LINAC SCALPEL FLOORSTAND
Device Facts
| Record ID | K964517 |
|---|---|
| Device Name | LINAC SCALPEL FLOORSTAND |
| Applicant | Surgical Navigation Technologies, Inc. |
| Product Code | IYE · Radiology |
| Decision Date | Apr 1, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.5050 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
This submission describes a floorstand based coordinate and fixation system that is intended to be used to place patients at the isocenter of a linear accelerator for stereotactic radiosurgery or radiotherapy procedures. The Linac Scalpel™ Floorstand is designed to provide secondary collimation and a fixed isocenter for Radiosurgery or Radiotherapy when used in conjunction with a commercially available Linear Accelerator.
Device Story
Mechanical floorstand system; provides secondary collimation and fixed isocenter for linear accelerator-based stereotactic radiosurgery/radiotherapy. System attaches via gimbal bearing to linear accelerator head; positioned at predetermined floor points. Controls collimation accuracy during radiation beam arcing and patient position accuracy during couch repositioning. Used in clinical radiation oncology settings by trained medical staff. Ensures precise alignment of radiation beam to target; facilitates accurate dose delivery; improves treatment precision for patient.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Mechanical structure; gimbal bearing attachment; floor-mounted coordinate and fixation system. No electronic, software, or energy-based components described.
Indications for Use
Indicated for patients undergoing stereotactic radiosurgery or radiotherapy procedures requiring precise patient positioning and secondary collimation at the isocenter of a linear accelerator.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Related Devices
- K971893 — SNT LINAC ACCESSORIES · Surgical Navigation Technologies, Inc. · Nov 13, 1997
- K983660 — EXACTRAC (BRAINLAB PATIENT POSITIONING SYSTEM) · Brainlab, Inc. · Jul 14, 1999
- K961484 — CASS SYSTEM (MODIFICATION) · Medical Instrumentation & Diagnostics Corp.(Midco) · Nov 7, 1996
- K053500 — FRAMELESS RADIOSURGERY COMPONENTS · Brainlab AG · Feb 3, 2006
- K223839 — TrueBeam, TrueBeam STx, Edge, VitalBeam · Varian Medical Systems, Inc. · May 22, 2023