GRASS MODEL 15 NEURODATA AMPLIFIER SYSTEM
Device Facts
| Record ID | K963669 |
|---|---|
| Device Name | GRASS MODEL 15 NEURODATA AMPLIFIER SYSTEM |
| Applicant | Astro-Med, Inc. |
| Product Code | GWL · Neurology |
| Decision Date | Dec 12, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 882.1835 |
| Device Class | Class 2 |
Device Story
Neurodata Amplifier System; processes physiological signals. Device houses up to 32 amplifier circuits in single enclosure; replaces manual switches with solid-state switches controlled via RS-232 serial port. Used in clinical/diagnostic settings for neurological signal amplification. Operates via capacity-coupled patient circuit; safety-extra-low-voltage (12VDC) operation. Isolates patient-applied circuits from earth ground. Provides signal amplification for healthcare provider review; facilitates neurological diagnostic assessment.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Physiological signal amplifier; 32-channel capacity; solid-state switching; RS-232 serial interface; capacity-coupled patient circuit; 12VDC operation; patient-to-earth isolation.
Regulatory Classification
Identification
A physiological signal amplifier is a general purpose device used to electrically amplify signals derived from various physiological sources (e.g., the electroencephalogram).
Predicate Devices
- Grass Model P511
- Melville Diagnostics MMC Series
Related Devices
- K172312 — eego amplifiers · Eemagine Medical Imaging Solutions GmbH · Mar 30, 2018
- K121996 — BE PLUS LTM AMPLIFIER GWI AMPLIFIER · Eb Neuro S.P.A. · Aug 3, 2012
- K041198 — NIC36 AMPLIFIER · Eb Neuro S.P.A. · Jun 1, 2004
- K060803 — G.USBAMP · G.Tec Medical Engineering GmbH · May 2, 2006
- K974495 — NEUROAMP EEG AMPLIFIER · Neurocorp, Ltd. · Feb 26, 1998