K963501 · Lacrimedics, Inc. · LZU · Apr 14, 1997 · OP
Device Facts
Record ID
K963501
Device Name
HERRICK ULTIMATE PUNCTUM PLUG
Applicant
Lacrimedics, Inc.
Product Code
LZU · OP
Decision Date
Apr 14, 1997
Decision
SESE
Submission Type
Traditional
Device Class
Class U
Attributes
Therapeutic
Intended Use
Used to enhance retention of eye fluids by occlusion of one or more punctum, reducing lacrimal flow. Useful for treatment of dry eye and other conditions where enhancement of tear retention is indicated.
Device Story
Flexible silicone ophthalmic device; inserted into punctum by clinician; occludes lacrimal duct; increases tear retention; treats dry eye, post-surgical conditions, and contact lens intolerance; enhances medication efficacy.
Clinical Evidence
No clinical data provided.
Technological Characteristics
Material: Medical grade silicone. Form factor: Flexible plug, 1.8 mm length, 0.8 mm diameter. Function: Mechanical occlusion of the punctum.
Indications for Use
Indicated for patients with dry eye, post-surgical ocular conditions, contact lens-related issues, or as an adjunctive treatment aid to enhance medication efficacy.
Predicate Devices
Tapered-Shaft Punctum Plug (Eagle Vision and Ciba Vision)
{0}
K963501
Lacrimedics, Inc.
190 N. Arrowhead Ave., Suite B • Rialto, CA 92376-9908 USA
APR 14 1997
# 510(k) NOTIFICATION SUMMARY
| Device name: | Herrick Ultimate Punctum Plug |
| --- | --- |
| Device common name: | Punctum Plug |
| Classification name: | Punctum Plug (86LZU) |
| Predicate devices: | Tapered-Shaft Punctum Plug (marketed by Eagle Vision and Ciba Vision) |
| | Soft Plug™ (marketed by Oasis) |
| | Punctum Plug (marketed by FCI Ophthalmics) |
| Device description: | A flexible silicone ophthalmic device designed to fit and be retained in the punctum. |
| Device use: | Used to enhance retention of eye fluids by occlusion of one or more punctum, reducing lacrimal flow. Useful for treatment of dry eye and other conditions where enhancement of tear retention is indicated. |
## Comparison to predicate devices:
| Device name
Features | Herrick Ultimate Punctum Plug (Lacrimedics) | Tapered Shaft Punctum Plug (Eagle Vision & Ciba) | Soft Plug (Oasis) | Punctum Plug (FCI Ophthalmics) |
| --- | --- | --- | --- | --- |
| 1. Indications claimed | Treatment of dry eye
After eye surgery
Enhance efficacy of medications
Treatment of contact lens problems
Adjunctive treatment aid | Treatment of dry eye | Treatment of dry eye
After eye surgery
Enhance efficacy of medications
Treatment of contact lens problems
Adjunctive treatment aid | (Not available) |
| 2. Function | Causes occlusion of punctum, resulting in greater tear retention | Causes occlusion of punctum, resulting in greater tear retention | Causes occlusion of punctum, resulting in greater tear retention | Causes occlusion of punctum, resulting in greater tear retention |
| 3. Design - length | 1.8 mm | 1.8 mm | 2.0 mm | 2.0 mm |
| diameter (typical) | 0.8 mm | 0.7 mm | 0.9 mm | 0.8 mm |
| 4. Material | Medical grade silicone | Medical grade silicone | Medical grade silicone | Medical grade silicone |
**Clinical tests:** None
**Adverse S&E information:** None
*David S. Herrick, MD*
CEO
*David S. Herrick, MD*
3-11-97
Telephone: (800)367-8327 (USA and Canada) • (909)873-3820 • Fax: (909)873-3823
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.