← Product Code [CEP](/productcode/CEP) · K964694

# ELECSYS LH ASSAY (K964694)

_Boehringer Mannheim Corp. · CEP · Jan 24, 1997 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K964694

## Device Facts

- **Applicant:** Boehringer Mannheim Corp.
- **Product Code:** [CEP](/productcode/CEP.md)
- **Decision Date:** Jan 24, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1485
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

Immunoassay for the in vitro quantitative determination of human luteinizing hormone in human serum and plasma.

## Device Story

Elecsys® LH Assay is an in vitro diagnostic immunoassay for quantitative determination of human luteinizing hormone (LH) in serum and plasma. Device utilizes sandwich principle: 20 µL sample incubated with biotinylated monoclonal LH-specific antibody and ruthenium-labeled monoclonal LH-specific antibody; complex binds to streptavidin-coated microparticles. Reaction mixture aspirated into measuring cell; microparticles magnetically captured on electrode surface. Voltage application induces chemiluminescent emission measured by photomultiplier. Results calculated via instrument-specific calibration curve generated by 2-point calibration and master curve from reagent bar code. Used on Elecsys® 2010 instrument. Provides clinicians with quantitative LH levels to assist in endocrine assessment.

## Clinical Evidence

Bench testing only. Method comparison study (N=166) vs. Enzymun-Test® LH showed high correlation (r=0.993). Precision studies (N=60 per level) demonstrated %CVs ranging from 0.81% to 5.17%. Lower detection limit 0.10 mIU/mL. Linearity 0.1–200 mIU/mL. Interference testing performed for bilirubin, hemoglobin, lipemia, biotin, and rheumatoid factor. Specificity testing showed <0.1% cross-reactivity with FSH, HCG, TSH, HGH, and HPL.

## Technological Characteristics

Sandwich immunoassay; electrochemiluminescence detection; streptavidin-coated microparticles; biotinylated monoclonal antibodies; ruthenium complex label. Requires Elecsys® 2010 instrument. Calibration via 2-point method and reagent bar code. Quantitative measurement range 0.1–200 mIU/mL.

## Regulatory Identification

A luteinizing hormone test system is a device intended to measure luteinizing hormone in serum and urine. Luteinizing hormone measurements are used in the diagnosis and treatment of gonadal dysfunction.

## Predicate Devices

- Enzymum® LH Assay ([K900799](/device/K900799.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

JAN 24 1997
K964694

BOEHRINGER MANNHEIM CORPORATION

# 510(k) Summary

## Introduction

According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

## 1. Submitter name, address, contact

Boehringer Mannheim Corporation
2400 Bisso Lane
P.O. Box 4117
Concord, CA 94524-4117
(510) 674 - 0690 extension 8240
FAX 510 687 - 1850

Contact Person: Betsy Soares-Maddox
Date Prepared: November 19, 1996

## 2. Device name

Proprietary name: Elecsys® LH Assay

Common name: Electrochemiluminescence assay for the determination of human luteinizing hormone (LH).

Classification name: System, Test, Human Luteinizing Hormone

## 3. Predicate device

We claim substantial equivalence to the Enzymum® LH Assay (K900799).

## 4. Device Description

Sandwich principle. Total duration of assay: 18 minutes.

- 1st incubation (9min.): 20 µL of sample, a biotinylated monoclonal LH-specific antibody (75 µL) and a monoclonal LH-specific antibody labeled with a ruthenium complex (75 µL)** react to form a sandwich complex.
- 2nd incubation (9min.): after addition of streptavidin-coated microparticles (30 µL), the complex becomes bound to the solid phase via interaction of biotin and streptavidin.

**Tris(2,2'-bipyridyl)ruthenium(II) complex (Ru(bpy)₂⁺)**

Continued on next page
page 26

{1}

page 27
# 510(k) Summary, Continued

## 4. Device Description

- The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier (0.4 second read frame).
- Results are determined via a calibration curve which is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent bar code.

## 5. Intended use

Immunoassay for the in vitro quantitative determination of human luteinizing hormone in human serum and plasma.

## 6. Comparison to predicate device

The Boehringer Mannheim Elecsys® LH Assay is substantially equivalent to other products in commercial distribution intended for similar use. Most notably it is substantially equivalent to the currently marketed Enzymun® LH Assay (K900779).

The following table compares the Elecsys® LH Assay with the predicate device, Enzymun® LH Assay. Specific data on the performance of the test have been incorporated into the draft labeling in attachment 5. Labeling for the predicate device is provided in attachment 6.

## Similarities:

- Intended Use: Immunoassay for the in vitro quantitative determination of human luteinizing hormone (LH)
- Sample type: Serum and plasma
- Antibody: Same pair of monoclonal mouse anti-LH antibodies
- Solid phase binding principle: Streptavidin/Biotin

Continued on next page

{2}

510(k) Summary, Continued

6. Comparison to predicate device cont.

Differences:

|  Feature | Elecsys® LH | Enzymun-Test® LH  |
| --- | --- | --- |
|  Assay Standardization | Enzmun® LH | WHO # 80/552  |
|  Detection method | Electrochemiluminescence | ELISA/1-step sandwich assay using streptavidin technology  |
|  Instrument required | Elecsys® 2010 | ES 300  |
|  Calibration Stability | A calibration is recommended every 7 days if kits is not consumed; 4 weeks with same reagent lot if reagent is consumed within 7 days. | Full calibration required every 2 weeks. One-point calibration required every run.  |

Performance Characteristics:

|  Feature | Elecsys® LH |   |   | Enzymun-Test® LH  |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|  Precision | Modified NCCLS (mIU/mL): |   |   | Modified NCCLS (mIU/mL):  |   |   |
|  Level | Low | Mid | High | Low | Mid | High  |
|  N | 60 | 60 | 60 | 120 | 120 | 120  |
|  Within-Run Mean | 0.54 | 9.38 | 50.72 | 3.6 | 13.5 | 59.5  |
|  %CV | 1.82 | 1.13 | 0.81 | 2.9 | 3.8 | 1.5  |
|  Total Mean | 0.54 | 9.38 | 50.72 | 3.6 | 13.5 | 59.5  |
|  %CV | 5.17 | 1.97 | 1.99 | 4.4 | 4.7 | 3.9  |
|  Lower Detection Limit | 0.10 mIU/mL |   |   | 0.50 mIU/mL  |   |   |

Continued on next page

{3}

510(k) Summary, Continued

6. Comparison to predicate device, (cont.)

Performance Characteristics:

|  Feature | Elecsys® LH | Enzymun-Test® LH  |
| --- | --- | --- |
|  Linearity | 0.1 - 200 mIU/mL (with a deviation from a linear line of ±10%) | 0.5 - 150 mIU/mL (with a deviation from a linear line of ±10%)  |
|  Method Comparison | Vs Enzymun-Test® LH
Least Squares
y = 1.00x - 0.199
r = 0.993
SEE = 1.141
N = 166

Passing/Bablok
y = 0.964x + 0.040
r = 0.993
SEE = 0.456
N = 166 | Vs Enzymun-Test® LH
Least Squares
y = 0.93x + 0.42
r = 0.953
SEE = 5.079
N = 62  |
|  Interfering substances | No interference at: | No interference at:  |
|  Bilirubin | 25.0 mg/dL | 64.5 mg/dL  |
|  Hemoglobin | 1 g/dL | 1 g/dL  |
|  Lipemia | 1500 mg/dL | 1250 mg/dL  |
|  Biotin | 30 ng/mL | 40 ng/mL  |
|  Rheumatoid Factor | 1500 U/mL | no interference  |
|  Specificity | Level tested % Cross-reactivity | Level tested % Cross-reactivity  |
|  FSH | 300 mIU/mL < 0.1 | 200 mU/mL 0.00  |
|  HCG | 600 IU/mL < 0.1 | 200 U/mL 0.00  |
|  TSH | 300 μIU/mL < 0.1 | 100 μU/mL 0.00  |
|  HGH | 600 μIU/mL < 0.1 | 200 ng/mL 0.00  |
|  HPL | 13.80 pmol/mL < 0.1 | ---  |

---

**Source:** [https://fda.innolitics.com/device/K964694](https://fda.innolitics.com/device/K964694)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
