← Product Code [FMQ](/productcode/FMQ) · K963450

# SAFETY VEST/SHOULDER STRAP VEST/ECONOMY SAFETY VEST (K963450)

_Skil-Care Corp. · FMQ · Oct 7, 1996 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K963450

## Device Facts

- **Applicant:** Skil-Care Corp.
- **Product Code:** [FMQ](/productcode/FMQ.md)
- **Decision Date:** Oct 7, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6760
- **Device Class:** Class 1
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

Safety Vest is intended for patients who require restraint while in the wheelchair or bed. Economy Safety Vest is intended for patients who require restraint while in the wheelchair or bed. Vest with Shoulder Straps is intended for patients who require restraint and posture support while in the wheelchair.

## Device Story

Vest-style upper body restraints; constructed from woven polyester or polyester mesh; secured via 1 1/8-inch polyester straps to wheelchair kickspurs or bed frames. Vest with Shoulder Straps includes additional webbing straps for attachment to wheelchair pushhandles. Used in clinical/nursing home settings to provide patient restraint or posture support. Operated by healthcare staff under facility regulations and accreditation standards to ensure appropriate, least-restrictive use. Benefits include patient safety and postural stability.

## Clinical Evidence

No clinical data; bench testing only. Devices have been commercially distributed for approximately 18 years.

## Technological Characteristics

Materials: Woven polyester or polyester mesh; 1 1/8-inch polyester webbing straps. Form factor: Vest-style garment with bias-cut binding. Mechanical restraint mechanism using strap-based securement to wheelchair or bed frames.

## Regulatory Identification

A protective restraint is a device, including but not limited to a wristlet, anklet, vest, mitt, straight jacket, body/limb holder, or other type of strap, that is intended for medical purposes and that limits the patient's movements to the extent necessary for treatment, examination, or protection of the patient or others.

## Predicate Devices

- Safety Vest (Skil-Care Corporation)
- Economy Safety Vest (Skil-Care Corporation)
- Vest with Shoulder Straps (Skil-Care Corporation)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}

Skil-Care CORPORATION K963450 167 Saw Mill River Road, Yonkers, New York 10701

(914) 963-2040

(800) 431-2972

FAX (914) 963-2567

TAB F: 510(k) Summary of Safety and Effectiveness

Name, address, phone and fax numbers for person submitting the 510(k) notification:

Arnold Silverman, President
Skil-Care Corporation
167 Saw Mill River Road
Yonkers, NY 10701
Phone: 1-914-963-2040
Fax: 1-914-963-2567

OCT - 7 1996

Contact person: Arnold Silverman

Date summary was prepared: August 27, 1996

Device name:

Trade name: Safety Vest, Economy Safety Vest, Vest with Shoulder Straps

Common name: Same

Classification name: Wheelchair Accessory

Predicate device:

Safety Vest, Economy Safety Vest, and Vest with Shoulder Straps marketed by Skil-Care Corporation.

Device Description:

Safety Vest: A vest-style, upper body device made from either woven polyester or polyester mesh. A 1 1/8-inch-wide polyester strap attached to one side of the vest at the waist is threaded through a slot on the other causing the vest to criss-cross and close in front of the patient. The waist strap secures to the kickspurs of the wheelchair or to the bed frame. The edges of the garment are finished with a bias cut binding, the color of which corresponds to a size chart included on the package insert. Device is intended for wheelchair or bed use.

Economy Safety Vest: Identical to above except that it is made from a lighter weight woven polyester material and is not available in polyester mesh. Unlike the above device which is intended for multiple uses, this is a single-patient-use device. This fact is noted on the garment label along with a notation that it is not to be laundered.

36

{2}

K963450

Vest with Shoulder Straps: Identical with Safety Vest with the addition of 1 1/8-inch-wide webbing straps sewn to the shoulder area. These straps are secured to the pushhandles to provide patient posture support. This device is for wheelchair use only.

## Indications for use:

Safety Vest is intended for patients who require restraint while in the wheelchair or bed.

Economy Safety Vest is intended for patients who require restraint while in the wheelchair or bed.

Vest with Shoulder Straps is intended for patients who require restraint and posture support while in the wheelchair.

## Comparative information:

The device (devices) used for comparative purposes is (are) currently marketed as described in this submission. Device (devices) is (are): Safety Vest, Economy Safety Vest, Vest with Shoulder Straps.

These devices are currently exempt from 510(k) Premarket Notification Procedures and Good Manufacturing Practice Regulations and are legally marketed by Skil-Care Corporation as of the date of this submission. Skil-Care Corporation has been marketing and commercially distributing these devices for approximately 18 years.

The difference from our currently marketed devices are that the labeling will be changed to incorporate many of the suggestions in FDA’s draft document, "Guidance on the Content of Premarket Notification [510(k)] Submissions for Protective Restraints.

The use of all patient restraints in nursing homes are subject to Health Care Financing Administration’s Regulations which prohibit the use of any restraint, physical or chemical, imposed for the purpose of discipline or convenience. Further, most health care facilities are accredited. HCFA rules governing appropriate use and accreditation standards for device use and personnel training provide the control necessary to ensure that the devices are used correctly. The application of these standards along with public awareness and health care provider training have contributed significantly to ensuring that the least restrictive restraint is used, that restraints are used only when needed for proper medical treatment, and that their use is under appropriate supervision.

37

---

**Source:** [https://fda.innolitics.com/device/K963450](https://fda.innolitics.com/device/K963450)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com