← Product Code [GEI](/productcode/GEI) · K964736

# BERGEN 610 BIPOLAR COAGULATOR (K964736)

_Bergen Mfg. · GEI · Feb 27, 1997 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K964736

## Device Facts

- **Applicant:** Bergen Mfg.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Feb 27, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Device Story

BERGEN 610 Coagulator; bipolar electrosurgical device. Features 0-50 digital power setting display (+/- 5 watts at 100 ohms); bar graph output current monitor for energy activation status. Handswitch powered by independent 7 Vdc isolated supply. Used by clinicians for coagulation; provides visual/audio monitoring of output. Safety features include RF and low frequency leakage compliance with IEC 601-1-2 and ANSI HF-18 standards for isolated body-floating devices.

## Clinical Evidence

No clinical data provided; safety and effectiveness based on compliance with IEC 601-1-2 and ANSI HF-18 standards.

## Technological Characteristics

Bipolar coagulator; isolated (body floating) output; 7 Vdc handswitch power supply; digital power display; bar graph current monitor. Compliant with IEC 601-1-2 and ANSI HF-18 safety standards.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

U.S. Food and Drug Administration
510K Document Mail Center HFZ-401
Center for Medical Devices
390 Piccard Drive
Rockville, Maryland 20850

November 15, 1996

To: Document Control Clerk

This summary of 510K safety and effectiveness for the BERGEN 610 Coagulator is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807,92.

A 0-50 digital display power setting indicator provides the operator a parallax free reading of the power setting (+/- 5watts, full scale @ 100 ohms).

A bar graph display output current monitor provides the operator a relative measure of the output current amplitude and to denote the presence of output energy during activation.

The handswitch is powered by an independent low voltage (7 Vdc) isolated power supply.

Audio and Visual monitors are in accord with IEC 601-1-2 and ANSI HF-18 guidelines.

RF and Low Frequency leakages are well within IEC 601-1-2 and ANSI HF-18 safety guide lines lines for isolated (body floating) bipolar coagulators.

This device is substantially equivalent to existing approved devices.

Sincerely,

![img-0.jpeg](img-0.jpeg)

Roger Oosten
Bergen Mfg
9345 Rookery Road
New Port Richey, Florida 34654

Date 11/15/96

---

**Source:** [https://fda.innolitics.com/device/K964736](https://fda.innolitics.com/device/K964736)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
