STA-LIATEST D-DI TEST KIT

K964728 · American Bioproducts Co. · DAP · Jun 26, 1997 · Hematology

Device Facts

Record IDK964728
Device NameSTA-LIATEST D-DI TEST KIT
ApplicantAmerican Bioproducts Co.
Product CodeDAP · Hematology
Decision DateJun 26, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7320
Device ClassClass 2

Intended Use

The STA®-Liatest® D-Di test kit is intended for use with STA® analyzers (Diagnostica Stago, France: STA® full-size model, K942117; STA® Compact model, K961579) to perform quantitative determinations of D-dimer antigen levels in citrated plasma by the immuno-turbidimetric method based on the measurement of light absorbance produced by a suspension of microlatex particles coated with mouse anti-human D-dimer monoclonal antibodies.

Device Story

In vitro diagnostic test kit; uses citrated plasma samples. Principle: immuno-turbidimetric assay; microlatex particles coated with mouse anti-human D-dimer monoclonal antibodies. Input: plasma sample; light absorbance measured by STA® analyzer. Output: quantitative D-dimer antigen levels. Used in clinical laboratory settings; operated by laboratory personnel. Provides quantitative data to assist in clinical assessment of D-dimer levels.

Technological Characteristics

Immuno-turbidimetric assay; reagents include Tris buffer and microlatex particles coated with mouse anti-human D-dimer monoclonal antibodies. Designed for use on STA® analyzer platforms. Reagent stability: 18 months at 2-8°C; 15 days on-board stability.

Indications for Use

Indicated for the quantitative determination of D-dimer antigen levels in citrated plasma using STA® analyzers.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} K964728 AMERICAN BIOPRODUCTS COMPANY Premarket 510(k) Notification STA®-Liatest® D-Di Test Kit JUN 26 1997 Page 33 of 33 ## VIII. Safety and Effectiveness Summary The STA®-Liatest® D-Di test kit is intended for use with STA® analyzers to perform quantitative assays of D-dimer antigen in citrated plasma by the immuno-turbidimetric method based on the measurement of light absorbance produced by a suspension of microlatex particles coated with specific mouse anti-human D-dimer monoclonal antibodies. Each STA®-Liatest® D-Di test kit provides: 6 x 5-ml vials of ready-for-use Tris buffer and 6 x 6-ml vials of a suspension of microlatex particles coated with mouse anti-human D-dimer monoclonal antibodies. Reagents in intact vials remain stable for 18 months after the date of manufacture, when stored at 2°-8°C. After opening of the vials, kit reagents (with STA®-Reducers and perforated caps installed on the vials) are stable for 15 days on board STA® analyzers. The proposed STA®-Liatest® D-Di test kit has demonstrated substantial equivalence to the commercially available ASSERACHROM® D-Di kit (Diagnostica Stago, France; K862156) which is a microtiter ELISA-based procedure for the quantitative determination of D-dimer antigen levels in citrated plasma. {1} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JUN 26 1997 Andrew Loc B. Le., Ph.D. Director, Regulatory Affairs and Quality Assurance American BioProducts Company Five Century Drive Parsippany, New Jersey 07054 Re: K964728 STA®-Liatest® D-Di Test Kit Regulatory Class: II Product Code: DAP Dated: April 16, 1997 Received: April 18, 1997 Dear Dr. Le: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. {2} Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3} Page 1 of 1 510(k) Number (if known): _________________________ Device Name: STA®-Liatest® D-Di Test Kit Indications for Use: The STA®-Liatest® D-Di test kit is intended for use with STA® analyzers (Diagnostica Stago, France: STA® full-size model, K942117; STA® Compact model, K961579) to perform quantitative determinations of D-dimer antigen levels in citrated plasma by the immuno-turbidimetric method based on the measurement of light absorbance produced by a suspension of microlatex particles coated with mouse anti-human D-dimer monoclonal antibodies. ![img-0.jpeg](img-0.jpeg) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) _________________________ _________________________ _________________________ (Concurrence of CDRH, Office of Device Evaluation (ODE)) Prescription Use ☐ OR Over-The-Counter Use ☐ (Per 21 CFR 801.109) (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...