ANTI-TUMOR EXPERT SYSTEM (ATES) FOR RADIATION THERAPY TREATMENT PLANNING

K963451 · Topslane · IYE · Aug 5, 1997 · Radiology

Device Facts

Record IDK963451
Device NameANTI-TUMOR EXPERT SYSTEM (ATES) FOR RADIATION THERAPY TREATMENT PLANNING
ApplicantTopslane
Product CodeIYE · Radiology
Decision DateAug 5, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

The ARTP (Anti-Tumor Radiation Treatment Planning System) is a radiation therapy treatment planning system for radiation dose planning and simulation of patients undergoing external beam treatment in the oncology clinic. ARTP is used to plan and simulate radiation treatments with linear accelerators and other similar teletherapy devices with x-ray energies from 1 to 25MV, as well as Cobalt-60, and electron energies from 1 to 25 MeV.

Device Story

Radiotherapy Treatment Planning System (RTPS) for oncology clinics; performs radiation dose planning and simulation for external beam treatments. Inputs: patient anatomical data for dose calculation. Processing: utilizes photon beam convolution dose calculation algorithm and electron pencil beam dose calculation algorithm. Outputs: radiation treatment plans for linear accelerators, teletherapy devices, and Cobalt-60 units. Used by clinicians to optimize radiation delivery; aids in clinical decision-making for cancer treatment; benefits patients by enabling precise dose distribution planning.

Clinical Evidence

Bench testing only. Performance evaluations confirmed the system consistently performed within design parameters and equivalently to the predicate device.

Technological Characteristics

RTPS software; photon beam convolution and electron pencil beam dose calculation algorithms. Complies with IEC 601-1, IEC 601-1.1, and IEC 878. Designed per FDA Reviewer Guidance for Computer Controlled Medical Devices (1991).

Indications for Use

Indicated for radiation dose planning and simulation of patients undergoing external beam radiation therapy using linear accelerators, teletherapy devices (1-25MV x-ray), Cobalt-60, or electron beams (1-25 MeV) in an oncology clinic setting.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} TOPSLANE K963457 1 of 2 # SUMMARY OF SAFETY AND EFFECTIVENESS AUG - 5 1997 ## 1. Submitter's Information: Dated: August 28, 1996 TOPSLANE 824 Ruth Dr., Pleasant Hill, CA 94523 Contact Person: Ann Zeng VP of Quality Assurance And Regulatory Affairs ## 2. Common or Usual Name: RADIATION TREATMENT PLANNING SYSTEM ATES Proprietary Name: System, Simulation, Radiation Therapy Classification Name: RA (90) KPQ Class 2 21 CFR 892.5050 Version Number: 2.0 ## 3. Predicate Device: Theraplan V05B Radiation Therapy Treatment Planning System K940237 Theratronics ## 4. Description of Device: The TOPSLANE ATES is a Radiotherapy Treatment Planning System (RTPS) for radiation dose planning of patients undergoing external beam treatment in the oncology clinic. ATES is using modem photon beam convolution dose calculation algorithm and electron pencil beam dose calculation algorithm. The system software is designed to be convenient and low cost to the user. ## 5. Statement of intended use: The TOPSLANE product ATES is a Radiation Therapy Treatment Planning System for radiation dose planning of patients undergoing external beam treatment in the oncology clinic. ATES is used to plan radiation treatments with linear accelerators and other similar teletherapy devices with x-ray energies from 1 to 25MV, as well as Cobalt-60, and electron energies from 1 to 25 MeV. The intended use is the same as the predicate device. 824 Ruth Dr., Pleasant Hill, CA94523, USA Tel 510 609 7496 Fax 510 609 7805 {1} TOPSLANE 2 of 2 ## 6. Statement of technological characteristics: The ATES has no significant change in design, materials, energy source or other technological characteristics compared to the predicate device. The intended use and the technological characteristics are the same as the predicate device and therefore we believe it is substantially equivalent to it. ## 7. Differences: The minor configuration differences between the ATES and the predicate device do not alter the intended use or affect the safety and effectiveness of the ATES when used as labeled. ## 8. Special Controls: Although there are no performance standards established by the FDA for these devices, ATES has been designed, and manufactured to meet the following standards: - IEC 601-1: Medical electrical equipment - General requirements for safety - IEC 601-1.1: Safety requirements for medical electrical systems - IEC 878: Graphical symbols for electrical equipment in medical practice The device and its development process also comply with the FDA, CDRH, ODE, August 29, 1991, Reviewer Guidance for Computer Controlled Medical Devices Undergoing 510(k) Review. ## 9. Performance Evaluations: Performance tests were conducted and the results indicated that the system consistently performed within the design parameters and equivalently to the predicate device. THE END 824 Ruth Dr., Pleasant Hill, CA94523, USA Tel 510 609 7496 Fax 510 609 7805 {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Ann Zeng VP of Quality Assurance and Regulatory Affairs TOPSLANE 824 Ruth Drive Pleasant Hill, CA 94523 Re: K963451 Anti-Tumor Radiation Treatment Planning System (ARTP) Dated: June 19, 1997 Received: June 23, 1997 Regulatory Class: II 21 CFR 892.5050/Procode: 90 IYE AUG - 5 1997 Dear Ms. Zeng: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3} 21/07 '97 19:49 86 28 4448145 CDHO1 CN ---> CDRH DRAER 002 Page 1 of 1 510(k) Number (if known): K963451 Device Name: Anti-Tumor Radiation Treatment Planning System (ARTP) Indications For Use: The ARTP (Anti-Tumor Radiation Treatment Planning System) is a radiation therapy treatment planning system for radiation dose planning and simulation of patients undergoing external beam treatment in the oncology clinic. ARTP is used to plan and simulate radiation treatments with linear accelerators and other similar teletherapy devices with x-ray energies from 1 to 25MV, as well as Cobalt-60, and electron energies from 1 to 25 MeV. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use ☑ (Per 21 CFR 801.109) OR Over The Counter Use ☐ (Optional Format 1-2-96) William Yri (Division Sign-Off) Division of Reproductive Abdominal, ENT, and Radiological Devices 510(k) Number K963451
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...