OPTICAL TRACKING SYSTEM (OTS)

K964801 · Radionics Software Applications, Inc. · HAW · Jun 2, 1997 · Neurology

Device Facts

Record IDK964801
Device NameOPTICAL TRACKING SYSTEM (OTS)
ApplicantRadionics Software Applications, Inc.
Product CodeHAW · Neurology
Decision DateJun 2, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2

Indications for Use

The Optical Tracking System (OTS) is a graphical planning tool that allows for pre-operative and operative planning of surgical procedures. The OTS is indicated for use in surgical procedures in which anatomical landmarks are not clearly visible or where a desired target is close to critical structures.

Device Story

Optical Tracking System (OTS) is an image-guided surgical planning and guidance tool. System components include a camera array, computer workstation, application software, and surgical probes. During surgery, the surgeon uses a probe to select anatomical points; the camera array tracks probe position; the computer interprets these inputs and relates them to pre-operative images (CT/MR). The system provides interactive localization of targets. Used in clinical/surgical settings by surgeons to assist in navigating targets near critical structures or where landmarks are not visible. Benefits include improved surgical precision and target localization.

Clinical Evidence

Bench testing only. Accuracy verified using a phantom with targets of known position. Camera array accuracy reported at a mean value of 0.3 mm, maintained over extended operation. Verification confirmed correct encoding of stereotactic CT and MR localizer transformation equations.

Technological Characteristics

System comprises a camera array, computer workstation, application software, and probes. Operates as an image-guided localization system using optical tracking. Connectivity involves workstation-based processing of probe position data relative to patient images. No specific materials or software architecture class provided.

Indications for Use

Indicated for surgical procedures where anatomical landmarks are obscured or targets are near critical structures.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Submission Summary (Full Text)

{1} K964801 PAGE 1 OF 2 Optical Tracking System 510(k) Submission # Appendix 1 JUN - 3 1997 ## Summary of Safety and Effectiveness ### General Information Classification: Class II Common Name: Surgical Planning and Guidance System Device Trade Name: Optical Tracking System (OTS) Intended Uses: The Optical Tracking System (OTS) is a graphical planning tool that allows for pre-operative and operative planning of surgical procedures. The OTS is indicated for use in surgical procedures in which anatomical landmarks are not clearly visible or where a desired target is close to critical structures. Predicate Devices: Radionics Operating Arm System, K951262 and K961844; StealthStation Stereotactic System, manufactured by Surgical Navigation Technologies, Inc., K954276. Establishment Name and Address: Radionics Software Applications, Inc. 22 Terry Avenue Burlington, MA 01803 Contact Name and Phone: Amy J. LaForte, PhD, (617) 272-1233 Establishment registration number: 1222895 Performance Standard: None established under Section 514. ### Substantial Equivalence Determination A summary of the information contained in this premarket notification that addresses safety and effectiveness follows. ### Description of the Device and Basis for Substantial Equivalence The Optical Tracking System (OTS), addressed in this premarket notification, has the same intended use and similar technological characteristics as the commercially available Radionics Operating Arm System and the StealthStation Stereotactic System. The OTS is a modification of the Operating Arm System but substitutes a camera array for the operating arm to track the position of probes and various tools. This is the same camera {2} Page 2 of 2 # Optical Tracking System 510(k) Submission array utilized by StealthStation. The OTS provides an interactive, image-guided means of localizing targets in surgical procedures. In addition to the camera array, the OTS consists of a computer workstation, application software, and probes. Upon calibration of the system in the surgical environment, the surgeon selects points on the patient using a probe which are interpreted by the computer and related to corresponding points on the image. ## Safety Summary RSA Optical Tracking System (OTS) systems testing verifies that the stereotactic CT and MR localizer transformation equations are correctly encoded into the application software. Further, it verifies that all target coordinate input and displays are accurate. The results from the OTS are compared to a phantom with targets of known position. The camera array is accurate to a mean value of 0.3 mm. This accuracy is maintained over a period of extended operation. ## General Safety and Effectiveness Concerns The device labeling contains instructions for use. It includes indications for use, cautions, warnings and user quality assurance procedures. The training and installation sessions ensure that the user understands all aspects of the Optical Tracking System: hardware, computer, and software and its intended functionality. This information promotes safe and effective use of the device. {3} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN - 3 1997 Amy J. LaForte, Ph.D. Senior Regulatory Engineer Radionics Software Applications, Inc. P.O. Box 358 22 Terry Avenue Burlington, Massachusetts 01803-0658 Re: K964801 Trade Name: Optical Tracking System Regulatory Class: II Product Code: 84HAW Dated: March 3, 1997 Received: March 4, 1997 Dear Dr. LaForte: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4} Page 2 - Amy J. LaForte, Ph.D. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5} Page 1 of 1 510(k) Number (if known): K964801 Device Name: ________ Optical Tracking System 510(k) Submission ## Indications for Use The following is the indications for use of the OTS : The Optical Tracking System (OTS) is a graphical planning tool that allows for pre-operative and operative planning of surgical procedures. The OTS is indicated for use in surgical procedures in which anatomical landmarks are not clearly visible or where a desired target is close to critical structures. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ![img-0.jpeg](img-0.jpeg) (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number K964801 Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ________
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...