← Product Code [CKF](/productcode/CKF) · K963383

# HICHEM ALP/AMP REAGENT KIT (K963383)

_Hichem Diagnostics · CKF · Sep 23, 1996 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K963383

## Device Facts

- **Applicant:** Hichem Diagnostics
- **Product Code:** [CKF](/productcode/CKF.md)
- **Decision Date:** Sep 23, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1050
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

HiChem ALP/AMP Reagent (product no. 70001) is intended for the quantitative determination of alkaline phosphatase in serum and plasma. The principal diagnostic indications of elevated serum alkaline phosphatase are diseases of the liver, bone, parathyroid and intestine.

## Device Story

Reagent kit for quantitative determination of alkaline phosphatase (ALP) in serum/plasma; enzymatic hydrolysis of p-nitrophenylphosphate to p-nitrophenoxide at alkaline pH; measured via spectrophotometry. Used in clinical laboratories; manual procedure or automated on clinical analyzers (e.g., Hitachi 704). Two liquid-stable components (substrate and buffer) combined before/during use. Output is ALP activity level (U/L). Results assist clinicians in diagnosing liver, bone, parathyroid, and intestinal diseases.

## Clinical Evidence

Bench testing only. Linearity validated up to 900 U/L (manual) and 1,200 U/L (automated). Precision studies (n=30-60) showed low SD. Method comparison (n=83 vs MAS; n=178 vs BMD) showed high correlation (r^2 ≥ 0.9986). Stability validated for working reagent and on-board analyzer use. Anticoagulant interference (heparin, lithium iodoacetate) <2%.

## Technological Characteristics

Enzymatic reagent kit; liquid-stable components (ALP/AMP Substrate and Reagent Buffer). Principle: enzymatic hydrolysis of p-nitrophenylphosphate to p-nitrophenoxide. Reaction temperatures: 30°C and 37°C. Compatible with manual procedures or automated clinical analyzers. No software or electronic components.

## Regulatory Identification

An alkaline phosphatase or isoenzymes test system is a device intended to measure alkaline phosphatase or its isoenzymes (a group of enzymes with similar biological activity) in serum or plasma. Measurements of alkaline phosphatase or its isoenzymes are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.

## Predicate Devices

- BMD Alkaline Phosphatase/AMP Reagent (product no. 704093)
- MAS ALP Reagent (product no. 139-154)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K963383
SEP 23 1996
DIAGNOSTICS

# SUMMARY OF 510(K) SAFETY AND EFFECTIVENESS INFORMATION

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

HiChem ALP/AMP Reagent (product no. 70001) is intended for the quantitative determination of alkaline phosphatase in serum and plasma. The principal diagnostic indications of elevated serum alkaline phosphatase are diseases of the liver, bone, parathyroid and intestine.

The HiChem ALP/AMP Reagent determines alkaline phosphatase by enzymatic hydrolysis of p-nitrophenylphosphate to p-nitrophenoxide at alkaline pH. The HiChem ALP/AMP Reagent is intended to be used either as a manual procedure or on clinical analyzers which can automate the required manipulations. The reagent is supplied as two liquid-stable components which are combined, either before or during use, in the approximate ratio of 1 part ALP/AMP Substrate and 5 parts ALP/AMP Reagent Buffer. The ALP/AMP Substrate can also be used as a start reagent and combined with the Reagent Buffer following sample addition.

The HiChem ALP/AMP Reagent is substantially equivalent to the BMD Alkaline Phosphatase/AMP Reagent, product no. 704093, manufactured by Boehringer Mannheim Corp., Indianapolis, IN., and the MAS ALP Reagent, product no. 139-154 manufactured by Medical Analysis Systems, Inc., Camarillo, CA. All three reagents support the same intended use and produce equivalent results with the same clinical purpose. In addition, they are all based on the same methodology which determines alkaline phosphatase (ALP) through the measurement of liberated p-nitrophenoxide. Finally, all reagents are sold in a generic format which supplies the reagent with a manual procedure and supports its use on various instruments through procedure supplements (application sheets).

The effectiveness of the manual procedure is shown by the recovery of linearity standards, the precision of control recoveries, the comparison of serum and plasma recoveries to the MAS ALP Reagent and the validation of the chemical additives and reconstituted stability claims.

The recovery of alkaline phosphatase using HiChem ALP/AMP Reagent as a manual method at both $30^{\circ}C$ and $37^{\circ}C$ reaction temperatures is linear to at least 900 U/L as shown by the recovery of linearity standards ranging from 0 to over 1,350 U/L. Regression statistics are shown below.

$$
\begin{array}{l}
\text{(Recoveries at } 30^{\circ} \mathrm{C}) = -1.7 \, \mathrm{U/L} + 1.005 \times (\text{Standard Activity}), \quad r^2 = 0.9998, \quad \mathrm{sy.x} = 6.5 \, \mathrm{U/L}. \\
\text{(Recoveries at } 37^{\circ} \mathrm{C}) = 2.9 \, \mathrm{U/L} + 0.969 \times (\text{Standard Activity}), \quad r^2 = 0.9998, \quad \mathrm{sy.x} = 8.0 \, \mathrm{U/L}.
\end{array}
$$

Precision, demonstrated by replicate assay of control sera at $37^{\circ}C$, is shown below.

|  Specimen | n | mean | within-run SD | total SD  |
| --- | --- | --- | --- | --- |
|  Serum control 1 | 30 | 55 U/L | 2.1 U/L | 2.2 U/L  |
|  Serum control 2 | 30 | 225 U/L | 3.7 U/L | 3.9 U/L  |
|  Serum control 3 | 30 | 765 U/L | 6.6 U/L | 9.0 U/L  |

Alkaline phosphatase recoveries of 83 mixed serum and plasma specimens are compared between the HiChem and MAS reagents. Least squares regression statistics are shown below.

$$
\text{(HiChem Results)} = -0.2 \, \mathrm{U/L} + 1.058 \times (\text{MAS Results}) \quad r^2 = 0.9986, \quad \mathrm{sy.x} = 5.5 \, \mathrm{U/L}.
$$

The use of heparin and lithium iodoacetate as anticoagulants is shown by the assay of spiked serum pools. In all cases, the bias due to the addition of anticoagulant is less than $2\%$ and statistically insignificant at a $95\%$ confidence level.

The stability of the combined working reagent over 1 month at $2-8^{\circ}C$ and 3 days at $18-25^{\circ}C$ are documented through the recovery of serum controls which range from approximately 50 to 750 U/L AMP at $37^{\circ}C$. For all controls, the observed shifts in recovery were less than the greater of 3 U/L or $5\%$.

The effectiveness of the automated Hitachi 704 procedure is shown by the recovery of linearity standards, the precision of control recoveries, the recovery of serum controls over both the calibration stability period and the on-board stability claim, and the comparison of patient specimen recoveries to the BMD Alkaline Phosphatase/AMP Reagent.

510(k) Notification, HiChem ALP/AMP Reagent Kit
HiChem Diagnostics, Brea, California
23 August, 1996
Page 53 of 54
HiCHEM DIAGNOSTICS 231 NORTH PUENTE STREET, BREA, CA 92821 TEL 714.871.8360 / 800.733.4424 / FAX 800.840.5237
TECHNOLOGY THAT MAKES A DIFFERENCE

{1}

The recovery of alkaline phosphatase using HiChem ALP/AMP Reagent as an automated method is linear to at least 1,200 U/L as shown by the recovery of ten linearity standards which span the claimed linear range. Regression statistics are shown below.

$$
(HiChem \text{ Recoveries}) = 3.6 \, U/L + 0.981 \times (\text{Activity}), \quad r^2 = 1.000, \quad s_y.x = 3.6 \, U/L.
$$

Precision, demonstrated by replicate assay of commercially available control sera, is shown below.

|  Specimen | n | mean | within-run SD | total SD  |
| --- | --- | --- | --- | --- |
|  Serum control 1 | 60 | 49 U/L | 0.4 U/L | 0.6 U/L  |
|  Serum control 2 | 60 | 196 U/L | 0.8 U/L | 1.1 U/L  |
|  Serum control 3 | 60 | 692 U/L | 2.5 U/L | 3.0 U/L  |

Alkaline phosphatase recoveries of 178 mixed serum and plasma specimens compared between the HiChem and BMD reagents using least squares regression, yield the following statistics.

$$
(HiChem \text{ Results}) = 1.2 \, U/L + 1.974 \times (\text{BMD Results}), \quad r^2 = 1.000, \quad s_y.x = 1.4 \, U/L.
$$

The 24 hour calibration stability claim is documented through the recovery of serum controls which span from approximately 50 to 750 U/L ALP. In all cases, the observed shifts in recoveries over 24 hours without calibration are less than 0.25%. The on-board stability claim of 2 weeks is documented by the recovery of serum controls using the same reagent, left on the analyzer, over 15 days. The largest observed control shift over the 15 day period was only 2%.

The HiChem ALP/AMP Reagent is shown to be safe and effective and substantially equivalent to the BMD Alkaline Phosphatase/AMP Reagent, product no. 704093, manufactured by Boehringer Mannheim Corp., Indianapolis, IN., and the MAS ALP Reagent, product no. 139-154 manufactured by Medical Analysis Systems, Inc., Camarillo, CA.

![img-0.jpeg](img-0.jpeg)

Wynn Stocking
Manager of Regulatory Affairs
HiChem Diagnostics

510(k) Notification, HiChem ALP/AMP Reagent Kit
HiChem Diagnostics, Brea, California
23 August, 1996
Page 54 of 54

---

**Source:** [https://fda.innolitics.com/device/K963383](https://fda.innolitics.com/device/K963383)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
