ROCHE COBAS AMPLICOR CHLAMYDIA TRACHOMATIS TEST

K964507 · Roche Molecular Systems, Inc. · MKZ · Jun 13, 1997 · Microbiology

Device Facts

Record IDK964507
Device NameROCHE COBAS AMPLICOR CHLAMYDIA TRACHOMATIS TEST
ApplicantRoche Molecular Systems, Inc.
Product CodeMKZ · Microbiology
Decision DateJun 13, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3120
Device ClassClass 1

Indications for Use

The COBAS AMPLICOR™ Chlamydia trachomatis (CT) Test is a qualitative in vitro diagnostic test for the detection of Chlamydia trachomatis in clinical specimens on the COBAS AMPLICOR analyzer. The test utilizes the Polymerase Chain Reaction (PCR) nucleic acid amplification technique and nucleic acid hybridization for the detection of Chlamydia trachomatis plasmid DNA in endocervical, urethral (male), and urine (male) samples.

Device Story

Automated in vitro diagnostic test for Chlamydia trachomatis detection; utilizes PCR amplification and nucleic acid hybridization. Input: endocervical swabs, male urethral swabs, male urine. Operation: automated pipetting, thermal cycling, hybridization, and detection on COBAS AMPLICOR analyzer. Detection uses magnetic microparticles as solid support; biotinylated primers (CP24/CP27) target CT DNA; biotinylated amplicon hybridizes to CP35 probe. Detection reaction measures chromophore absorbance at 660 nm via horseradish peroxidase oxidation of TMB. Used in clinical laboratory settings by trained personnel. Output: qualitative result based on optical density. Benefits: provides automated, high-throughput screening for Chlamydia infection, replacing manual microwell plate methods.

Clinical Evidence

Comparative clinical performance evaluated using 639 clinical specimens (266 endocervical swabs, 93 male urethral swabs, 280 male urine). Results demonstrated equivalence between the COBAS AMPLICOR and the predicate AMPLICOR microwell plate test. Analytical sensitivity is 1 Inclusion Forming Unit (IFU) for all 15 Chlamydia serovars. Analytical specificity tested against 105 bacteria, yeasts, and viruses with no cross-reactivity. Reproducibility confirmed via multi-operator, multi-analyzer study over three days with 100% correct results for spiked samples.

Technological Characteristics

Automated nucleic acid amplification and detection system. Uses magnetic microparticles as solid support for DNA probes. Detection via horseradish peroxidase-mediated oxidation of TMB, measured at 660 nm. Employs PCR amplification with biotinylated primers (CP24, CP27). Automated thermal cycling, pipetting, and washing. System is networked/automated analyzer-based. Reagents include denaturation solution, substrate A/B, and wash concentrate.

Indications for Use

Indicated for qualitative detection of Chlamydia trachomatis plasmid DNA in endocervical swabs, male urethral swabs, and male urine samples using PCR and nucleic acid hybridization on the COBAS AMPLICOR analyzer.

Regulatory Classification

Identification

Chlamydia serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to chlamydia in serum. Additionally, some of these reagents consist of chlamydia antisera conjugated with a fluorescent dye used to identify chlamydia directly from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genus Chlamydia and provides epidemiological information on these diseases. Chlamydia are the causative agents of psittacosis (a form of pneumonia), lymphogranuloma venereum (a venereal disease), and trachoma (a chronic disease of the eye and eyelid).

Predicate Devices

Submission Summary (Full Text)

{0} K964507 Roche Molecular Systems, Inc. Somerville, New Jersey 08876 COBAS AMPLICOR™ Chlamydia Trachomatis Test 510(k) Summary 510(k) Summary JUN 13 1997 Roche COBAS AMPLICOR™ Chlamydia trachomatis Test Roche Molecular Systems, Inc. 1080 U.S. Highway 202 Somerville, New Jersey 08876-1760 (908) 253-7200 ## Intended Use: The COBAS AMPLICOR™ Chlamydia trachomatis (CT) Test is a qualitative *in vitro* diagnostic test for the detection of *Chlamydia trachomatis* in clinical specimens on the COBAS AMPLICOR analyzer. The test utilizes the Polymerase Chain Reaction (PCR) nucleic acid amplification technique and nucleic acid hybridization for the detection of *Chlamydia trachomatis* plasmid DNA in endocervical, urethral (male), and urine (male) samples. ## Description of the Device: The Roche COBAS AMPLICOR™ Chlamydia trachomatis test performed on the COBAS AMPLICOR analyzer is an automated modification of the Roche AMPLICOR Chlamydia trachomatis Microwell Plate Test performed using the Perkin Elmer 9600 thermal cycler and a conventional photometric microwell plate reader (K922906/C). The AMPLICOR™ Chlamydia trachomatis Test was previously shown to be substantially equivalent to tissue culture with immunofluorescent staining for the detection of *Chlamydia trachomatis* in urogenital swab and male urine samples. The AMPLICOR Chlamydia trachomatis Test has been modified to allow the automation of the amplification and detection test procedure using the COBAS AMPLICOR analyzer. Specimen collection, transport and processing and target amplification with the COBAS AMPLICOR Chlamydia Trachomatis Test are identical to the AMPLICOR Chlamydia Trachomatis Test. COBAS AMPLICOR and microwell AMPLICOR Chlamydia Trachomatis Test detection reagents incorporate identical probes using the appropriate solid support. The clinical and non-clinical performance of the COBAS AMPLICOR Chlamydia Trachomatis Test is substantially equivalent to the AMPLICOR Chlamydia Trachomatis Test. ## Similarities and Differences to Predicate Device: The COBAS AMPLICOR Chlamydia trachomatis Test is similar in design, reagent composition, function and intended use to the commercial microwell plate format AMPLICOR Chlamydia trachomatis Test (K922906/C). The COBAS AMPLICOR Chlamydia trachomatis Test uses without modification the AMPLICOR Chlamydia trachomatis Test STD Swab Specimen Collection and Transport Kit, the AMPLICOR STD Swab and the Urine Specimen Preparation Kits for specimen collection, specimen preparation and PCR amplification (see Table 1). The COBAS AMPLICOR Detection Kit reagents are similar in function and formulation to those in the AMPLICOR Chlamydia trachomatis Detection Test. The solid support for the DNA detection probe has been changed from a polystyrene microwell plate to magnetic microparticles to permit automation of the detection reaction on the COBAS AMPLICOR. ## Similarities - AMPLICOR STD Specimen Collection and Transport Kit (P/N 83075) for the collection and transport of male and female swab specimens. - AMPLICOR STD Specimen Preparation Kits (P/N 83002 and 83006) for the processing of swab and urine specimens. - AMPLICOR CT Amplification Kit (P/N 83008). - Identical biotinylated primers (CP24 and CP27) for defining the CT DNA target sequence for amplification. f\reg\regdocs\ctdocs\ctcobas\fdaresp\ctresp.new\26 Page 036 {1} Roche Molecular Systems, Inc. Somerville, New Jersey 08876 COBAS AMPLICOR™ Chlamydia Trachomatis Test 510(k) Summary - Hybridization of biotinylated amplicon generated in the PCR reaction to the same CP35 CT-specific DNA probe - Identical Detection Reagent formulations (Denaturation Solution, Substrate A, Substrate B and 10-X Wash Concentrate) - Identical detection reactions based on the measurement chromophore absorbance following oxidation of 3,3′,5,5′-tetramethybenzidine by horseradish peroxidase in the presence of hydrogen peroxide. ## Differences - Automated pipetting of amplified samples and reagent, washing, thermal cycling, hybridization and detection steps of the test procedure. - The avidin-HRP Conjugate Buffer in the COBAS AMPLICOR *Chlamydia trachomatis* Test contains bovine serum albumin compared to bovine gamma globulin in the AMPLICOR *Chlamydia trachomatis* MWP Test. - Change of solid support from polystyrene microwell plate to magnetic microparticles, - Slightly modified thermal cycle profile for the COBAS AMPLICOR to give equivalent performance to the Perkin Elmer 9600 thermal cycler. - No Stop Reagent is required with the COBAS AMPLICOR assay. Since the automated analyzer precisely times each step of the hybridization and detection procedures, it is not necessary to add Stop Reagent at the conclusion of the substrate incubation step. Accordingly, the optical density of the oxidation products from the detection reaction is measured at 660 nm in the COBAS AMPLICOR *Chlamydia trachomatis* Test as compared to 450 nm in the AMPLICOR *Chlamydia trachomatis* MWP Test system since the Stop Reagent modifies the oxidation state of the TMB reaction product, from the detection reaction, shifting the absorbance maximum of the chromophore from 660 nm to 450 nm. ## Non-Clinical Performance: The analytical sensitivity (limit of detection) of the COBAS AMPLICOR *Chlamydia trachomatis* Test is 1 Inclusion Forming Unit for all 15 Chlamydia serovars (A, B, Ba, C, D, E, F, G, H, I, J, K, L1, L2, L3). The analytical specificity for the AMPLICOR *Chlamydia trachomatis* Test was evaluated using 105 bacteria, yeasts and viruses. No organism gave a result greater than 0.016 A₆₆₀. These results are consistent with those observed for the AMPLICOR *Chlamydia Trachomatis* Test. The reproducibility of the Roche COBAS AMPLICOR *Chlamydia trachomatis* Test on the COBAS AMPLICOR analyzer was determined in a multi-operator analysis of swab samples spiked with *Chlamydia trachomatis* organism at four levels, AMPLICOR *C. trachomatis* positive and negative kit controls, and six patient samples spanning the reportable range for the assay (24 samples per test run; two complete A-tube rings). Samples were analyzed by independent operators for three days using three different analyzers. The test results for all samples (STM specimens spiked at 50, 10, 1 and 0 IFU/PCR, positive and negative kit controls, and the patient specimens) were 100% correct. The reproducibility of the assay was calculated from the absorbance values obtained from the replicates of the spiked STM specimens and the Kit Controls. The complete results of the Analysis of Variance calculations for these data are provided in Table 2. ## Clinical Performance: The comparative clinical performance between the COBAS AMPLICOR *Chlamydia trachomatis* Test and the AMPLICOR *Chlamydia Trachomatis* Test was determined using 639 clinical specimens (266 endocervical swabs, 93 male urethral swabs, and 280 male urine specimens). Equivalent results were obtained for specimens tested by the two test procedures. March 31, 1997 f\reg\regdocs\ctdocs\ctcobas\fdaresp\ctresp.new\27 Page 037 {2} Roche Molecular Systems, Inc. Somerville, New Jersey 08876 COBAS AMPLICOR™ Chlamydia Trachomatis Test 510(k) Summary # TABLE 1 ## COBAS AMPLICOR Chlamydia trachomatis Test Reproducibility ### Analysis of Variance - Statistical Summary | PRECISION MEASUREMENT | TEST SAMPLE | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | C. trachomatis Spiked STM (IFU/PCR Test) | | | | Kit Controls | | | | 0 | 1 | 10 | 50 | Negative | Positive | | Total Replicates | 27 | 27 | 27 | 27 | 27 | 27 | | Mean Absorbance | 0.004 | 3.856 | 3.685 | 3.639 | 0.004 | 3.831 | | Minimum | 0.000 | 3.646 | 3.220 | 3.220 | 0.000 | 3.279 | | Maximum | 0.017 | 4.000 | 4.000 | 4.000 | 0.011 | 4.000 | | Between Day Variance | 0.0000 | 0.0000 | 0.0040 | 0.0037 | 0.0000 | 0.0022 | | Standard Deviation | 0.0013 | 0.0000 | 0.0632 | 0.0610 | 0.0000 | 0.0468 | | CV (%) | 32.0 | 0.0 | 1.7 | 1.7 | 0.0 | 1.2 | | Between Operator Variance | 0.0000 | 0.0145 | 0.0271 | 0.0230 | 0.0000 | 0.0587 | | Standard Deviation | 0.0029 | 0.1206 | 0.1647 | 0.1517 | 0.0027 | 0.2422 | | CV (%) | 72.4 | 3.1 | 4.5 | 4.2 | 69.7 | 6.3 | | Within Operator Variance | 0.0000 | 0.0059 | 0.0162 | 0.0125 | 0.0000 | 0.0109 | | Standard Deviation | 0.0035 | 0.0767 | 0.1274 | 0.1118 | 0.0026 | 0.1045 | | CV (%) | 87.0 | 2.0 | 3.5 | 3.1 | 67.4 | 2.7 | | Total Variance | 0.0000 | 0.0204 | 0.0474 | 0.0392 | 0.0000 | 0.0718 | | Standard Deviation | 0.0047 | 0.1429 | 0.2176 | 0.1980 | 0.0037 | 0.2679 | | CV (%) | 117.6 | 3.7 | 5.9 | 5.4 | 97.0 | 7.0 | # TABLE 2 ## Comparative Clinical Performance ### COBAS AMPLICOR vs AMPLICOR Chlamydia Trachomatis Test | | Culture Positive | | Culture Negative | | | | --- | --- | --- | --- | --- | --- | | | | | AMPLICOR Positive | | | | | AMPLICOR Positive | AMPLICOR Negative | MOMP Positive | MOMP Negative | AMPLICOR Negative | | FEMALE SWABS | | | | | | | COBAS Positive | 19 | 1 | 13 | 2* | 0 | | COBAS Negative | 0 | 1 | 0 | 1 | 229 | | MALE SWABS | | | | | | | COBAS Positive | 13 | 0 | 3 | 0 | 1** | | COBAS Negative | 0 | 0 | 0 | 0 | 76 | | MALE URINE | | | | | | | COBAS Positive | 32 | 2 | 30 | 1 | 2 | | COBAS Negative | 2 | 3 | 3 | 0 | 202 | * inconclusive results by MOMP alternative primer pair testing ** COBAS negative on repeat testing f\reg\regdocs\ctdocs\ctcobas\fdaresp\ctresp.new\28 Page 038 {3} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JUN 13 1997 Alex Wesolowski Director, Regulatory and Clinical Affairs Roche Molecular Systems, Inc. 1080 U.S. Highway 202 Branchburg, NJ 08876-1760 Re: K964507 Trade Name: Roche Cobas Amplicor™ Chlamydia Trachomatis Test Regulatory Class: I Product Code: MKZ Dated: November 7, 1996 Received: November 8, 1996 Dear Mr. Wesolowski: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4} Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5} Page 1 of 1 510(k) Number (if known) _______________ Device Name: COBAS AMPLICOR™ Chlamydia Trachomatis Test Indications for Use: The COBAS AMPLICOR™ Chlamydia trachomatis Test is a qualitative *in vitro* diagnostic test for the detection of *Chlamydia trachomatis* in clinical specimens on the COBAS AMPLICOR analyzer. The test utilizes the Polymerase Chain Reaction (PCR) nucleic acid amplification technique and nucleic acid hybridization for the detection of *Chlamydia trachomatis* plasmid DNA in endocervical, male urethral, and male urine samples. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number K 96 4507
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...