K963293 · Augustine Medical, Inc. · MSA · Mar 28, 1997 · General, Plastic Surgery
Device Facts
Record ID
K963293
Device Name
AUGUSTINE MEDICAL WOUND CARE SYSTEM MODEL 68XXX
Applicant
Augustine Medical, Inc.
Product Code
MSA · General, Plastic Surgery
Decision Date
Mar 28, 1997
Decision
SN
Submission Type
Traditional
Regulation
21 CFR 878.4020
Device Class
Class 1
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K963293 · Mar 28, 1997
AUGUSTINE MEDICAL WOUND CARE SYSTEM MODEL 68XXX
Augustine Medical, Inc.
Clinical study data; Investigator reports
The sponsor provided clinical data from seven studies comparing the Augustine Medical Wound Care System to standard of care treatments for various ulcer types to demonstrate that the device does not impede healing and performs at least as well as standard therapies.
Wound healing; Standard of care; Clinical performance; Ulcer management
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Seven studies of Augustine Medical Wound Care System; Comparative clinical evaluation against standard of care; Study Period: Prior to March 1997
Patients with diabetic ulcers, venous stasis ulcers, pressure ulcers, and chronic leg ulcers
Standard of care treatments
Median percentage reduction of wound area per day
Indications for Use
The Augustine Medical Wound Care System is intended to be used for the local management of wounds by maintaining moisture and body temperature in the wound bed and is indicated for partial and full thickness wounds, such as, stage I through IV venous, arterial, diabetic, and pressure ulcers.
Device Story
System manages wounds by maintaining moisture and periwound temperature; comprises adhesive Wound Cover, radiant heater card, heater control unit, and power supply. Wound Cover provides gas-permeable barrier, absorbs exudate, and supports heater card; heater card irradiates periwound area with infrared energy. Used in clinical settings by healthcare professionals; prescription-only. Control unit offers temperature settings (38°C/42°C) with safety limits to prevent overheating. Output allows visual assessment via clear window; aids wound healing by maintaining warm, moist environment. Benefits patient by supporting healing process without impeding standard care.
Clinical Evidence
Clinical data from seven studies involving diabetic, venous stasis, pressure, and chronic leg ulcers. Compared Augustine Medical Wound Care System to standard of care. Results showed no adverse effects or extended healing times; median percentage reduction of wound area per day was consistently equal to or higher than standard of care (e.g., 2.30% vs 1.86% for diabetic ulcers; 3.36% vs 3.11% for venous stasis).
Technological Characteristics
System includes adhesive Wound Cover (gas-permeable), radiant heater card, heater control unit, and power supply. Operates via infrared energy to raise periwound temperature (38°C or 42°C). Features temperature safety limits and overtemperature shut-off. Prescription use. Sterile, single-use dressing components.
Indications for Use
Indicated for local management of partial and full thickness wounds, including stage I-IV venous, arterial, diabetic, and pressure ulcers.
Regulatory Classification
Identification
An occlusive wound dressing is a nonresorbable, sterile or non-sterile device intended to cover a wound, to provide or support a moist wound environment, and to allow the exchange of gases such as oxygen and water vapor through the device. It consists of a piece of synthetic polymeric material, such as polyurethane, with or without an adhesive backing. This classification does not include an occlusive wound dressing that contains added drugs such as antimicrobial agents, added biologics such as growth factors, or is composed of materials derived from animal sources.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 27, 2022
Augustine Medical, Inc.
Scott D. Augustine, M.D.
Chief Executive Officer
10393 West 70th Street
Eden Prairie, Minnesota 55344
Re: K963293
Trade/Device Name: Augustine Medical Wound Care System Model 68XXX
Regulation Number: 21 CFR 878.4020
Regulation Name: Occlusive Wound Dressing
Regulatory Class: Class I
Product Code: MSA
Dear Scott D. Augustine, M.D.:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated March 28, 1997.
Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under regulation number, 878.4020.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Julie Morabito, OHT4: Office of Surgical and Infection Control Devices, 240-402-3839, Julie.Morabito@fda.hhs.gov.
Sincerely,
Julie A. Morabito -S
Julie Morabito, Ph.D.
Assistant Director
DHT4B: Division of Infection Control and Plastic Reconstructive Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
MAR 28 1997
Scott D. Augustine, MD
CEO
Augustine Medical, Inc.
10393 West 70th Street
Eden Prairie, Minnesota 55344
Re: K963293
Augustine Medical Wound Care System, Model 68XXX
Regulatory Class: Unclassified
Product Code: MSA
Dated: February 14, 1997
Received: February 18, 1997
Dear Dr. Augustine:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market your device subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (Act) and the following limitations:
1. This device may not be labeled for use on third degree burns.
2. This device may not be labeled as having any accelerating effect on the rate of wound healing or epithelization.
3. This device may not be labeled as a long-term, permanent, or no-change dressing, or as an artificial (synthetic) skin.
4. This device may not be labeled as a treatment or a cure for any type of wound.
{3}
Page 2 - Dr. Scott D. Augustine
The labeling claims listed above would be considered a major modification in the intended use of the device and would require a premarket notification submission (21 CFR 807.81). The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations (CFR), Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practices (GMP) for Medical Devices: General GMP regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note
2
{4}
Page 3 - Dr. Scott D. Augustine
the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or 301-443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Celia M. Witten, Ph.D., M.D.
Director
Division of General and Restorative Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}
# Indications for Use
510(k) number: K963293
Device name: Augustine Medical (Trade Name) Wound Care System
Model 68XXX
Indications for use:
The Augustine Medical Wound Care System is intended to be used for the local management of wounds by maintaining moisture and body temperature in the wound bed and is indicated for partial and full thickness wounds, such as, stage I through IV venous, arterial, diabetic, and pressure ulcers.
PLEASE DO NOT WRITE BELOW THIS LINE—CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☑ or
(Per 21 CFR 801-109)
or
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K963293
Over the Counter Use
Augustine Medical, Inc.
iv
3/18/97
{6}
510 (K) ROUTE SLIP
OK
510 (k) NUMBER K963293 PANEL SU DIVISION DGRD BRANCH PRSB
TRADE NAME AUGUSTINE MEDICAL WOUND CARE SYSTEM MODEL 68XXX
COMMON NAME AUGUSTINE MEDICAL WOUND CARE SYSTEM
PRODUCT CODE MGP DRESSING, WOUND AND BURN, OCCLUSIVE
APPLICANT AUGUSTINE MEDICAL, INC.
SHORT NAME AUGUMEDI
CONTACT SCOTT D AUGUSTINE
DIVISION
ADDRESS 10393 WEST 70TH ST.
EDEN PRAIRIE, MN 55344
PHONE NO. (612) 947-1200 FAX NO. (612) 947-1400
MANUFACTURER AUGUSTINE MEDICAL, INC. REGISTRATION NO. 2183725
STERIGENICS INTL., INC. 1420032
STERIGENICS INTL., INC. 1058584
ISOMEDIX OPERATIONS, INC. 1450662
DATE ON SUBMISSION 19-AUG-96 DATE DUE TO 510 (K) STAFF 04-NOV-96
DATE RECEIVED IN ODE 21-AUG-96 DATE DECISION DUE 19-NOV-96
DECISION __ DECISION DATE __
| SUPPLEMENTS | SUBMITTED | RECEIVED | DUE POS | DUE | OUT |
| --- | --- | --- | --- | --- | --- |
| S001 | 21-NOV-96 | 22-NOV-96 | 05-FEB-97 | 20-FEB-97 | 31-JAN-97 |
| S002 | 14-FEB-97 | 18-FEB-97 | 04-MAY-97 | 19-MAY-97 | |
CORRESPONDENCE SENT DUE BACK
C001 05-NOV-96 05-DEC-96 HOLD LETTER
C002 31-JAN-97 02-MAR-97 HOLD LETTER
OTHER
SUBMISSIONS SUBMITTED RECEIVED DUE POS DUE OUT
ADD-TO-FILE 02-OCT-96 02-OCT-96 01-DEC-96
Is this 510 (k) identified as a Class III device ☐ YES ☑ NO
{7}
Scott D. Augustine, MD
CEO
Augustine Medical, Inc.
10393 West 70th Street
Eden Prairie, Minnesota 55344
Re: K963293
Augustine Medical Wound Care System, Model 68XXX
Regulatory Class: Unclassified
Product Code: MSA
Dated: February 14, 1997
Received: February 18, 1997
Dear Dr. Augustine:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market your device subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (Act) and the following limitations:
1. This device may not be labeled for use on third degree burns.
2. This device may not be labeled as having any accelerating effect on the rate of wound healing or epithelization.
3. This device may not be labeled as a long-term, permanent, or no-change dressing, or as an artificial (synthetic) skin.
4. This device may not be labeled as a treatment or a cure for any type of wound.
{8}
Page 2 - Dr. Scott D. Augustine
The labeling claims listed above would be considered a major modification in the intended use of the device and would require a premarket notification submission (21 CFR 807.81). The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations (CFR), Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practices (GMP) for Medical Devices: General GMP regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the *Federal Register*. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note
{9}
Page 3 - Dr. Scott D. Augustine
the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or 301-443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Celia M. Witten, Ph.D., M.D.
Director
Division of General and Restorative Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
FILE
SMPY
| OFFICE | SURNAME | DATE | OFFICE | SURNAME | DATE |
| --- | --- | --- | --- | --- | --- |
| HF2-410 | Knaus | 3/23/97 | | | |
| Z-410 | S. Rhodes | 3/27/97 | | | |
| Z410 | Oel | 3/28/97 | | | |
{10}
bcc: HFZ-401 DMC
HFZ-403 RChissler
HFZ-410 DGRD
D.O.
d/t:
f/t:HFZ-410:DXK:BJT:3/25/97
a
{11}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food And Drug Administration
March 21, 1997
From: Reviewer(s) - Name(s) David Krause
Memorandum
Subject: 510(k) Number K963293/52
To: The Record - It is my recommendation that the subject 510(k) Notification:
☐ Refused to accept.
☐ Requires additional information (other than refuse to accept).
☐ Accepted for review
☑ Is substantially equivalent to marketed devices. Use K-27 SN letter for wound dressings
☐ NOT substantially equivalent to marketed devices.
☐ Other (e.g., exempt by regulation, not a device, duplicate, etc.)
Is this device subject to Postmarket Surveillance?
☐ YES ☑ NO
Is this device subject to the Tracking Regulation?
☐ YES ☑ NO
Was clinical data necessary to support the review of this 510(k)?
☑ YES ☐ NO
Is this a prescription device?
☑ YES ☐ NO
Was this 510(k) reviewed by a Third Party?
☐ YES ☑ NO
This 510(k) contains:
Truthful and Accurate Statement ☐ Requested ☑ Enclosed
(required for originals received 3-14-95 and after)
☑ A 510(k) summary OR ☐ A 510(k) statement
☐ The required certification and summary for class III devices
☑ The indication for use form (required for originals received 1-1-96 and after)
The submitter requests under 21 CFR 807.95 (doesn't apply for SEs):
☐ No Confidentiality ☐ Confidentiality for 90 days ☐ Continued Confidentiality exceeding 90 days
Predicate Product Code with panel and class: Additional Product Code(s) with panel (optional):
79 MSA - Dressing, Wood & Burn, Heated Occlusive
Unclassified
Heated Occlusive
Review: (Branch Chief) 79 MGP Occlusive Wood & Burn Dressing
(Dr. Rhoads) (Branch Code) 3/27/97
(Date)
Final Review: (Division Director) 3/27/97
(Date)
Revised: 11-20-96
{12}
# 510(k) "SUBSTANTIAL EQUIVALENCE" DECISION-MAKING PROCESS (DETAILED)

- 510(k) submissions compare new devices to marketed devices. FDA requests additional information if the relationship between marketed and "predicate" (pre-Amendments or reclassified post-Amendments) devices is unclear.
- This decision is normally based on descriptive information alone, but limited testing information is sometimes required.
- Data may be in the 510(k), other 510(k)s, the Center's classification files, or the literature.
{13}
REVISED:3/14/95
"SUBSTANTIAL EQUIVALENCE" (SE) DECISION MAKING DOCUMENTATION
# K963293
Reviewer: David Krause, Biologist
Division/Branch: DGRD/PRSB
Device Name: Augustine Medical Wound Care System, Model 68XXX
Product To Which Compared [510(K) Number If Known]: ClearSite Wound Dressing (K914207 & K920677), Lyofoam Wound Dressing (K860085) and Seabrook Micro Temp Pump and Pads (K843146).
YES NO
| 1. Is Product A Device | X | | If NO = Stop |
| --- | --- | --- | --- |
| 2. Is Device Subject To 510(k)? | X | | If NO = Stop |
| 3. Same Indication Statement? | X | | If YES = Go To 5 |
| 4. Do Differences Alter The Effect Or Raise New Issues of Safety Or Effectiveness? | NA | | If YES = Stop NE |
| 5. Same Technological Characteristics? | | X | If YES = Go To 7 |
| 6. Could The New Characteristics Affect Safety Or Effectiveness? | X | | If YES = Go To 8 |
| 7. Descriptive Characteristics Precise Enough? | NA | | If NO = Go To 10
If YES = Stop SE |
| 8. New Types Of Safety Or Effectiveness Questions? | | X | If YES = Stop NE |
| 9. Accepted Scientific Methods Exist? | X | | If NO = Stop NE |
| 10. Performance Data Available? | X | | If NO = Request Data |
| 11. Data Demonstrate Equivalence? | X | | Final Decision: SE |
1. Intended Use: The subject device is intended to be used for the local management of wounds by maintaining moisture and body temperature in the wound bed and is indicated for partial and full thickness wounds, such as stage I through IV venous, arterial, diabetic, and pressure ulcers. When used in combination with the Wound Cover and Heater Control Unit, the Heater Card is designed to provide heat to the periwound area. The Wound Cover is intended to provide a gas permeable barrier which will absorb wound exudate and maintain a warm and moist periwound environment.
2. Device Description: This wound care device (no trade name has been assigned yet) is a wound care treatment system comprised of an adhesive bandage, radiant heater (warming insert), and a heater controller. The system is designed to irradiate a cutaneous ulceration or other wound with infrared energy to raise the periwound temperature. The Augustine Medical Wound Care System is comprised of: a heater control unit (two heater control units are available for the application of temperatures at 38°C and 42°C); a heater card; a sterile/bandage called a Wound Cover which supports the heater card and protects the wound (the wound cover is designed with a clear or opaque window which is elevated from the wound); and a power supply/battery charger. The top of the bandage has a pocket
{14}
that accepts the warming insert. The warming insert is connected to the heater controller via a 48.0 inch long detachable electrical cord. The device is considered to be a portable, hand held device.
# EXPLANATIONS TO "YES" AND "NO" ANSWERS TO QUESTIONS ON PAGE 1 AS NEEDED
1. Explain why not a device: The Augustine Medical Wound Care System is a medical device.
2. Explain why not subject to 510(k): This device is subject to 510(k).
3. How does the new indication differ from the predicate device's indication: The indications are generally the same.
4. Explain why there is or is not a new effect or safety or effectiveness issue: N/A.
5. Describe the new technological characteristics: This device has the capability to generate heat at the wound site.
6. Explain how new characteristics could or could not affect safety or effectiveness: The heat generated could lead to problems with healing of the wound.
7. Explain how descriptive characteristics are not precise enough: N/A.
8. Explain new types of safety or effectiveness questions raised or why the questions are not new: The ultimate safety and effectiveness question is will the wound heal.
9. Explain why existing scientific methods can not be used: Existing methods can be used.
10. Explain what performance data is needed: Clinical work which compares healing of wounds when this device is used and when standard of care treatments are used.
11. Explain how the performance data demonstrates that the device is or is not substantially equivalent: Healing rates were not slower when this device was applied.
ATTACH ADDITIONAL SUPPORTING INFORMATION HERE
{15}
MEMO TO THE RECORD
510(K) REVIEW
K963293/S2
DATE: March 26, 1997
FROM: Dr. David Krause, Biologist/Reviewer
OFFICE: HFZ-410
DIVISION: DGRD/PRSB
DEVICE NAME: Augustine Medical Wound Care System, Model 68XXX
COMPANY NAME and ADDRESS: Augustine Medical, Inc., 10393 West 70th Street, Eden Prairie, MN 55344
**Background:** Augustine Medical, Inc. submitted the original 510(k) for this device on August 19, 1996. In a letter dated November 5, 1996, FDA asked for additional information in the form of 7 deficiencies. In a supplement to the original 510(k), dated November 21, 1996, Augustine Medical responded to these deficiencies. All of the questions were adequately addressed except for one. In a letter dated January 31, 1997, FDA responded to K963293/S001. In that response FDA asked Augustine Medical to respond to the following statement: Additional Information is required. This information needs to be in the form of data which evaluate whether or not the device establishes an environment in which healing can take place in the same amount of time as the standard wound care therapies available to patients. Data should be supplied for each type of wound that the sponsor places on his "Indications for Use" form. This supplement (K963293/S002) is Augustine Medical’s reply to that statement.
**Intended Use:** The subject device is intended to be used for the local management of wounds by maintaining moisture and body temperature in the wound bed and is indicated for partial and full thickness wounds, such as stage I through IV venous, arterial, diabetic, and pressure ulcers. When used in combination with the Wound Cover and Heater Control Unit, the Heater Card is designed to provide heat to the periwound area. The Wound Cover is intended to provide a gas permeable barrier which will absorb wound exudate and maintain a warm and moist periwound environment. Both the ClearSite Wound Dressing and the Lyofoam Wound Dressing predicates have almost identical intended uses, except for the heating of the periwound area. The Seabrook device has the intended use of localized heat therapy which is similar to heating the periwound area, although this device is not for open wounds.
**Device Description:** This wound care device (no trade name has been assigned yet) is a wound care treatment system comprised of an adhesive bandage, radiant heater (warming insert), and a heater controller. The system is designed to irradiate a cutaneous ulceration or other wound with infrared energy to raise the periwound temperature. The Augustine Medical
1
{16}
Wound Care System is comprised of: a heater control unit (two heater control units are available for the application of temperatures at 38°C and 42°C); a heater card; a sterile/bandage called a Wound Cover which supports the heater card and protects the wound (the wound cover is designed with a clear or opaque window which is elevated from the wound); and a power supply/battery charger. The top of the bandage has a pocket that accepts the warming insert. The warming insert is connected to the heater controller via a 48.0 inch long detachable electrical cord. The device is considered to be a portable, hand held device.
**Review of Supplement K963293/S002:** The deficiency to which Augustine Medical is responding is included in *italics*, Augustine Medical’s response will be summarized in **bold print**, while this reviewer’s comments will follow in normal type.
*Additional Information is required. This information needs to be in the form of data which evaluate whether or not the device establishes an environment in which healing can take place in the same amount of time as the standard wound care therapies available to patients. Data should be supplied for each type of wound that the sponsor places on his "Indications for Use" form.*
The sponsor states that, to date, there are seven studies completed or in progress involving different wound types being treated with the Augustine Medical Wound Care System. During all of these studies there have been no reports of any adverse effects or extended healing times when using the Augustine Medical Wound Care System. For Diabetic Ulcers, the median percentage reduction of the wound area per day was 1.86% for standard of care treatment and 2.30% for the Augustine Medical Wound Care Device. For Venous Stasis Ulcers, these values were 3.11% and 3.36%, respectively, for Pressure the values were 1.40% in one study and 2.38% in a second study, while it was 4.85% for the Augustine Device, and finally for Chronic Leg Ulcers the percentage of reduction of wound area was 2.00% for standard of care and 2.65% for the Augustine Wound Care Device.
The sponsor includes a table which shows the data for these studies. The sponsor also includes letters from 4 investigators who are using or have used the Augustine Medical Wound Care System in studies. All of these letters indicate that the Augustine Medical Wound Care System does not impede the healing process and may, indeed, augment healing to some degree. This is an adequate response.
**Substantial Equivalence Rationale:** When this 510(k) originally came to FDA, on August 19, 1996, there were questions as to whether or not this device would fit the product description of a wound and burn dressing. The fact that the device was heated also raised the possibility that the device would need to be found "Not Substantially Equivalent" because of new types of safety and effectiveness questions. Finally, FDA also needed to know whether
2
{17}
or not the device actually allowed wounds to heal or did it, in fact, inhibit the healing process. All of these issues were raised with the sponsor in a series of telephone and letter requests for additional information.
The Intended Use of the device is equivalent to a wound and burn dressing. As stated on the "Indications for Use" Form: "The Augustine Medical Wound Care System is intended to be used for the local management of wounds by maintaining moisture and body temperature in the wound bed and is indicated for partial and full thickness wounds, such as, stage I through IV venous, arterial, diabetic, and pressure ulcers." Both the ClearSite and Lyofoam Wound Dressings include wound management in the intended uses and specify pressure ulcers (stages I to IV) as indications. The added indications (other types of ulcers) for the Augustine Medical Wound Care System are supported by clinical data which indicate that the wound dressing does, indeed, allow healing to take place at least as well as "standard of care" for these types of ulcers.
The question of heating the periwound area does not raise new questions of safety and effectiveness because the device never raises the temperature of the wound bed above "normal body temperature" which is 37-38°C. Further, the sponsors of the Lyofoam wound dressing included the following claim in their 510(k): "maintains wound temperature to optimize wound healing." The device contains two heater cards, one at 39°C and one at 42°C. The 42°C heater card is not indicated for leg wounds where compromised circulation may be involved, in fact, there is a warning against using the 42°C heater card for these leg wounds. Also, the device is prescription and comes with adequate instructions for use which will assure that the device is only applied by a health care professional.
This device and the additional information which was sent to FDA in response to telephone and letter requests is reviewed in detail in previous memos.
**Product Code Update:** I received a new product code for this device from Bill Huff on March 18, 1997. The new product code is 79 MSA, Dressing, Wound and Burn, Occlusive, Heated. Please see memos dated March 17 and March 18 for details.
**Recommendation:** Based on my review of the information included in the original 510(k) submission and the subsequent supplements, I recommend that the subject device, The Augustine Medical Wound Care System, is substantially equivalent to the listed legally marketed predicates (see original review).
3
14
{18}
Classification: 79 MSA, Dressing, Wound and Burn, Occlusive, Heated (Unclassified)
David Krause, Ph.D.
(B) (Date)
Biologist/Reviewer,
Division of General and Restorative Devices
Plastic and Reconstructive Surgery Branch
4
{19}
# Indications for Use
510(k) number (if known):
Device name: The Augustine Medical (Trade Name) Wound Care System
Indications for use:
The Augustine Medical Wound Care System is intended to be used for wound care.
PLEASE DO NOT WRITE BELOW THIS LINE—CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ________ or Over the Counter Use ________
(Per 21 CFR 801-109)
Augustine Medical, Inc.
iv
{20}
K963293
# 510(k) Summary
## Augustine Medical Wound Care System
MAR 28 1997
## General Information
| Classification | Unclassified |
| --- | --- |
| Trade Name | Augustine Medical Wound Care System |
| Submitter | Augustine Medical, Inc.
10993 West 70th Street
Eden Prairie, MN 55344
(612) 947-1200 |
| Contact | Scott D. Augustine, MD
CEO |
## Predicate Devices
The Augustine Medical Wound Care System is substantially equivalent in safety and effectiveness in its intended use to these predicate devices:
**ClearSite Wound Dressing**
K914207 & K920677
Manufactured by NDM Corporation, 3040 E. River Road, Dayton, Ohio 45439-1436, (800) 783-1767
**Lvofoam Wound Dressing**
K860085
Manufactured by Acme United Corporation, Medical Division, 75 Kings Highway Cutoff, Fairfield, CT 06430-5340, (800) 243-9852
**Seabrook MicroTemp Pump and Pads**
K843146
Manufactured by Seabrook, Inc. 4043 McMann Road, Cincinnati, OH 45245, (800) 477-7757
Augustine Medical, Inc.
3/24/97
{21}
# Device Description
The wound care system is comprised of a power supply/battery charger, heater control unit, heater card, and a sterile dressing/bandage called a Wound Cover which supports the heater card and holds it away from the wound and skin.
The Wound Cover is a disposable, single-use, sterile, non wound-contact dressing. The Wound Cover may be used either as a stand-alone wound dressing or as part of the Augustine Medical Wound Care System. The Wound Cover incorporates a clear or opaque window. The clear window allows the caregiver to view and assess the tissue. The Wound Cover surrounds and protects the wound, absorbs wound exudate and allows active warming of the periwound area by the application of radiant heat.
The warming components of the system include a thin heater card, a control unit, and a power supply/battery charger. The system operates on standard wall outlet electricity or the self-contained batteries. The heater card is sized appropriately to the Wound Cover and attaches easily over the clear window of the Wound Cover. The heater card does not contact the wound and is intended for single patient use. The control unit incorporates temperature safety limits.
# Intended Use
The Wound Care System is intended to be used for the local management of wounds by maintaining moisture and body temperature in the wound bed and is indicated for partial and full thickness wounds, such as, Stage I through IV venous, arterial, diabetic, and pressure ulcers.
# Testing
Biocompatibility testing was performed on the materials used in the construction of the sterile dressing. All materials passed biocompatibility testing and are suitable for this application.
Physical testing of the product included: dimensional inspection, functional performance, electrical safety, electromagnetic compatibility, pressure switch activation, overtemperature safety shut-off, and performance under clinical conditions. All testing of the product yielded acceptable results.
Augustine Medical, Inc.
2
3/18/97
{22}
Augustine Medical, Inc.
3
# Summary of Substantial Equivalence
The Augustine Wound Care System components are constructed of the same or substantially equivalent materials as the predicate products. The sizes and configurations available along with the packaging and sterilization methods are also equivalent.
The clinical indications for use are substantially equivalent to the predicate devices as well.
Therefore, because of the similarity of materials to other wound treatment products, the test results, and the similar indications for use of the predicate devices, Augustine Medical believes these products do not raise any new safety or effectiveness issues.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.