INNOVATIVE TECHNOLOGIES ALIGINATE/PECTIN WOUND DRESSINGS
Device Facts
| Record ID | K963598 |
|---|---|
| Device Name | INNOVATIVE TECHNOLOGIES ALIGINATE/PECTIN WOUND DRESSINGS |
| Applicant | Innovative Technologies , Ltd. |
| Product Code | KMF · General Hospital |
| Decision Date | Nov 27, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.5090 |
| Device Class | Class 1 |
| Attributes | Therapeutic |
Intended Use
Innovative Technologies' Alginate/Pectin Wound Dressings, are highly conformable, sterile, primary wound dressings intended to provide an environment ideally suited for the management of moderate to heavily exuding partial to full thickness wounds. The non-woven Alginate/Pectin fibre preparation, which includes various sizes of flat dressings and wound packing, reacts with wound exudate to form an integral, gelatinous, mass providing for a moist healing environment. The gel may easily be lifted away from the wound, reducing the potential for delicate peri-wound tissue damage during dressing changes.
Device Story
Non-woven, sterile, primary wound dressing composed of calcium alginate and citrus pectin fibers; intended for moderate to heavily exuding wounds. Applied by clinician to wound site; reacts with wound exudate to form integral gelatinous mass; maintains moist healing environment; facilitates non-traumatic removal by lifting gel away from wound bed. Reduces peri-wound tissue damage during dressing changes. Supplied in various sizes as flat dressings or packing.
Clinical Evidence
Bench testing only. Biocompatibility testing completed including cytotoxicity, haemolysis, acute systemic toxicity, skin irritation, and sensitisation per ISO/Tripartite guidelines.
Technological Characteristics
Material: Calcium Alginate and Citrus Pectin fibers. Form: Non-woven, conformable, flat dressings or packing. Sterilization: Gamma radiation per ANSI/AAMI/ISO11137-1994 and AAMI TIR8-1991; SAL 10^-6.
Indications for Use
Indicated for management of moderate to heavily exuding partial to full thickness wounds, including pressure, venous, diabetic, and arterial ulcers, donor sites, trauma wounds, dermal lesions, incisions, and 1st and 2nd degree burns.
Regulatory Classification
Identification
A liquid bandage is a sterile device that is a liquid, semiliquid, or powder and liquid combination used to cover an opening in the skin or as a dressing for burns. The device is also used as a topical skin protectant.
Predicate Devices
- Kaltostat Alginate Wound Dressings (Convetec)
- Granuflex (Duoderm) Hydrocolloid Dressings (Convetec)
Related Devices
- K980989 — 3M TEGAGEN HI ALGINATE DRESSING & 3M TEGAGEN HG ALGINATE DRESSING · Innovative Technologies , Ltd. · May 27, 1998
- K982638 — 3M TEGAGEN HI ALGINATE DRESSING & 3 M TEGAGEN HG ALGINATE DRESSING, MODEL #'S 90110,90112,90114,90120, 90210,90212,90214 · Innovative Technologies , Ltd. · Aug 21, 1998
- K962421 — SORBCEL ALGINATE/CMC WOUND DRESSING · Innovative Technologies , Ltd. · Jul 11, 1996
- K973509 — KENDALL CURASORB ZN + MG ALGINATE WOUND DRESSING · Kendall Healthcare Products Co. Div.Of Tyco Health · Dec 4, 1997
- K982234 — AL-GEN CALCIUM ALGINATE DRESSING · Innovative Technologies , Ltd. · Sep 8, 1998