K963389 · American Laser Medical · GEX · Nov 26, 1996 · General, Plastic Surgery
Device Facts
Record ID
K963389
Device Name
FLEX-LASE 600 SERIES ION LASER SYSTEM
Applicant
American Laser Medical
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 26, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Flex-lase 600 Series Ion Laser is intended for use in the treatment of selected vascular and pigmented lesions of the skin. Laser light of a selected wavelength is used to photocalculate tissue based on the absorption characteristic of the targeted chromophore within the tissue.
Device Story
Flex-Lase 600 Series Ion Laser (Model 620 argon; Model 640 krypton) delivers continuous wave laser energy for dermatological procedures. System inputs include electrical power (208 VAC) and water cooling; operator activates via footswitch. Device emits specific wavelengths (488/514 nm for argon; 568/520-530 nm for krypton) delivered via fiber-optic handpiece to target chromophores in skin lesions. Aiming beam provided by adjustable incoherent white light. Used in clinical settings by trained professionals. Output energy photocoagulates targeted vascular or pigmented tissue, facilitating lesion treatment. System is not computer-controlled.
Clinical Evidence
No clinical data provided; substantial equivalence is based on bench testing and comparison of technical specifications and intended use with predicate devices.
Technological Characteristics
Continuous wave ion laser (argon or krypton). Model 620: 488/514 nm; Model 640: 568/520-530 nm. Power output up to 5.5W (argon) or 3.0W (krypton). Beam delivery via fiber-optic handpiece. Cooling: filtered flow-through tap water (2.5 gal/min). Electrical: 208-240 VAC, 40 amp, single phase. Aiming beam: adjustable incoherent white light. Not computer-controlled.
Indications for Use
Indicated for treatment of selected vascular and pigmented skin lesions, including capillary (port wine) hemangioma, strawberry hemangioma, telangiectasia, rosacea, freckles, age spots, cafe-au-lait, and lentigo.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
11/08/96 10:55 *000000000000000 + 301 594 2358
NO. 655 008
NOV 26 1996
K963399
Flax-Lase 600 Series Ion Laser
American Laser Medical
November 7, 1996
# Summary of Safety and Effectiveness Information
## Regulatory Authority:
Safe Medical Devices Act of 1990, 21 CFR 807.92
## Company Name/Contact:
**Company:** American Laser Medical
1832 South 3850 West
Salt Lake City, UT 84104
(801) 972-1311, FAX (801) 972-5251
**Contact:** Mr. Daniel Hoefer
Regulatory Affairs
American Laser Medical
1832 South 3850 West
Salt Lake City, UT 84104
(801) 972-1311, FAX (801) 972-5251
## Name of Device:
**Trade Name:** Flax-Lase 600 Series Ion Laser
**Common Name:** Dermatology Laser
## Classification:
Powered surgical laser instruments are assigned the unique device classification code 79GEX. The published physical description of this device is found in 21 CFR, § 878.4810 (a) (2). This device classification section specifically describes argon laser systems. The American Laser Medical device may be configured as either an argon laser system or as a krypton laser system, however American Laser Medical maintains that there is no substantial difference between argon and krypton lasers for the purpose of classification. Presently, these lasers are considered Class II medical devices.
1
{1}
11/25/96 12:23 *000000000000000000000000000000000000000000000000000000000000000000000000000
NO. 674 D03
Flex-Lase 600 Series Ion Laser
American Laser Medical
November 7, 1996
## Description of Device:
The Flex-Lase 600 Series Ion Laser submission covers all of the internal parts, subassemblies and components as well as the completed device. The Flex-Lase Model 640 is a continuous wave krypton laser system emitting light at the 568 nm (yellow) and 520/530 nm (green) wavelengths. The Flex-Lase Model 620 is a continuous wave argon laser system emitting light at the 488 nm (blue) and the 514 nm (green) wavelengths.
Treatment beam output for the Flex-Lase Model 640 (krypton) is 1.0 Watt (yellow) and 3.0 Watts (yellow-green simultaneous). The power output for the Flex-Lase Model 620 (argon) is 5.5 watts blue-green or 2.0 watts green. The aiming beam is provided by an adjustable intensity incoherent white light source. Exposure times for the Flex-Lase 600 Series Ion Laser (in seconds) are 0.02, 0.05, 0.1, 0.25, 0.5, 1.0, and continuous.
Laser activation occurs by footswitch. Overall weight of the system is 112 lbs. (51 kg). The Flex-Lase 600 Series Ion Laser has a footprint of 16 X 24 inches (38 X 60 cm) and is 36 inches (90 cm) high.
The electrical power requirement is 208 VAC 40 amp single phase. Cooling is provided by external filtered flow-through tap water at a minimum rate of 2.5 gallons (9.5 liters) per minute.
Accessories available for use with the Flex-lase 600 Ion Laser include a focusing handpiece with interchangeable guide tips and/or the Hexascan Mark II. The Flex-laser 600 series is not a computer controlled device.
## Intended Use:
The Flex-lase 600 Series Ion Laser is intended for use in the treatment of selected vascular and pigmented lesions of the skin. Laser light of a selected wavelength is used to photocalculate tissue based on the absorption characteristic of the targeted chromophore within the tissue.
## Indications for Use:
The Flex-Lase 600 Ion Laser is intended for use in the treatment of selected pigmented and vascularized lesions of the skin. Representative indications for use are:
2
{2}
11/25/96 12:24 +000000000000000 - 301 594 2358 NO. 674 D04
Flex-Lase 600 Series Ion Laser
American Laser Medical
November 7, 1996
Flex-Lase Model 640 Laser System
For vascular lesions: (Yellow - 568 nm)
1. Capillary Hemangioma (port wine hemangioma)
2. Strawberry hemangioma
3. Telangectasia
4. Rosacea
For pigmented lesions: (Yellow/Green - 568 nm and 520-530 nm).
1. Freckles
2. Age Spots
3. Cafe-au-lait
4. Lentigo
Flex-Lase Model 620 Laser System
For Vascular Lesions: (Blue/Green - 488 nm and 514.5 nm)
1. Capillary Hemangioma (port wine hemangioma)
2. Strawberry Hemangioma
3. Telangectasia
4. Rosacea
For pigmented lesions: (Green - 514.5 nm)
1. Freckles
2. Age Spots
3. Cafe-au-lait
4. Lentigo
Comparison of Device Characteristics:
For vascular lesions:
The Flex-Lase 600 Ion Laser System is substantially equivalent to the HGM Spectrum K1, the Metalaser Vasculase and the Coherent Artisan already in commercial distribution. The materials, design, intended use, method of manufacture, warnings, cautions, precautions and treatment parameters are substantially the same.
For pigmented lesions:
The Flex-Lase 600 Series Ion Laser system is substantially equivalent to the HGM Spectrum K1, the Metalaser Vasculase and the Coherent Artisan already in commercial distribution. The
3
{3}
11/08/96 10:56 *000000000000000 + 301 594 2358 NO. 655 D11
Flex-Lase 600 Series Ion Laser
American Laser Medical, Inc.
November 7, 1996
materials, design, intended use, method of manufacture, warnings, cautions and precautions are substantially the same.
Feature comparison tables which follow illustrate the equivalence of features of the Flex-Lase 600 and equivalent available systems. See also appendix II for marketing materials on substantially equivalent devices.
4
{4}
11/08/96 10:56 *00000000000000 + 301 594 2356 NO. 655 D12
# Laser Equivalency Table
## Vascular Lesions
| | FLEX-LASE
MODEL 620 | FLEX-LASE
MODEL 640 | HGM K1 | METALASER
VASCULASE | COHERENT
ARTISAN |
| --- | --- | --- | --- | --- | --- |
| Type | Continuous.
wave
argon | Continuous
wave
krypton | Continuous
Wave
krypton | Copper
Vapor
Laser | Continuous
wave
Argon/Dye |
| Wavelength
(nanometers) | 514, or
488/514 | 568, or
568/520-
530 | 568,
520/530 | 511, 578 | 514 only,
or
488/514,
or 577-630
selectable |
| Exposure
Time (sec.) | 0.02,
0.05,
0.1,
0.25,
0.5, 1.0,
and cont. | Same | 0.05, 0.1,
0.2, 0.5,
1.0, 2.0,
5.0, and
continuous | 0.075,
0.1, 0.2,
0.3, and
continuous | 0.01,
0.02,
0.05, 0.1,
0.2, 0.5,
1.0, 2.0,
5.0, cont. |
| Output Mode | Multimode | Multimode | Multimode | Multimode | Multimode |
| Beam
Delivery | Handpiece
via fiber
optic | Handpiece
via fiber
optic | Handpiece
via fiber
optic | Handpiece
via fiber
optic | Handpiece
via fiber
optic |
| Stability
of Output
Energy | +/- 10% | +/- 10% | +/- 10% | +/- 20% | +/- 5% |
| Spot Size
(mm) | 0.25,
0.5, 1.0 | 0.25, 0.5,
1.0 | 0.1, 0.2,
1.0 | 0.1, 0.2,
1.0 | 0.1, 0.25,
0.5, 1.0,
fixed,
variable
0.2-2.5 |
| Energy
density | 2-30
J/cm² | 2-30 J/cm² | 2-30 J/cm² | 4-15 J/cm² | 2-30 J/cm² |
| Aiming
beam | Adjust-
able in-
coherent
white
light
source | Adjustable
incoherent
white
light
source | HeNe
(633nm)
adjustable
to 5 mW | HeNe (633
nm) | HeNe (633
nm) |
5
{5}
11/08/96 10:57 +0000000000000000000000000000000000000000000000000000000000
| PARAMETER | FLEX-LASE MODEL 620 | FLEX-LASE MODEL 640 | HOM K1 | METALASER VASCULASE | COHERENT ARTISAN |
| --- | --- | --- | --- | --- | --- |
| Cooling | Filtered flow-through tap water 2.5 gal/min | Filtered flow-through tap water 2.5 gal/min | Internal liquid cooled | Air-cooled, 40F-80F ambient air temp. | Filtered flow-through tap water 2.5 gal/min |
| Electrical Requirements | 208-240 VAC 40 amp single phase 50/60 Hz | 208-240 VAC 40 amp single phase 50/60 Hz | 208-240 VAC 40 amp 50/60 Hz single phase | 208 VAC +/- 10%, 20 amp 60Hz single phase | 208 VAC 50 amp 50/60 Hz 3 phase |
| Indications for Vascular lesions | Capillary (port wine) hemangioma, Strawberry hemangioma, Telangec-tasia, Rosacea | Capillary or port wine hemangioma Strawberry hemangioma Telangec-tasia Rosacea | Vascular Lesions (same) | Vascular Lesions (same) | Vascular Lesions (same) |
| K Number | | | K913428, K913569 | K903883 | na |
{6}
11/08/96 10:57 +0000000000000000000000000000000000000000000000000000000000
NO.655
Laser Equivalency Table
Pigmented Lesions
| | FLEX-LASE
MODEL 620 | FLEX-LASE
MODEL 640 | HGM K1 | METALASER
VASCULASE | COHERENT
ARTISAN |
| --- | --- | --- | --- | --- | --- |
| Type | Continuous wave
argon | Continuous wave
krypton | Continuous Wave
krypton | Copper Vapor
Laser | Continuous wave
Argon/Dye |
| Wavelength (nanometers) | 514, or
488/514 | 568, or
568/520-530 | 568,
520/530 | 511, 578 | 514 only, or
488/514, or
577-630 selectable |
| Exposure Time (sec.) | 0.02,
0.05,
0.1,
0.25,
0.5, 1.0,
and cont. | Same | 0.05, 0.1,
0.2, 0.5,
1.0, 2.0,
5.0, and
continuous | 0.075,
0.1, 0.2,
0.3, and
continuous | 0.01,
0.02,
0.05, 0.1,
0.2, 0.5,
1.0, 2.0,
5.0, cont. |
| Output Mode | Multimode | Multimode | Multimode | Multimode | Multimode |
| Beam Delivery | Handpiece
via fiber
optic | Handpiece
via fiber
optic | Handpiece
via fiber
optic | Handpiece
via fiber
optic | Handpiece
via fiber
optic |
| Stability of Output Energy | +/- 10% | +/-10% | +/-10% | +/-20% | +/- 5% |
| Spot Size (millimeters) | 0.25,
0.5, 1.0 | 0.25, 0.5,
1.0 | 0.1, 0.2,
1.0 mm | 0.1, 0.2,
1.0 mm | 0.1, 0.25,
0.5, 1.0,
fixed,
variable
0.2-2.5 mm |
| Energy density | 2-30 J/cm² | 2-30 J/cm² | 2-30 J/cm² | 4-15 J/cm² | 2-30 J/cm² |
| Aiming beam | Adjustable incoherent white light source | Adjustable incoherent white light source | HeNe (633nm)
adjustable to 5 mW | HeNe (633 nm) | HeNe (633 nm) |
{7}
11/08/96 10:58 +00000000000000 + 301 594 2358 NO. 655 D15
| PARAMETER | FLEX-LASE
MODEL 620 | FLEX-LASE
MODEL 640 | HGM K1 | METALASER
VASCULASE | COHERENT
ARTISAN |
| --- | --- | --- | --- | --- | --- |
| Cooling | Filtered flow-through tap water 2.5 gal/min | Filtered flow-through tap water 2.5 gal/min | Internal liquid cooled | Air-cooled, 40F-80F ambient air temp. | Filtered flow-through tap water 2.5 gal/min |
| Electrical Requirements | 208-240 VAC 40 amp single phase 50/60 Hz | 208-240 VAC 40 amp single phase 50/60 Hz | 208-240 VAC 40 amp 50/60 Hz single phase | 208 VAC +/- 10%, 20 amp 60Hz single phase | 208 VAC 50 amp 50/60 Hz 3 phase |
| Indications for Pigmented lesions | Freckles, Age spots, cafe-au-lait, lentigo | Freckles, Age spots, cafe-au-lait lentigo | Pigmented Lesions (same) | Pigmented Lesions (same) | Pigmented Lesions (same) |
| K Number | | | K913428, K913569 | K903883 | na |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.