CANON TX-10 TONOMETER

K963079 · Canon USA, Inc. · HKX · Mar 3, 1997 · Ophthalmic

Device Facts

Record IDK963079
Device NameCANON TX-10 TONOMETER
ApplicantCanon USA, Inc.
Product CodeHKX · Ophthalmic
Decision DateMar 3, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.1930
Device ClassClass 2

Indications for Use

The TX-10 tonometer is intended to be used for the measurement of intraocular pressure of the human eye.

Device Story

TX-10 is a non-contact tonometer using air-puff technology to measure intraocular pressure. Device features auto-alignment system for measurement; includes auto and manual alignment modes. Operated by clinician in clinical setting. System includes power-assisted chin rest and 4-inch color LCD display. Output consists of intraocular pressure measurements, date, time, and patient ID. Device provides automated measurement, reducing need for special operator skills. Safety mechanism includes software-controlled stopper.

Clinical Evidence

Bench testing comparing TX-10 and T-2 performance. TX-10 demonstrated regression slope of 0.996, Y-intercept of 0.462 mmHg, and standard deviation of difference (Sd) of 1.56 mmHg. Results meet proposed ISO specifications (slope 0.95-1.05; intercept -1.0 to +1.0; Sd < 2.5 mmHg).

Technological Characteristics

Non-contact air-puff tonometer. Features auto-alignment system, power-assisted chin rest, and 4-inch color LCD. Software-controlled safety stopper. Measuring range 0-60 mmHg; 1 mmHg increment. Measuring time 3 milliseconds. LED fixation target.

Indications for Use

Indicated for measurement of intraocular pressure in human eyes.

Regulatory Classification

Identification

A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.

Predicate Devices

Submission Summary (Full Text)

{1} 03/03/97 16:58 FAX 316 328 5.52 104 10000000 Canon K963079 MAR - 3 1997 CANON U.S.A., INC. ONE CANON PLAZA LAKE SUCCESS, NY 11042-1198 TELEPHONE: (516) 328-5000 GENERAL FAX TELEPHONE (516) 328-5069 # 510(k) Summary 1. Date Prepared: July 23, 1996 2. Applicant's Name: Canon USA, Inc. One Canon Plaza Lake Success, NY 11042 U.S.A. Mr. Glenn Impal / Mr. Ken Shadoff Phone: (516) 328-5600 Fax: (516) 328-5169 3. Model Name: Canon TX-10 Tonometer 86HKX, Tonometer, AC Powered Document Control Number: ? 4. Predicate Model: Canon T-2 Tonometer 86HKX, Tonometer, AC Powered Document Control Number: K943939 5. Device Description: The TX-10 is a tonometer designed using a non-contact measurement system. Air puff gently measures the intraocular pressure with the help of an auto-alignment system. Features quick and easy measurement that does not require any special skills. 6. Intended Use: The TX-10 tonometer is intended to be used for the measurement of intraocular pressure of the human eye. 7. Summary of Technological Characteristics: See the attached comparison chart. 8. Assessment of Performance Data: See Attachment 1 {2} Comparison in the measurement accuracy among Canon TX-10 and Canon T-2 | | TX-10 | T-2 | ISO Specifications (proposed) | | --- | --- | --- | --- | | Regression Slope | 0.996 | 0.907 | 0.95 to 1.05 | | Regression Y intercept | 0.462mmHg | 1.08mmHg | -1.0 to +1.0 | | Standard Deviation of Difference, Sd | 1.56mmHg | 2.12mmHg | 2.5mmHg | | Coefficient of Correlation | 0.926 | 0.858 | Unspecified | {3} K963079 # CANON TONOMETERS COMPARISON CHART | MODEL | | CANON NON CONTACT TONOMETER T-2 | CANON TONOMETER TX-10 | | --- | --- | --- | --- | | Specification | Measuring Range Setting | 0~30/0~60mmHg (Automatic) | Same as T-2 | | | Alignment Method | One bright dot with alignment level display | **Auto Mode** Monitor and auto fine alignment | | | | | **Manual Mode** Monitor with two dots | | | Increment | 1 mmHg | Same as T-2 | | | Measuring Time | 3 millisecond (0.003 second) | Same as T-2 | | | Fixation Target | LED (green) | Same as T-2 | | | Chin Rest | Manual | Power assisted (driving at controller) | | Safety Mechanism | | Mechanical Stopper | Software Controlled Stopper | | Printout Format | | R/L, date, time, avarage, ID, 3 or 10 data for each eye | Same as T-2 | | Display | | 5" (inch) B/W CRT | 4" (inch) Color LCD | | Dimentions (width x depth x height) mm | 286 x 445 x 472 | **Body** 299 x 340 x 177 | | | | | **Controller** 170 x 183 x 177 | | | Weight (kg) | 22.0 | **Body** 16.0 | | | | | **Controller** 1.5 | |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...