ELECSYS CALCHECK CK-MB
Device Facts
| Record ID | K963182 |
|---|---|
| Device Name | ELECSYS CALCHECK CK-MB |
| Applicant | Boehringer Mannheim Corp. |
| Product Code | JJX · Clinical Chemistry |
| Decision Date | Aug 23, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1660 |
| Device Class | Class 1 |
Intended Use
Elecsys CalCheck CK-MB is intended for use in the quantitative verification of the calibration curve established by the Elecsys CK-MB reagents and calibrators on Elecsys 1010 or 2010 immunoassay analyzers.
Device Story
Elecsys CalCheck CK-MB consists of three-level lyophilized, human-based materials; used for calibration verification of CK-MB assays on Elecsys 1010/2010 immunoassay analyzers. User reconstitutes material, performs triplicate assay, and compares results against target values to verify calibration curve integrity. Used in clinical laboratory settings by laboratory personnel. Provides quality control verification to ensure accurate patient test results.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Three-level lyophilized human-based material; single analyte (CK-MB) control; intended for use on Elecsys 1010/2010 immunoassay analyzers.
Indications for Use
Indicated for quantitative verification of calibration curves for CK-MB assays on Elecsys 1010 or 2010 immunoassay analyzers.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
- Tosoh Medic AIA-Pack HCG Calibration Verification Test Set (K924862)
Related Devices
- K133284 — ELECSYS CK-MB CALCHECK 5 · Roche Diagnostics · Nov 22, 2013
- K093582 — ELECSYS CK-MB CALCHECK 5 · Roche Diagnostics · Mar 29, 2010
- K963143 — ELECSYS CALCHECK TROPONIN T · Boehringer Mannheim Corp. · Aug 23, 1996