CM02 CUVETTE, 1/2, CM02 CUVETTE, 3/8 AND CM02 CUVETTE, 1/4
Device Facts
| Record ID | K963698 |
|---|---|
| Device Name | CM02 CUVETTE, 1/2, CM02 CUVETTE, 3/8 AND CM02 CUVETTE, 1/4 |
| Applicant | International Biophysics Corp. |
| Product Code | DTY · Cardiovascular |
| Decision Date | Mar 20, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.4410 |
| Device Class | Class 2 |
Intended Use
The IBC CMO2 Flow Through Cuvette is a single piece, single use, disposable component intended for use with the Bentley OxySat Monitor as a direct substitute for the Bentley OTC Flow Through Cuvette.
Device Story
Single-use, disposable flow-through cuvette; used with Bentley OxySat Monitor; facilitates blood oxygen saturation monitoring. Input: blood flow through cuvette; Output: optical path for transducer measurement. Operation: single-piece injection-molded polycarbonate housing; replaces multi-piece assembled cuvettes to improve structural integrity, reduce leak risk, and ensure consistent geometry for transducer attachment. Used in clinical settings where OxySat monitoring is required. Benefits: reduced manufacturing cost, improved reliability, and consistent performance compared to predicate.
Clinical Evidence
Bench testing only. Comparative performance analysis against predicate demonstrated identical accuracy (r=.998). Biocompatibility testing performed per Tripartite Guidelines; device confirmed non-hemolytic.
Technological Characteristics
Single-piece injection-molded polycarbonate; available in 1/4", 3/8", and 1/2" sizes; sterile (Ethylene Oxide); disposable; non-hemolytic.
Indications for Use
Indicated for use with the Bentley OxySat Monitor as a flow-through cuvette for blood monitoring in patients requiring oxygen saturation measurement.
Regulatory Classification
Identification
A cardiopulmonary bypass in-line blood gas sensor is a transducer that measures the level of gases in the blood.
Predicate Devices
- Bentley OTC Flow Through Cuvette
Related Devices
- K092549 — OXIPROBE, MODELS BM-100, BM-200, BM-400, BM-300, BM-300S, BM-600P · Bio Medical Technologies Co., Ltd. · Nov 16, 2009
- K141281 — CRIT-LINE CLIP (CLIC) BLOOD CHAMBER · Fresenius Medical Care Renal Therapies Group, LLC · Aug 14, 2014
- K082546 — REUSABLE OHMEDA (U410-02), DATEX (U410-09) COMPATABLE ADULT FINGER CLIP SENSOR, PHILLIPS ADULT SOFT TIP SENSOR(U403S-91) · Unimed Medical Supplies, Inc. · Dec 2, 2008
- K992215 — MULTICOMPATIBLE ENVITEC FINGER CLIP SENSOR, MULTICOMPATIBLE ENVITEC ADULT DISPOSABLE SENSOR, MULTICOMPATIBLE ENVITEC PED · Envitec-Wismar GmbH · Mar 29, 2000