← Product Code [DSP](/productcode/DSP) · K963920

# 8 FR.-30CC NARROWFLEX INTRA-AORTIC BALLOON CATHETER, (K963920)

_Arrow Intl., Inc. · DSP · Jun 17, 1997 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K963920

## Device Facts

- **Applicant:** Arrow Intl., Inc.
- **Product Code:** [DSP](/productcode/DSP.md)
- **Decision Date:** Jun 17, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.3535
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

Refractory left ventricular power failure. Cardiogenic shock unstable refractory angina. Mechanical complication due to acute myocardial infarction; i.e., ventricular septal defect mitral regurgitation or papillary muscle rupture. Impending infraction, ischemia related intractable ventricular arrhythmias. Septic shock. Support for failed angioplasty and valvuloplasty. Cardiac support for high risk general surgical patients.

## Device Story

Arrow 8 Fr. - 30cc Narrowflex™ Intra-Aortic Balloon Catheter; provides mechanical circulatory support. Device consists of 30cc balloon on 8 Fr. catheter; utilizes Nitinol inner lumen and kink-resistant outer lumen. Operates via inflation/deflation cycles to provide hemodynamic support. Used in clinical settings for patients with acute cardiac failure or shock. Physician-operated; provides cardiac assistance to improve perfusion and reduce myocardial workload. Benefits include stabilization of hemodynamics in critical cardiac conditions.

## Clinical Evidence

Bench testing only. Performance metrics evaluated include total cycle time, displaced volume, and first response, showing comparability to predicate devices.

## Technological Characteristics

8 Fr. intra-aortic balloon catheter; Nitinol inner lumen; kink-resistant outer lumen; 30cc balloon volume. Mechanical circulatory support device.

## Regulatory Identification

An intra-aortic balloon and control system is a prescription device that consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning during certain life-threatening emergencies, and a control system for regulating the inflation and deflation of the balloon. The control system, which monitors and is synchronized with the electrocardiogram, provides a means for setting the inflation and deflation of the balloon with the cardiac cycle.

## Special Controls

*Classification.* (1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. The special controls for this device are:(i) Appropriate analysis and non-clinical testing must be conducted to validate electromagnetic compatibility and electrical safety of the device;
(ii) Software verification, validation, and hazard analysis must be performed;
(iii) The device must be demonstrated to be biocompatible;
(iv) Sterility and shelf-life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components;
(v) Non-clinical performance evaluation of the device must demonstrate mechanical integrity, durability, and reliability to support its intended purpose; and
(vi) Labeling must include a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 31, 2014, for any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976. Any other intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- 8 Fr-40cc NarrowFlex™ ([K960713](/device/K960713.md))
- 8 Fr-30cc ([K892799](/device/K892799.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K963920

JUN 17 1997

510(k) SUMMARY - K963920

The device is substantially equivalent to the following legally marketed Arrow intra-aortic balloon devices:

1. 8 Fr-40cc NarrowFlex™, with Nitinol inner lumen and kink resistant outer lumen, product no. IAB-04840, premarket notification K960713.
2. 8 Fr-30cc, with stainless steel inner lumen and plastic outer lumen, product no. IAB-04230, premarket notification K892799.

This device is identical in construction to the predicate 8 Fr-40cc NarrowFlex™ device IAB-04840 with the exception of the 30cc balloon and shorter overall length. The balloon volume and the overall length have been adapted from the 8 Fr-30cc IAB-04230 predicate.

The device is indicated for the following conditions:

Refractory left ventricular power failure. Cardiogenic shock unstable refractory angina. Mechanical complication due to acute myocardial infarction; i.e., ventricular sepial defect mitral regurgitation or papillary muscle rupture. Impending infraction, ischemia related intractable ventricular arrhythmias. Septic shock. Support for failed angioplasty and valvuloplasty. Cardiac support for high risk general surgical patients.

The device has comparable technological characteristics to the predicate devices.

The nonclinical test results included in the submission showing comparable performance to the predicate devices are as follows:

Performance Test
- Total cycle time
- Displaced volume
- First response

96123

{1}

DEPARTMENT OF HEALTH &amp; HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850

JUN 17 1997

Mr. Thomas D. Nickel
Vice President, Regulatory Affairs and Quality Assurance
Arrow International, Inc.
3000 Bernville Road
Reading, Pennsylvania 19605

Re: K963920
Arrow 8 Fr. - 30cc Narrowflex™ Intra-Aortic Balloon Catheter
Regulatory Class: III (Three)
Product Code: 74 DSP
Dated: January 14, 1997
Received: January 17, 1997

Dear Mr. Nickel:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}

Page 2 - Mr. Thomas D. Nickel

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "dsmo@fdadr.cdrh.fda.gov."

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health

Enclosure

{3}

Page ___ of ___

510(k) Number (if known): K963920

Device Name: Arrowflex 8Fr 30cc Neuroplex Tm

P.O. Box 12888
Reading, PA 19612

ARROW
INTERNATIONAL

3000 Bernville Road
Reading, PA 19605
(610) 378-0131
FAX: (610) 374-5360

## Section 11 - Indications

Refractory left ventricular power failure. Cardiogenic shock unstable refractory angina. Mechanical complication due to acute myocardial infarction; i.e., ventricular septal defect mitral regurgitation or papillary muscle rupture. Impending infraction, ischemia related intractable ventricular arrhythmias. Septic shock. Support for failed angioplasty and valvuloplasty. Cardiac support for high risk general surgical patients.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

![img-0.jpeg](img-0.jpeg)

(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K963920

Prescription Use ☑
(Per 21 CFR 801.109)

OR

Over-The-Counter Use ☐

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/device/K963920](https://fda.innolitics.com/device/K963920)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
