KINEMATIC POSITIONING DEVICES; C-SPINE, KNEE,SHOULDER
K964752 · Philips Medical Systems (Cleveland), Inc. · LNH · Feb 24, 1997 · Radiology
Device Facts
| Record ID | K964752 |
| Device Name | KINEMATIC POSITIONING DEVICES; C-SPINE, KNEE,SHOULDER |
| Applicant | Philips Medical Systems (Cleveland), Inc. |
| Product Code | LNH · Radiology |
| Decision Date | Feb 24, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1000 |
| Device Class | Class 2 |
Intended Use
The Picker International Outlook system is intended for use as a NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR, (2) depend upon NMR parameters (proton density, flow velocity, spin-lattice relaxation time T1, spin-spin relaxation time T2) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis. The Kinematic Positioning Devices enable the physician to evaluate the anatomy and the functional interactions of the surrounding tissues (ligaments, cartilage, bone, muscle and fat). The Kinematic Positioning Devices are indicated for use in the Cervical Spine, Knee and Shoulder.
Device Story
Kinematic positioning devices for Outlook MRI system; used to constrain patient anatomy (c-spine, knee, shoulder) in specific positions during imaging. C-spine device allows flexion, extension, lateral movement, and rotation. Knee device allows flexion and extension. Shoulder device allows abduction, adduction, and rotation. Devices include integrated holders for multipurpose receiver coils. Manually operated by clinician; no active components or motors. Enables physician to evaluate dynamic tissue interactions; aids in diagnostic interpretation of MRI images.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Non-ferromagnetic materials; manually operated mechanical positioning frames; integrated receiver coil holders; no active energy sources or motors; dimensions/form factor specific to anatomical region (c-spine, knee, shoulder).
Indications for Use
Indicated for patients requiring MRI imaging of the cervical spine, knee, or shoulder to evaluate anatomy and functional tissue interactions (ligaments, cartilage, bone, muscle, fat).
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
- Siemen's Kinematic Knee Device for the Magnetom Open System (K961121)
Related Devices
- K982761 — JOINT MOTION DEVICE PACKAGE PN 1032,KNEEL CERVICAL PN 1026, LUMBAR 1027, WRIST 1028, ANKLE 1029, SHOULDER 1030, HIP 1031 · Chamco, Inc. · Oct 22, 1998
- K972704 — KINEMATIC POSITIONERS · Ge Medical Systems, Inc. · Sep 29, 1997
- K961121 — KINEMATIC KNEE DEVICE/MAGNETOM OPEN SYSTEM · Siemens Medical Solutions USA, Inc. · Jun 19, 1996
- K972296 — MAGNETIC RESONANCE DIAGNOSTIC SYSTEM · Ge Medical Systems, Inc. · Sep 15, 1997
- K093738 — MRI PATIENT POSITIONING DEVICES · Medtec, Inc. · May 14, 2010
Submission Summary (Full Text)
{0}
216-473-5729
12:48
# FEB 24 1997
# SUMMARY OF SAFETY AND EFFECTIVENESS
(As required by 21 CFR 807.92)
## 1. General Information
K964752
Classification: Class II
Magnetic Resonance (MR) Imaging System
Common/Usual Name: Magnetic Resonance Imaging Accessories
Proprietary Name: Kinematic Positioning Devices: C-Spine, Knee and Shoulder
Establishment Registration
Manufacturer: Picker Nordstar, Inc.
P.O. Box 33
FIN-00511 Helsinki, FINLAND
Phone: +358-9-394 127
Fax: +358-9-146 2811
FDA Facility Registration #9680194
United States Representative: Picker International, Inc.
595 Miner Road
Cleveland, OH 44143
FDA Owner Number #1580240
Performance Standards: No applicable performance standards have been issued under section 514 of the Food, Drug and Cosmetic Act.
## 2. Intended Uses
The Picker International Outlook system is intended for use as a NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR, (2) depend upon NMR parameters (proton density, flow velocity, spin-lattice relaxation time T1, spin-spin relaxation time T2) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.
The Kinematic Positioning Devices enable the physician to evaluate the anatomy and the functional interactions of the surrounding tissues (ligaments, cartilage, bone, muscle and fat). The Kinematic Positioning Devices are indicated for use in the Cervical Spine, Knee and Shoulder.
PICKER INTERNATIONAL, INC.
(KINE)
2/19/97
{1}
12:48
3. Device Description
C-Spine Positioning Device
The Outlook C-spine positioning device is designed to keep the head of the patient constrained in different positions while the C-spine is being imaged in the Outlook scanner. The device allows flexion, extension, lateral movement and lateral rotation. There is also a possibility to set the distance between the horizontal (flexion/extension) axis and the vertical axis (lateral motion). The device has an integrated holder for the Multipurpose receiver coil.
Knee Positioning Device
The Outlook Knee positioning device is designed to keep the leg constrained in different positions while the knee joint is imaged in the Outlook system with a Multipurpose coil. The device is used for flexion and extension of the knee.
Shoulder Positioning Device
The Outlook Knee positioning device is designed to keep the arm of the patient constrained in different positions while the shoulder joint is being imaged in the Outlook scanner. The device allows abduction, adduction and internal/external rotation. The device has an integrated holder for the Multipurpose receiver coil.
4. Safety and Effectiveness
The Picker International Kinematic Positioning Devices for the Outlook are similar in technological characteristics and intended use to the Siemen's Kinematic Knee Device for the Magnetom Open System (K961121). The following chart has been created to demonstrate their substantial equivalence.
PICKER INTERNATIONAL, INC. (KINE) 2/19/97 I-2
{2}
216-473-5721 11111111111111111111111111111111111
10 97 12:48
# SUBSTANTIAL EQUIVALENCE CHART
| Parameter | Kinematic Positioning Devices for the Outlook | Predicate Device Kinematic Knee Device for the Magnetom Open System (K961121) |
| --- | --- | --- |
| Material | Same | No ferromagnetic materials which could affect the scan field. |
| Device operation | Same | Manually operated by the user. No active components such as motors. |
| Indications for Use | C-Spine, Knee, Shoulder scanning | Knee scanning |
| Intended use | The Siemens Magnetom Open is a magnetic resonance imaging system which uses time-varying magnetic field gradients and rf energy to spatially encode the anatomy of the patient. The kinematic knee device will enable the physician to evaluate the anatomy of the knee and dynamic interaction of the different tissues (ligaments, cartilage, bone, muscle, fat).
The Picker International Outlook system is intended for use as a NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR, (2) depend upon NMR parameters (proton density, flow velocity, spin-lattice relaxation time T1, spin-spin relaxation time T2) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.
The Kinematic Positioning Devices enable the physician to evaluate the anatomy and the functional interactions of the surrounding tissues (ligaments, cartilage, bone, muscle and fat). | |
PICKER INTERNATIONAL, INC.
(KINE)
2/19/97