K964624 · Siemens Medical Solutions USA, Inc. · DSI · Sep 11, 1997 · Cardiovascular
Device Facts
Record ID
K964624
Device Name
SIEMENS SC9000/SC9015 BEDSIDE MONITORING SYSTEM
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
DSI · Cardiovascular
Decision Date
Sep 11, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The intended use of this device is to measure heart rate, respiration rate, invasive pressure, non-invasive pressure, arrhythmia (adult), temperature, cardiac output, arterial oxygen saturation, pulse rate, cardiac output, end-tidal carbon dioxide and (central) apnea. This device will produce visual and aural alarms if any of these parameters vary beyond preset limits and produce timed or alarm recordings. This device will connect to the Siemens SIRENET or Infinity(Olympus)network. The SC9000/SC9015 Modified Patient Monitoring System is intended to be used on Adult, Pediatric and Neonatal populations, with the exception of the parameter Cardiac Output, which is intended for use in the adult and pediatric populations only, and Arrhythmia, which is intended for use in the adult population only.
Device Story
Bedside monitoring system; processes physiological signals including ECG, pressure, temperature, SpO2, and CO2. Modification involves updated software (VB1.1-W) to improve arrhythmia detection via increased sampling rate (250 samples/sec vs 100) and 2-lead processing. System provides visual/aural alarms for parameter limit violations; generates recordings. Used in clinical settings by healthcare professionals to support patient assessment and decision-making. Connects to Siemens SIRENET/Infinity networks and third-party devices (SV300 ventilator, Baxter Vigilance monitor). Benefits include enhanced arrhythmia monitoring accuracy for adult patients.
Clinical Evidence
No clinical trial data provided. Performance equivalence supported by bench testing against AAMI standards (ECAR-D-94, EC13-1992) and FDA guidelines for arrhythmia detectors. Clinical performance assessment mentioned as attached in appendix V, claiming equivalence or superiority to the predicate.
Indicated for monitoring heart rate, respiration, invasive/non-invasive pressure, arrhythmia (adults only), temperature, cardiac output (adults/pediatrics only), SpO2, pulse rate, and end-tidal CO2/apnea in adult, pediatric, and neonatal populations. Used by healthcare professionals (physicians, nurses, technicians) in clinical care environments.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}
510 (K) Notification
Siemens SC9000/ SC9015 Bedside Monitoring System with modified Arrhythmia Monitoring
K964624
SEP 11 1997
**510(k) SUMMARY**
as required per 807.92(c)
2: Submitter’s Name, Address:
Siemens Medical Systems, Inc.
16 Electronics Avenue
Danvers, MA 01923
Tel: (508) 750-7500
Fax: (508) 777-3398
Official Correspondent: Robert W. Becker
Contact person for this submission: Jacqueline E. M. Emery or Fred Geheb
Date submission was prepared: October 24, 1996
3: Trade Name, Common Name and Classification Name:
A. Trade Name: Siemens SC9000/ SC9015 Bedside Monitoring System
B. Common Name, Classification Number, Class and Regulation Number:
| Common Name | Classification Number | Class | Regulation Number |
| --- | --- | --- | --- |
| Cardiac monitor | 74DRT | II | 21 CFR 870.2300 |
| Arrhythmia detector & Alarm System | 74DSI | III | 21 CFR 870.1025 |
| Breathing frequency monitor | 73BZQ | II | 21 CFR 868.2375 |
| Pulse rate monitor | 74BWS | II | 21 CFR 870.2300 |
| Non-indwelling blood pressure monitor | 74DXN | II | 21 CFR 870.1130 |
| Clinical electronic thermometer | 80BWX | II | 21 CFR 880.2910 |
| Pulse Oximeter | 74DQA | II | 21 CFR 870.2700 |
| Cardiac Output Monitor | 74KFN | II | 21 CFR 870.1435 |
| end-tidal Carbon-Dioxide Monitor | 73CCK | II | 21 CFR 868.1400 |
| Indwelling Blood Pressure Monitor | 74CAA | II | 21 CFR 870.1110 |
| Indwelling blood pressure monitor | 74CAA | II | 21 CFR 870.1110 |
| Heart Rate Monitor, Neonatal * | 74FLO | II | 21 CFR 870.2300 |
| Ventilatory Effort Monitor * (Apnea Detector) | 73FLS | II | 21 CFR 868.2375 |
| Monitor Blood Pressure, Neonatal, * Invasive | 74FLP | II | 21 CFR 870.1110 |
* Pending FDA clearance of 510(K) K962291
Siemens Medical Systems, Inc.
Electromedical Group
16 Electronics Avenue
Danvers, MA 01923
USA
Tel: (508) 750-7500
Fax: (508) 777-3398
Telex: 511958 (Siemensm SD)
{1}
510 (K) Notification
Siemens SC9000/ SC9015 Bedside Monitoring System with modified Arrhythmia Monitoring
## 4: Predicate Device Identification:
The Siemens SC9000/ SC9015 Bedside Monitoring System with modified Arrhythmia Monitoring is an updated software version of the Siemens SC9000/ SC9015 Bedside Monitoring System. The hardware of the SC9000/SC9015 is unchanged. The modification improves the detection and monitoring of arrhythmia. The Siemens SC9000/SC9015 Bedside Monitoring Systems was granted 510(K) clearance under the following 510(K) numbers:
K946306 - Siemens SC9000/SC9015 Monitor (Original Submission)
K954632 - Siemens SC9000/SC9015 Monitor with etCO₂ Functionality
The following submission is currently awaiting FDA 510(K) clearance:
K962291 - Siemens SC9000/SC9015 Monitor modified with Neonatal Functionality.
## 5. Device Description
The Siemens SC9000/ SC9015 Bedside Monitoring System with modified Arrhythmia Monitoring is an updated software version of the Siemens SC9000/ SC9015 Bedside Monitoring System. The modification improves the detection and monitoring of arrhythmia. The hardware of the SC9000/SC9015 is unchanged.
## 6. Intended Use:
The intended use of this device is to measure heart rate, respiration rate, invasive pressure, non-invasive pressure, arrhythmia (adult), temperature, cardiac output, arterial oxygen saturation, pulse rate, cardiac output, end-tidal carbon dioxide and (central) apnea. This device will produce visual and aural alarms if any of these parameters vary beyond preset limits and produce timed or alarm recordings. This device will connect to the Siemens SIRENET or Infinity(Olympus)network.
The SC9000/SC9015 Modified Patient Monitoring System is intended to be used on Adult, Pediatric and Neonatal populations, with the exception of the parameter Cardiac Output, which is intended for use in the adult and pediatric populations only, and Arrhythmia, which is intended for use in the adult population only.
Siemens Medical Systems, Inc.
Electromedical Group
16 Electronics Avenue
Danvers, MA 01923
USA
Tel: (508) 750-7500
Fax: (508) 777-3398
Telex: 511958 (Siemensm SD)
{2}
510 (K) Notification
Siemens SC9000/ SC9015 Bedside Monitoring System with modified Arrhythmia Monitoring
7. Table of device similarities and differences to predicate device
| | Modified Algorithm | Predicate Device Algorithm | Explanation of the modified version |
| --- | --- | --- | --- |
| Manufacturer | Siemens Medical Systems EMG | same | |
| Device name | SC9000 with modified arrhythmia algorithm | SC9000 | 2 leads improve arrhythmia detection |
| 510(K) Number | To be assigned | K946306 | |
| Intended Population | Adult, pediatric and neonatal * | same | Same-application areas. Arrhythmia detection is inactive in the neonatal mode. |
| Software Revision | VB1.1-W | VA2-W | |
| Display: | up to 2 leads | same | |
| Available leads: | I, II, III, aVR, aVF, aVL, V, MCL1, MCL6 (MCL1, MCL6 only with 3-lead set; V, aVR, aVF, aVL only with 5 lead set) | same | |
| Measuring range: | 15 - 300 bpm | same | |
| Accuracy: | ±5 bpm /min. or ± 5% (whichever is greater) | same | |
| QRS detection: amplitude: duration: | 0.5 - 5.0 mV
40-120 msec | same
70 - 120 msec | extended to 40 for neonatal. |
| Frequency ranges: filter setting = Monitor: filter setting = ESU: filter setting = Off: | 0.5 to 40 Hz
0.5 to 20 Hz
0.05 to 40 Hz | same
same
same | |
| Degree of protection against electrical shock | Type CF | same | |
| Defibrillation protection | In accordance with IEC 6-1-2-27 | same | |
| Arrhythmia Detection: | Yes | same | |
| Pacer Detection: | Yes | same | |
| Samples/Sec | 250 Samples/Second | 100 Samples/Sec | Increased sample rate for improved QRS processing. |
* with the exception of the parameter Cardiac Output, which is intended for use in the adult and pediatric populations only, and Arrhythmia, which is intended for use in the adult population only.
Siemens Medical Systems, Inc.
Electromedical Group
16 Electronics Avenue
Danvers, MA 01923
USA
Tel: (508) 750-7500
Fax: (508) 777-3398
Telex: 511958 (Siemensm SD)
{3}
510 (K) Notification
Siemens SC9000/ SC9015 Bedside Monitoring System with modified Arrhythmia Monitoring
8. Assessment of non-clinical performance data for equivalence:
Currently there are no FDA standards for this device. However, the Siemens SC9000/SC9015 Monitor with modified arrhythmia detection complies with:
AAMI (ECAR - D-94) “Draft: Recommended Practice for Testing and Reporting Performance Results of Ventricular Arrhythmia Detection Algorithms” March 1994.
AAMI (EC13 - 1992) “Cardiac monitors, heart rate meters, and alarms”.
FDA “Guidelines for Submitting Data in Support of Pre-market Notification (510(K)) Applications for Arrhythmia Detectors (1990)”.
9. Assessment of clinical performance data for equivalence:
Clinical performance is equal to, or surpasses that of the predicate device. The assessment of clinical performance is attached in appendix V.
10. Bio-compatibility Data:
Not applicable
11. Sterilization data:
Not applicable
Siemens Medical Systems, Inc.
Electromedical Group
16 Electronics Avenue
Danvers, MA 01923
USA
Tel: (508) 750-7500
Fax: (508) 777-3398
Telex: 511958 (Siemensm SD)
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20856
SEP 11 1997
Ms. Jacqueline E.M. Emery
Siemens Medical Systems, Inc.
16 Electronics Avenue
Danvers, Massachusetts 01923
Re: K964624
Siemens SC9000/SC9015 Bedside Monitoring System with Modified Arrhythmia Monitoring
Regulatory Class: III (three)
Product Code: 74 DSI
Dated: March 8, 1997
Received: June 10, 1997
Dear Ms. Emery:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}
Page 2 - Ms. Jacqueline E.M. Emery
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular,
Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{6}
Page 1 of 1
510(k) Number (if known): K964624
Device Name: Siemens SC9000/SC9015 Bedside Monitoring System with Modified Arrhythmia Monitoring
Indications for Use:
This device is capable of monitoring:
- heart rate
- respiration rate
- invasive pressure
- non-invasive pressure
- arrhythmia
- temperature
- cardiac output
- arterial oxygen saturation
- pulse rate
- cardiac output
- end-tidal carbon dioxide and (central) apnea
This device can be connected to third party devices, Siemens SV300™ ventilator and the Baxter Vigilance™ blood gas/continuous cardiac output monitor. The device is intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition. The device is intended to be used on Adult, Pediatric and Neonatal populations, with the exception of the parameter Cardiac Output which is intended for use in the adult and pediatric populations only; and Arrhythmia which is intended for use in the adult population only.
## MRI Compatibility Statement:
The Siemens SC9000/SC9015 Series is not compatible for use in an MRI magnetic field.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number
703170
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.