HEELBO SECURITY BELT

K963060 · Heelbo, Inc. · FMQ · Nov 4, 1996 · General Hospital

Device Facts

Record IDK963060
Device NameHEELBO SECURITY BELT
ApplicantHeelbo, Inc.
Product CodeFMQ · General Hospital
Decision DateNov 4, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6760
Device ClassClass 1

Intended Use

The Heelbo Security Belt is intended to help secure the patient while seated in a wheelchair or geriatric chair.

Device Story

Padded belt (4.75" x 22") with cotton webbed straps; straps tie to wheelchair or geriatric chair frame; used to secure patient in seated position; prevents patient from sliding or falling; operated by caregivers in clinical or home settings.

Clinical Evidence

Bench testing only; no clinical data presented. Biocompatibility assessment based on history of safe use of materials in clothing/garment industry and lack of reported skin irritation.

Technological Characteristics

Padded belt, 4.75 inches wide by 22 inches long; cotton webbed straps; non-powered; mechanical restraint.

Indications for Use

Indicated for patients requiring assistance to remain secured while seated in a wheelchair or geriatric chair.

Regulatory Classification

Identification

A protective restraint is a device, including but not limited to a wristlet, anklet, vest, mitt, straight jacket, body/limb holder, or other type of strap, that is intended for medical purposes and that limits the patient's movements to the extent necessary for treatment, examination, or protection of the patient or others.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 11/01/96 12:46 847 918 3860 010/011 Heelbo, Inc. 1134 N. Homan Ave. Chicago, IL 60651 312/489-7711 Telex RCA 297-248 800/323-5444 Fax 312/489-7744 K963060 Heelbo, Inc. Security Belt 510(k) Summary 1. Submitter's name, Address and Contact Person Submitter: Heelbo, Inc. 1134 N. Homan Chicago, IL 60651 Contact Person: Joseph S. Tokarz, Manager, Regulatory Affairs Hollister Incorporated 2000 Hollister Drive Libertyville, IL 60048 Ph (847)680-2849, Fax (847)918-3860 Date Summary Prepared - November 1, 1996 2. Name of Device: Heelbo Security Belt 3. Name of Predicate Device(s) Heelbo Security Belt 4. Description of Device The Heelbo Security Belt is a padded belt 4¾ inches wide by 22 inches long with cotton webbed straps sewn onto both ends. The webbed straps tie to the frame of the wheelchair or geriatric chair to help secure the patient while seated. 5. Statement of Intended Use The Heelbo Security Belt is intended to help secure the patient while seated in a wheelchair or geriatric chair. 6. Statement of Technological Characteristics of the Device The Heelbo Security Belt is a padded belt 4¾ inches wide by 22 inches long with cotton webbed straps sewn onto both ends. The webbed straps tie to the frame of the wheelchair or geriatric chair to help secure the patient while seated. The subject devices are identical in intended use, design, materials, manufacturing process, physical and mechanical specifications and issues of safety and effectiveness to the devices prior to the submission of this notification. The only difference is that the product labeling has been revised to comply with the Agency's labeling requirements set forth in the draft Guidance on the Content of Premarket Notification [510(k)] Submissions for Protective Restraints dated December 1995. Page 18 {1} 11/01/96 12:47 847 918 3860 011/011 Heelbo, Inc. 1134 N. Homan Ave. Chicago, IL 60651 312/489-7711 Telex RCA 297-248 800/323-5444 Fax 312/489-7744 Heelbo, Inc. Security Belt 7. Biocompatibility Assessment The subject devices are identical in components materials to the predicate devices. The suppliers of the materials used in the fabrication of these devices have stated that there is a history of safe use of their materials in the clothing and garment industry. In addition, Heelbo, Inc., is not aware of any reports or complaints of skin irritation associated with these materials. 8. Conclusion Based upon the information presented above it is concluded that the proposed Heelbo Security Belt is safe and effective for its intended use and is substantially equivalent to the predicate device. Page 19
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...