K964723 · Racal Filter Technologies, Inc. · FXX · Feb 7, 1997 · General, Plastic Surgery
Device Facts
Record ID
K964723
Device Name
RACAL N95 RESIRATOR
Applicant
Racal Filter Technologies, Inc.
Product Code
FXX · General, Plastic Surgery
Decision Date
Feb 7, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Intended Use
The Racal N95 Respirator and Surgical Mask will be used for the protection of the health care practitioner from blood and body fluids and particulates and to meet the NIOSH / CDC TB Prevention Guidelines.
Device Story
Disposable, lightweight, one-piece half-mask respirator and surgical mask. Constructed with multi-layer filtering media, including inner/outer scrim layers, netting, and a meltblown protection layer. Features a wire nose clip, foam nose piece, and elastic headbands to ensure a secure face-to-mask seal. Used by healthcare practitioners in clinical settings to provide a fluid-resistant barrier and 95% filtration efficiency against non-oil-based particulate aerosols. Protects the wearer from exposure to blood, body fluids, and airborne pathogens, specifically aiding in the prevention of Mycobacterium tuberculosis transmission.
Clinical Evidence
Bench testing only. Device is NIOSH approved (TC-84A-0008) with 95% filtration efficiency against particulate aerosols free of oil, meeting 42 CFR Part 84 standards and CDC guidelines for Mycobacterium tuberculosis transmission prevention.
Technological Characteristics
N95-class particulate respirator. Materials: multi-layer filtering media (inner/outer scrim, netting, meltblown layer), wire nose clip, foam nose piece, elastic headband. Dimensions: four sizes (XS, S, M, L). Performance: 95% filtration efficiency against non-oil aerosols. NIOSH approved (TC-84A-0008).
Indications for Use
Indicated for health care practitioners to provide respiratory protection against particulates, blood, and body fluids, and to assist in compliance with NIOSH/CDC tuberculosis prevention guidelines.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Louis M Gerson Isolair APR Type N95 Respirators (K960778)
Related Devices
K964722 — DELTA N95 RESPIRATOR · Racal Filter Technologies, Inc. · Feb 7, 1997
K974068 — PFR95 PARTICULATE FILTER RESPIRATOR AND SURGICAL MASK REGULAR SIZE, PFR95 PARTICULATE FILTER RESPIRATOR AND SURGICAL · Tecnol Medical Products, Inc. · Dec 22, 1997
K062663 — MOLDEX HEALTH CARE N95 PARTICULATE RESPIRATORS AND SURGICAL MASKS · Moldex/Metric, Inc. · Oct 11, 2006
K061859 — INOVEL HEALTH CARE N95 PARTICULATE RESPIRATORS & SURGICAL MASKS · Inovel, LLC · Jul 12, 2006
K970448 — PCM 2000; ISOLAIR APR MODEL 2735 · Tecnol Medical Products, Inc. · Apr 30, 1997
Submission Summary (Full Text)
{0}
FILE.No. 708 01/29 '97 16:06 ID:RACAL HEALTH & SAFETY 301 695 3017 PAGE 2
RACAL HEALTH & SAFETY INC. PREMARKET SUBMISSION
RACAL N95 RESPIRATOR AND SURGICAL MASK 510(K) K964723
510(k) SUMMARY - K96 4723
FEB - 7 1997
RACAL N95 RESPIRATOR & SURGICAL MASK
January 29, 1997
This summary is provided in accordance with the Safe Medical Devices Act of 1990 (SMDA). The information provided in the 510(k), Premarket Notification, was in accordance with 21 CFR 807.87.
1. Submitter of 510(k):
Company Name: Racal Health & Safety Inc.
7305 Executive Way
Frederick, Maryland
U.S.A. 21704-8368
Contact Person (U.S.A.): Mr. J. Randall Hoggle
Telephone: (301)695-8200
Fax: (301)695-4413
Contact Person (Canada): John Campbell
Telephone: (613)345-0111
Fax: (613)345-2639
2. Name of Device
Trade/Proprietary Name: RACAL N95 RESPIRATOR AND SURGICAL MASK
Common/Usual Name: Masks, Surgical
Classification Name: 21 CFR 878.4040 "Surgical Apparel", Class II
3. Legally Marketed Predicate Devices
RACAL N95 RESPIRATOR AND SURGICAL MASK is substantially equivalent to many legally marketed Surgical masks, including those listed in Table I below.
Page 1 of 3
CONFIDENTIAL
{1}
FILE No. 708 01/29 '97 16:06 ID:RACAL HEALTH & SAFETY 301 695 3017 PAGE 3
RACAL HEALTH & SAFETY INC. PREMARKET SUBMISSION
RACAL N95 RESPIRATOR AND SURGICAL MASK 510(K) K964723
# 510(k) SUMMARY
TABLE 1 COMPARISON OF TECHNOLOGICAL CHARACTERISTICS PREDICATE DEVICES AND DELTA N95 RESPIRATOR AND SURGICAL MASK
| Manufacturers | Racal Health & Safety | 3M | Louis M Gerson |
| --- | --- | --- | --- |
| Device | RACAL N95 Respirators | Model 1860 Health Care Respirators | Isolair APR Type N95 Respirators |
| 510(k) Number | K964723 | K955382 | K960778 |
| Standards Met | Performance Code of Federal Regulations Title 42 Part 84 for N95-class of Respiratory Protective Devices
Meets CDC guidelines for preventing transmission of Mycobacterium tuberculosis | Performance Code of Federal Regulations Title 42,Part 84 for N95-class of Respiratory Protective Devices
Meets CDC guidelines for preventing transmission of Mycobacterium tuberculosis | Performance Code of Federal Regulations Title 42, Part 84 for N95 - class of Respiratory Protective Devices
Meets CDC guidelines for preventing transmission of Mycobacterium tuberculosis |
| NIOSH Approval Number | TC-84A-008 | TC-84A-006 | TC-84A-160 |
| Device Description | • N95 Class particulate Respirator
• Multi-Layer Filtering media
• Fluid resistant barrier
• Wire nose clip
• Foam nose piece inside mask
• Elastic Headband
• Peach colored | • N95 Class particulate Respirator
• Multi-Layer Filtering media
• Fluid resistant barrier
• Wire nose clip
• Foam nose piece inside mask
• Elastic Headband
• Blue colored | • N95 Class particulate Respirator
• Multi-Layer Filtering media
• Fluid resistant barrier
• Wire nose clip
• Foam nose piece inside mask
• Elastic Headband
• Blue-green colored |
| Intended Use | Respiratory Protection of Health Care workers | Respiratory Protection of Health Care of workers | Respiratory Protection of Health Care workers |
## 4. Device Description
RACAL N95 RESPIRATOR AND SURGICAL MASK is indicated for respiratory protective device. This disposable Surgical mask is a one-piece, light weight half mask with filter material forming the mask. The component includes an inner scrim layer, netting material forming the mask. The component includes an inner scrim layer, netting layer, meltblown protection layer, outer scrim layer and elastic band. RACAL
Page 2 of 3
CONFIDENTIAL
{2}
FILE No. 708 01/29 '97 16:06 ID:RACAL HEALTH & SAFETY 301 695 3017 PAGE 4
RACAL HEALTH & SAFETY INC. PREMARKET SUBMISSION
RACAL N95 RESPIRATOR AND SURGICAL MASK 510(K) K964723
N95 RESPIRATOR AND SURGICAL MASK is NIOSH approved (Approval #TC-84A-0008) and has a filtration efficiency level of 95% and is effective against particulate aerosols free of oil. RACAL N95 RESPIRATOR AND SURGICAL MASK is available in four sizes, extra small, small, medium and large. The nose clip and elastic headband ensures effective face to mask seal. Based on the information provided, RACAL N95 RESPIRATOR AND SURGICAL MASK is substantially equivalent to legally marketed predicate devices.
5. Indications For Use
The Racal N95 Respirator and Surgical Mask will be used for the protection of the health care practitioner from blood and body fluids and particulates and to meet the NIOSH / CDC TB Prevention Guidelines.
Page 3 of 3
CONFIDENTIAL
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.