off the CD-R jukebox employ the MPEG-1 (ISO/IEC 11172) standard for video and audio compression at a 30:1 compression ratio.

The UNIVISION™ Echocardiology System is a network system that uses digital technology to acquire images from an echocardiography ultrasound system. The images may be viewed outside the exam room and are stored and retrieved on digital media. The substantially equivalent device is TomTec Imaging Systems', TomTec P90 system, FDA 510K Number K950279.

## COMPARISONS TO PREDICATE DEVICE:

The substantially equivalent device is TomTec Imaging Systems', TomTec P90 system, FDA 510K Number K950279. The TomTec P90 has several options. The TomTec P90 configuration that is equivalent to the UNIVISION™ Echocardiology System includes the Stress Echo and Networking application packages. The TomTec 3D/4D Ultrasound and Contrast Densitometry applications are not functions implemented in the UNIVISION™ Echocardiology System.

The UNIVISION™ Echocardiology System is intended for use in ultrasound Echocardiology laboratories. It acquires echocardiograms from ultrasound imaging systems during the exam and transmits these over a network for review and archive onto CD-R optical discs.

To increase speed of transmission over the network and make storage economically viable, MPEG-1 compression is employed. The images show no significant difference between the original image from an ultrasound echocardiography system and that from the same image after MPEG-1 compression has been employed.

## CONCLUSION:

The UNIVISION™ Echocardiology System is substantially equivalent to the legally marketed predicate device with regards to safety, effectiveness and intended use.

---

**Source:** [https://fda.innolitics.com/device/K964803](https://fda.innolitics.com/device/K964803)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
