K964219 · Goulter Medical, Inc. · EYZ · Aug 11, 1997 · Gastroenterology, Urology
Device Facts
Record ID
K964219
Device Name
ZIP CONDOM CATHETER
Applicant
Goulter Medical, Inc.
Product Code
EYZ · Gastroenterology, Urology
Decision Date
Aug 11, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5250
Device Class
Class 1
Indications for Use
1. For Penile Attachment Only 2. Non-indwelling Urinary Continence Aid 3. Management of Urinary Incontinence 4. Urine Collection Device
Device Story
ZIP™ Condom Catheter is a reusable, single-piece external male urine collection system. Components include a natural latex condom/sheath with an integral 450 mL urine collection compartment, a Velcro-like band/support strap, an application ring, and specialized male underpants. Device operates via gravity-fed urine flow from the sheath into the collection compartment, separated by non-return flow valves to prevent backflow. Distal end features a twist-open/close drain valve for emptying. Used by patients for urinary incontinence management; intended for daily living activities. Patients clean the device with soap/detergent and surface disinfectants; recommended replacement after one week due to valve performance degradation from sediment/crystals. System design allows for stand-up urination and is intended to be unobtrusive under clothing.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on design, material composition, and intended use comparisons to legally marketed predicate devices.
Technological Characteristics
Constructed from natural rubber latex. Features a single-piece design with an integral 450 mL collection compartment, non-return flow valves, and a twist-open drain valve. Secured via Velcro-like band and support strap. Reusable for approximately one week. Not sterile; requires manual cleaning and surface disinfection.
Indications for Use
Indicated for male patients requiring a non-indwelling urinary continence aid or management of urinary incontinence via penile attachment for urine collection.
Regulatory Classification
Identification
A urine collector and accessories is a device intended to collect urine. The device and accessories consist of tubing, a suitable receptacle, connectors, mechanical supports, and may include a means to prevent the backflow of urine or ascent of infection. The two kinds of urine collectors are:(1) A urine collector and accessories intended to be connected to an indwelling catheter, which includes the urinary drainage collection kit and the closed urine drainage system and drainage bag; and (2) A urine collector and accessories not intended to be connected to an indwelling catheter, which includes the corrugated rubber sheath, pediatric urine collector, leg bag for external use, urosheath type incontinence device, and the paste-on device for incontinence.
Special Controls
*Classification* —(1)*Class II (special controls) for a urine collector and accessories intended to be connected to an indwelling catheter.* The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.(2)
*Class I (general controls).* For a urine collector and accessories not intended to be connected to an indwelling catheter, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
Uri-Drain® - Male Urinary Control Device (Preamendment)
Uri-Drain® - Leg Bag - Reusable Deluxe (Preamendment)
Submission Summary (Full Text)
{1}
K964219 (1 of 3)
| Premarket Notification
Section 510(k) | Summary of Safety and Effectiveness Information
for the **ZIP™ Condom Catheter** |
| --- | --- |
**Regulatory Authority:** Safe Medical Devices Act of 1990, 21 CFR 807.92
1. **Device Trade Name:** *ZIP™ Condom Catheter* AUG 11 1997
**Common Name:** External male urine collection device
**Classification Name:** Device, Incontinence, Urosheath Type, Urine Collector and Accessories
2. **Establishment Name & Registration Number:**
**Name:** Goulter Medical, Inc.
**Number:** Pending
3. **Classification:**
§ 876.5250 Urine collector and accessories. (a) Identification. A urine collector and accessories is a device intended to collect urine. The device and accessories consist of tubing, a suitable receptacle, connectors, mechanical supports, and may include a means to prevent the back-flow of urine or ascent of infection. The two kinds of urine collectors are: (1) A urine collector and accessories intended to be connected to an indwelling catheter, which includes the urinary drainage collection kit and the closed urine drainage system and drainage bag; and (2) a urine collector and accessories not intended to be connected to an indwelling catheter, which includes the corrugated rubber sheath, pediatric urine collector, leg bag for external use, urosheath type incontinence device, and the paste-on device for incontinence. (b) Classification. (1) Class II (performance standards) for a urine collector and accessories intended to be connected to an indwelling catheter. (2) Class I (general controls) for a urine collector and accessories not intended to be connected to an indwelling catheter. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice regulations in Part 820, with the exception of 820.180, with respect to general requirements concerning records, and 820.198, with respect to complaint files.
* **UNDERLINE ADDED**
**Device Class:** Class I for the proposed indications.
**Classification Panel:** Urology
**Product Code:** 78EXJ
**Tier Class:** Tier I Device
4. **Company Contact:**
Ms. Susan Clymer
One Embarcadero Center
Suite 1020
San Francisco, California 94111-3600
415.217.5656 - 415.217.5655 fax
5. **Special Controls:**
FDA Mandated Special Controls do not apply to this device.
Urores.doc
5
{2}
K964219 (P2 of 3)
## 6. Substantially Equivalent Device(s):
1. Mentor Freedom Cath® - Self Adhering External Male Catheter
2. Everyday - Self Adhesive Urinary External Catheter
3. Uri-Drain® - Male Urinary Control Device
4. Uri-Drain® - Leg Bag - Reusable Deluxe
## 7. Device Description:
The ZIP™ Condom Catheter has four basic elements: 1) the condom/sheath with integral urine collection compartment; 2) the Velcro-like band and support strap; 3) the Application ring; and 4) the specially designed male underpants brief.
The condom/sheath with integral urine collection compartment forms a bulbous tube approximately 10 ½ inches long. The tube is divided into two sections, the condom/sheath and the urine collection compartment. Non-return flow valves separate the condom/sheath portion from the urine collection compartment. The distal end of the urine collection compartment is affixed with a twist-open twist-closed stop-cock drain valve. The urine collection capacity is approximately 450 ml. Typical human urine output thus yields a time related capacity of about 4 to 6 hours.
A series of 4 sizes differing in sheath diameter and length will be offered. Specific name designators for the sizes are yet to be determined.
**Design:** The ZIP™ Condom Catheter expands upon existing male urine continence devices by first approaching the device as a “total” or “system” concept rather than just a collection of items originally developed for other urine collection purposes.
Only the essential elements of the ideal male urine continence system were considered. First, the system had to be easy to apply and use. Second, it should allow use of traditional male urination methods, i.e., the stand-up urinals found in public restrooms. Third, the system should be small enough to go unnoticed during the typical activities of daily living. Fourth, the system should have adequate capacity so that urine disposal intervals are not unreasonably short. Lastly, the system should be reusable and easily cleaned and sanitized.
This is a natural latex containing product. The ZIP™ Condom Catheter is made from the exact same natural latex formulation as the Everyday - Self-Adhesive Urinary External Catheter made by Hollister, Inc. The use of pure latex to construct the ZIP™ Condom Catheter requires the inclusion of certain FDA required cautionary statements. Latex containing devices must legibly bear an appropriate allergic reaction statement on the device labeling. In conformance with this FDA requirement, Goulter Medical, Inc. attaches a label containing the following cautionary statement to the device. “This product is made from natural rubber latex which may cause allergic reactions in some individuals”.
## The Cleared Indications for use are:
1. For penile attachment only
2. Non-Indwelling Urinary continence aid
3. Management of Urinary Incontinence
4. Urine Collection Device
Urores.doc
6
{3}
K964219 (P3.43)
## Claims:
1. Unobtrusive
2. Reusable
3. Cost effective
4. Single piece system
5. Enhanced self-image and esteem
6. Improves social freedom
7. Lightweight
8. Less physically restrictive
9. Easy to use
## 8. Cleaning/Sterilization/Re-sterilization:
The device may not be sterilized or re-sterilized. Cleaning and washing using soap/detergent/water followed by surface disinfection may be employed using commercially available liquid germicidal agents. The device is reusable for a limited period of time. Generally, with reasonable care, the device may be used for about a week. Though in individual cases this period may be shorter or longer. Most often the limiting factor is the performance of the anti-reverse flow valves which tend to be performance affected by accumulating urine sediments and crystals. It is recommended that the device not be reused for more than a week and it should be replaced at once if the valves fail to operate properly.
## 9. Equivalence:
Based on the materials, intended uses, design and clinical effectiveness, the ZIP™ Condom Catheter is substantially equivalent to the previously referenced legally marketed male external urinary collection devices.
The feature comparison chart on the following page graphically demonstrates equivalence.
## 10. Feature Comparison Table:
| FEATURE | ZIP™ Condom Catheter | Mentor Freedom Cath® | Everyday External Catheter | Uri-Drain® | SE? |
| --- | --- | --- | --- | --- | --- |
| Materials: | Medical Grade Latex | Latex | Identical Latex Formulation | Latex | Yes |
| Intended Use(s): | Non-Indwelling Male Urinary Continence Aid Management of Male Urinary Incontinence Male Bladder Training Aid to Improve Urinary Continence | Same | | | Yes |
| Design: | Single Piece Unit | Condom Only | Condom Only | Condom Only | Yes |
| Method of Operation: | One Way Flow to Collection Bag | Same | Same | Same | Yes |
| Bag: | Integral to Condom Sheath | Requires Secondary Leg Bag | Requires Secondary Leg Bag | Requires Secondary Leg Bag | Yes |
| Bag Capacity: | Approx. 450 mL | About 1 Liter | About 1 Liter | About 1 Liter | Yes |
| Underpant: | Male Brief W/ Pouch | NA | NA | NA | No |
| How Secured: | Velcro Band & Strap | Adhesive Sheath | Spiral Foam Band | Spiral Foam Band | Yes |
| Sizes: | 4 | 4 | 3 | 3 | Yes |
| Additional Items Needed to Use the Device: | None - Sold Complete | Leg Bag Connectors Tubing Adhesive tube Holder | Leg Bag Connectors Tubing Tube clamp | Leg Bag Connectors Tubing Adhesive tube Holder | No |
| Manufacturer: | Goulter Medical | Mentor | Hollister | Sherwood Medical | Yes |
| Product Code: | 78EXJ | 78EXJ | 78EXJ | 78EXJ | Yes |
| K - Number | K964219 | Preamendment | Preamendment | Preamendment | |
Urores.doc
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
AUG 11 1997
Goulter, Medical, Inc.
c/o Buckman Company, Inc.
Mr. David W. Schlerf
1000 Burnett Avenue, Suite 450
Concord, California 94520
Re: K964219
Zip™ Condom Catheter
Dated: July 12, 1997
Received: July 21, 1997
Regulatory Class: I
21 CFR §876.5250/Product Code: 78 EXJ & EYZ
Dear Mr. Schlerf:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Lillian Yin, Ph.D.
Director, Division of Reproductive,
Abdominal, Ear, Nose and Throat,
and Radiological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{5}
Page 1 of 1
510(k) Number: K964219
Device Name: ZIP™ Condom Catheter
Indications For Use:
1. For Penile Attachment Only
2. Non-indwelling Urinary Continence Aid
3. Management of Urinary Incontinence
4. Urine Collection Device
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY
Concurrence of CDRH, Office of Device Evaluation (ODE)
Colin M. Gillard
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K964219
Prescription Use ☐
OR
Over-The-Counter Use ☑
(Per 21 CFR 801.109)
(Optional format 1-2-96)
Urores.doc
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.