ENDOBEAM

K963044 · Corin USA · GCT · Sep 20, 1996 · Gastroenterology, Urology

Device Facts

Record IDK963044
Device NameENDOBEAM
ApplicantCorin USA
Product CodeGCT · Gastroenterology, Urology
Decision DateSep 20, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

Endoscopic light source to provide illumination for fibre optic endoscopy to joint cavities and abdominal space.

Device Story

Endobeam is an endoscopic light source providing illumination for fibre optic endoscopy in joint cavities and abdominal space. Device utilizes a 35W Xenon microdischarge lamp with 4000-hour life and 3700K color temperature. Operated via manual or automatic illumination control with stepless intensity adjustment. Used in clinical settings by physicians; provides light via fibre optic cables to surgical sites. Safety tested to IEC 601/1 1988 and DIN VDE 0750 standards; includes electromagnetic compatibility testing.

Clinical Evidence

Bench testing only. Device tested for safety per IEC 601/1 1988 and DIN VDE 0750 TI/11.91, and for electromagnetic compatibility.

Technological Characteristics

Xenon microdischarge lamp (35W, 3700K, 4000-hour life). Features manual/automatic illumination control, stepless intensity adjustment, and UV/IR filters. Power input 100-240V, 0-440 Hz. Safety standards: IEC 601/1 1988, DIN VDE 0750 TI/11.91. Electromagnetic compatibility compliant.

Indications for Use

Indicated for patients undergoing fibre optic endoscopy of joint cavities and abdominal space.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} Page 13 K963044 SEP 20 1996 (g) Detailed Similarities/Differences Comparison: | Endospot 510(k) Device | Autobrite (Predicate Device) | | --- | --- | | Power Source | Power Source | | On/Off Button | On/Off Button | | Manual Illumination Control | Manual Illumination Control | | Automatic Illumination Control | Automatic Illumination Control | | UV/IR Filters | UV/IR Filters | | 35 Watt | 100 Watt | | Xenon Microdischarge Lamp | Xenon Lamp Module | | 1 Lamp in Unit | 1 Lamp in Unit | | Stepless Light Intensity Adjustment | Stepless Light Intensity Adjustment | | 4000 Lamp Hours | 1000 Lamp Hours | ## 8. 510(k) Summary: Classification Name: Accessory to Endoscope Device Trade Name: Endobeam Sponsor: Corin U.S.A. 10500 University Center Drive Suite 130 Tampa, FL 33612 USA Reason for Submission: New device Intended Use: Endoscopic light source to provide illumination for fibre optic endoscopy to joint cavities and abdominal space. Major Components: - Power Source - Xenon Microdischarge Lamp Lamp/Data: - Light intensity 35W - Lamp life - 4000 hours - Color temperature - 3700K Power Source: 100 - 240V ± 10% 0 - 440 Hz Safety Testing: - TUV tested to IEC 601/1 1988 - DIN VDE 0750 TI/11.91 - Electromagnetic compatibility tested {1} Page 14 Comparison to Legally Marketed Endoscopic Light Source: Dyobrite 3000 Xenon Illuminator Smith and Nephew Dyonics 160 Dascomb Road Andover, Massachusetts 01810 USA 9. Truthful and Accurate Statement: I hereby confirm that the attached information is, to the best of my knowledge, truthful and accurate. PRINT NAME: Jack Thomas SIGNED: ![img-0.jpeg](img-0.jpeg) POSITION: President DATE: 1 August 1996
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...