PRIME & BOND 2.1 M.P. DENTIN/ENAMEL BONDING AGENT W/ACTIVATOR

K964525 · Dentsply Intl. · KLE · Jan 22, 1997 · Dental

Device Facts

Record IDK964525
Device NamePRIME & BOND 2.1 M.P. DENTIN/ENAMEL BONDING AGENT W/ACTIVATOR
ApplicantDentsply Intl.
Product CodeKLE · Dental
Decision DateJan 22, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3200
Device ClassClass 2

Indications for Use

Prime & Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent is a one-component, no-mix visible light curable dental bonding agent that contains fluoride. A new accessory, Prime & Bond™ Self-Curing Activator, can be used with Prime & Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent, enabling the practitioner to use self-cure bonding techniques for the current Indications for Use of the Bonding Agent where a dual-cured adhesive may be desired. Also because of the self-cure capability, use of the Bonding Agent and Activator together allows adhesion of amalgam restorations to tooth structures, which is a NEW INDICATION FOR USE.

Device Story

One-component, no-mix visible light curable dental bonding agent containing fluoride; used with new Prime & Bond™ Self-Curing Activator (acetone/ethanol solution of aromatic sulfinic acid salt). Activator initiates self-cure free radical polymerization of monomers/oligomers in bonding agent. Used by dental practitioners in clinical settings. Enables dual-cured adhesive techniques and bonding of amalgam restorations to tooth structures. Benefits include expanded clinical utility for self-cure applications and improved restorative adhesion.

Clinical Evidence

Bench testing only; includes performance data for the bonding agent/activator combination and cytotoxicity test results for the activator's active ingredient.

Technological Characteristics

One-component, no-mix visible light curable resin bonding agent; contains fluoride. Activator is an acetone/ethanol solution of an aromatic sulfinic acid salt. Initiates free radical polymerization.

Indications for Use

Indicated for dental bonding of restorative materials to dentin and enamel, including dual-cured adhesive applications and adhesion of amalgam restorations to tooth structures.

Regulatory Classification

Identification

A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.

Predicate Devices

Submission Summary (Full Text)

{0} JAN 22 1997 510(k) SUMMARY K964525 Page 1 of 2 DENTSPLY NAME & ADDRESS: DENTSPLY International 570 West College Avenue P.O. Box 872 York, PA 17405-0872 (717) 845-7511 Fax (717) 854-2343 P. J. Lehn Telefax (717) 849-6343 CONTACT: P. Jeffery Lehn DATE PREPARED: November 11, 1996 TRADE NAME: PRIME & BOND™ 2.1 MULTIPURPOSE DENTIN/ENAMEL BONDING AGENT WITH ACTIVATOR COMMON OR USUAL NAME: Bonding agent CLASSIFICATION NAME: Resin tooth bonding agent 21 CFR, 872.3200 PREDICATE DEVICES: Scotchbond M.P. Plus Dental Adhesive System K942493 Prime & Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent K962348 DESCRIPTION AND INTENDED USE: Prime & Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent is a one-component, no-mix visible light curable dental bonding agent that contains fluoride. No changes have been made to Prime & Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent in design, material, chemical composition or manufacturing process. A new accessory, Prime & Bond™ Self-Curing Activator, can be used with Prime & Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent, enabling the practitioner to use self-cure bonding techniques for the current Indications for Use of the Bonding Agent where a dual-cured adhesive may be desired. Also because of the self-cure capability, use of the Bonding Agent and Activator together allows adhesion of amalgam restorations to tooth structures, which is a NEW INDICATION FOR USE. Prime & Bond™ Self-Curing Activator is an acetone/ethanol solution of an aromatic sulfinic acid salt. When applied in combination with Prime & Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent, the Activator initiates a self-cure, free radical polymerization of the monomers and oligomers present in the Prime & Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent. {1} 510(k) SUMMARY (cont'd.) Page 2 of 2 # TRADE NAME: PRIME & BOND™ 2.1 MULTIPURPOSE DENTIN/ENAMEL BONDING AGENT WITH ACTIVATOR TECHNOLOGICAL CHARACTERISTICS: No changes have been made to Prime & Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent in design, material, chemical composition or manufacturing process. All of the components of the Prime & Bond™ Self-Curing Activator have been used in legally marketed devices or have been proven to be safe for dental use. We believe the fact that Prime & Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent has not been changed, the prior use of the Prime & Bond™ Self-Curing Activator components in predicate devices, the performance data presented for the use of Prime & Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent with the Prime & Bond™ Self-Curing Activator, and the cytotoxicity test results of the Activator active ingredient support the safety and effectiveness of Prime & Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent when used with Prime & Bond™ Self-Curing Activator for the new indications for use.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...