ACROMED PEDICLE SCREW

K963639 · Acromed Corp. · KWP · Nov 14, 1996 · Orthopedic

Device Facts

Record IDK963639
Device NameACROMED PEDICLE SCREW
ApplicantAcromed Corp.
Product CodeKWP · Orthopedic
Decision DateNov 14, 1996
DecisionSN
Submission TypeTraditional
Regulation21 CFR 888.3050
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ISOLA Spinal System, when used with pedicle screws, is intended for use in grade 3 or 4 spondylolisthesis at L5-S1 utilizing autologous bone graft and intended to be removed after solid fusion is attained. The ISOLA Spinal System, when not used with pedicle screws, is intended for use in grade 3 or 4 spondylolisthesis at L5-S1 utilizing autologous bone graft and intended to be removed after solid fusion is attained.

Device Story

AcroMed Pedicle Screw (4.75 mm) is a spinal fixation component for the ISOLA Spine System. Device features two-section design: long cancellous section with integral fixed lower nut; machine-threaded section with hexagonal drive head. Integral nut provides bolt connection strength and reduces stress concentration. Supplied with adjustable tapered nut and optional lock nut for rod-to-screw connection. Used by surgeons in spinal fusion procedures to stabilize L5-S1 segments in spondylolisthesis patients. Removed after solid fusion achieved.

Clinical Evidence

Bench testing only. Static and fatigue testing performed to demonstrate consistency with previously cleared components.

Technological Characteristics

Material: Implant grade stainless steel (ASTM F1314). Dimensions: 4.75 mm diameter; cancellous thread lengths 25-50 mm (5 mm increments). Design: Two-section screw with integral fixed lower nut, machine-threaded section, and hexagonal drive head. Mechanical fixation.

Indications for Use

Indicated for patients with grade 3 or 4 spondylolisthesis at L5-S1 requiring spinal fixation with autologous bone graft.

Regulatory Classification

Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} K963639 1 of 2 NOV 14 1996 # ISOLA SYSTEM ## AcroMed Pedicle Screw (4.75 mm) ### 510(k) SUMMARY **COMPANY:** AcroMed Corporation 3303 Carnegie Avenue Cleveland, OH 44115 **TRADENAME:** AcroMed Pedicle Screw **CLASSIFICATION:** Spondylolisthesis Spinal Fixation Device Unclassified, preamendment device system **DESCRIPTION:** The AcroMed Pedicle Screw is composed of two sections: a long cancellous section with an integral fixed lower nut, and a machine threaded section topped with an hexagonal drive head. The integral nut serves two purposes: it creates a stronger bolt connection between the screw and rod and it eliminates the “claw hammer” effect associated with other screw designs. The AcroMed Pedicle Screw is supplied with an adjustable tapered nut and an optional lock nut, which enables a pedicle screw and a slotted connector to be adjoined utilizing a locking nut system. The junction point of the adjustable integral nut and machine thread portion of the pedicle screw is also rounded to reduce potential stress concentration. The screw is available in cancellous thread lengths from 25-50 mm, in 5 mm increments. **MATERIAL:** The AcroMed Pedicle Screw is manufactured from implant grade stainless steel conforming to ASTM F1314 specifications. 1 L1/7 'd 8160 'oN Wd02:ε 9661 '8 'AON {1} 2 of 2 # INDICATIONS: The AcroMed Pedicle Screw is designed to be utilized with the ISOLA Spine System (K944737). The ISOLA Spinal System, when used with pedicle screws, is intended for use in grade 3 or 4 spondylolisthesis at L5-S1 utilizing autologous bone graft and intended to be removed after solid fusion is attained. The ISOLA Spinal System, when not used with pedicle screws, is intended for use in grade 3 or 4 spondylolisthesis at L5-S1 utilizing autologous bone graft and intended to be removed after solid fusion is attained. The ISOLA Spinal System, when not used with pedicle screws, is intended for use in grade 3 or 4 spondylolisthesis at L5-S1 utilizing autologous bone graft and intended to be removed after solid fusion is attained. ## PERFORMANCE DATA: Static and fatigue testing show the AcroMed pedicle screws (4.75 mm) to perform consistently with previously cleared components. ## SUBSTANTIAL EQUIVALENCE: The AcroMed 4.75 mm diameter pedicle screw is equivalent to other AcroMed pedicle screws as cleared under K951116. L1/S d 8160°N Wd02:ε 96618
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...