← Product Code [HKX](/productcode/HKX) · K963079

# CANON TX-10 TONOMETER (K963079)

_Canon USA, Inc. · HKX · Mar 3, 1997 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K963079

## Device Facts

- **Applicant:** Canon USA, Inc.
- **Product Code:** [HKX](/productcode/HKX.md)
- **Decision Date:** Mar 3, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.1930
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The TX-10 tonometer is intended to be used for the measurement of intraocular pressure of the human eye.

## Device Story

TX-10 is a non-contact tonometer using air-puff technology to measure intraocular pressure. Device features auto-alignment system for measurement; includes auto and manual alignment modes. Operated by clinician in clinical setting. System includes power-assisted chin rest and 4-inch color LCD display. Output consists of intraocular pressure measurements, date, time, and patient ID. Device provides automated measurement, reducing need for special operator skills. Safety mechanism includes software-controlled stopper.

## Clinical Evidence

Bench testing comparing TX-10 and T-2 performance. TX-10 demonstrated regression slope of 0.996, Y-intercept of 0.462 mmHg, and standard deviation of difference (Sd) of 1.56 mmHg. Results meet proposed ISO specifications (slope 0.95-1.05; intercept -1.0 to +1.0; Sd < 2.5 mmHg).

## Technological Characteristics

Non-contact air-puff tonometer. Features auto-alignment system, power-assisted chin rest, and 4-inch color LCD. Software-controlled safety stopper. Measuring range 0-60 mmHg; 1 mmHg increment. Measuring time 3 milliseconds. LED fixation target.

## Regulatory Identification

A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.

## Predicate Devices

- Canon T-2 Tonometer ([K943939](/device/K943939.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

03/03/97 16:58 FAX 316 328 5.52
104 10000000
Canon
K963079
MAR - 3 1997
CANON U.S.A., INC.
ONE CANON PLAZA
LAKE SUCCESS, NY 11042-1198
TELEPHONE: (516) 328-5000
GENERAL FAX TELEPHONE
(516) 328-5069

# 510(k) Summary

1. Date Prepared: July 23, 1996

2. Applicant's Name: Canon USA, Inc.
One Canon Plaza
Lake Success, NY 11042
U.S.A.

Mr. Glenn Impal / Mr. Ken Shadoff
Phone: (516) 328-5600
Fax: (516) 328-5169

3. Model Name: Canon TX-10 Tonometer
86HKX, Tonometer, AC Powered
Document Control Number: ?

4. Predicate Model: Canon T-2 Tonometer
86HKX, Tonometer, AC Powered
Document Control Number: K943939

5. Device Description: The TX-10 is a tonometer designed using a non-contact measurement system. Air puff gently measures the intraocular pressure with the help of an auto-alignment system. Features quick and easy measurement that does not require any special skills.

6. Intended Use: The TX-10 tonometer is intended to be used for the measurement of intraocular pressure of the human eye.

7. Summary of Technological Characteristics: See the attached comparison chart.

8. Assessment of Performance Data: See Attachment 1

{1}

Comparison in the measurement accuracy among Canon TX-10 and Canon T-2

|   | TX-10 | T-2 | ISO Specifications (proposed)  |
| --- | --- | --- | --- |
|  Regression Slope | 0.996 | 0.907 | 0.95 to 1.05  |
|  Regression Y intercept | 0.462mmHg | 1.08mmHg | -1.0 to +1.0  |
|  Standard Deviation of Difference, Sd | 1.56mmHg | 2.12mmHg | 2.5mmHg  |
|  Coefficient of Correlation | 0.926 | 0.858 | Unspecified  |

{2}

K963079

# CANON TONOMETERS COMPARISON CHART

|  MODEL |   | CANON NON CONTACT TONOMETER T-2 | CANON TONOMETER TX-10  |
| --- | --- | --- | --- |
|  Specification | Measuring Range Setting | 0~30/0~60mmHg
(Automatic) | Same as T-2  |
|   |  Alignment Method | One bright dot with alignment level display | **Auto Mode**
Monitor and auto fine alignment  |
|   |   |   |  **Manual Mode**
Monitor with two dots  |
|   |  Increment | 1 mmHg | Same as T-2  |
|   |  Measuring Time | 3 millisecond
(0.003 second) | Same as T-2  |
|   |  Fixation Target | LED (green) | Same as T-2  |
|   |  Chin Rest | Manual | Power assisted
(driving at controller)  |
|  Safety Mechanism |   | Mechanical Stopper | Software Controlled Stopper  |
|  Printout Format |   | R/L, date, time, avarage, ID, 3 or 10 data for each eye | Same as T-2  |
|  Display |   | 5" (inch) B/W CRT | 4" (inch) Color LCD  |
|  Dimentions
(width x depth x height) mm | 286 x 445 x 472 | **Body**
299 x 340 x 177  |   |
|   |   |  **Controller**
170 x 183 x 177  |   |
|  Weight (kg) | 22.0 | **Body**
16.0  |   |
|   |   |  **Controller**
1.5  |   |

---

**Source:** [https://fda.innolitics.com/device/K963079](https://fda.innolitics.com/device/K963079)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
