TRUZONE PEAK FLOW METER
Device Facts
| Record ID | K963089 |
|---|---|
| Device Name | TRUZONE PEAK FLOW METER |
| Applicant | Monaghan Medical Corp. |
| Product Code | BZH · Anesthesiology |
| Decision Date | Nov 6, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 868.1860 |
| Device Class | Class 2 |
Intended Use
The TruZone Peak Flow Meter is a simple mechanical device that responds to and indicates the Peak Expiratory Flow Rate (PEFR) generated by the user during a forced exhalation maneuver.
Device Story
Mechanical device; measures Peak Expiratory Flow Rate (PEFR). User places inlet in mouth; performs forced exhalation from full lungs. Internal piston extends spring via pressure spike; pushes indicator to maximum displacement position. Indicator remains at peak value; user reads PEFR in liters per minute from logarithmic scale. Used by patients for self-monitoring; provides objective data for respiratory assessment.
Clinical Evidence
Bench testing only; device performance verified against November 1994 ATS Standards.
Technological Characteristics
Mechanical peak flow meter; logarithmic scale for increased resolution at low flow rates. Piston-spring mechanism. No electronic components or software.
Indications for Use
Indicated for patients requiring measurement of Peak Expiratory Flow Rate (PEFR) via forced exhalation maneuver to monitor respiratory function.
Regulatory Classification
Identification
A peak-flow meter for spirometry is a device used to measure a patient's maximum ventilatory flow rate.
Predicate Devices
- TruZone Peak Flow Meter (K926307)
- TruZone Peak Flow Meter (K955262)
Related Devices
- K955234 — TRUPEAK PEAK FLOW METER · Monaghan Medical Corp. · Mar 18, 1996
- K963095 — TRUPEAK PEAK FLOW METER · Monaghan Medical Corp. · Nov 4, 1996
- K955262 — TRUZONE PEAK FLOW METER · Monaghan Medical Corp. · Mar 7, 1996