DSL DHEA-S-7

K962895 · Diagnostic Systems Laboratories, Inc. · JKC · Sep 10, 1996 · Clinical Chemistry

Device Facts

Record IDK962895
Device NameDSL DHEA-S-7
ApplicantDiagnostic Systems Laboratories, Inc.
Product CodeJKC · Clinical Chemistry
Decision DateSep 10, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1245
Device ClassClass 1

Indications for Use

The DSL C2700 DHEA-S-7 RIA assay is intended for the quantitative determination of DHEA-S in human serum. The measurement of DHEA-S is used for in vitro diagnostic use in the diagnosis and treatment of DHEA-secreting adrenal carcinomas.

Device Story

Competitive binding radioimmunoassay (RIA) kit; measures DHEA-S in human serum. Input: serum sample; radio-labeled DHEA-S tracer; DHEA-S antiserum. Process: competitive binding between labeled and unlabeled DHEA-S for antibody sites; centrifugation and decantation for separation; gamma counter analysis of bound counts. Output: radioactivity counts inversely proportional to DHEA-S concentration. Used in clinical laboratories by technicians. Results assist physicians in diagnosing and monitoring adrenal carcinomas.

Clinical Evidence

Bench testing only. Comparative study of 418 patient serum samples using subject device and predicate. Linear regression analysis: Y = 0.76(X) + 36.2; correlation coefficient r = 0.98.

Technological Characteristics

Competitive binding radioimmunoassay. Analyte: DHEA-S. Separation method: centrifugation and decantation. Detection: gamma counter. Format: in vitro diagnostic kit.

Indications for Use

Indicated for quantitative measurement of DHEA-S in human serum to aid in the diagnosis and treatment of DHEA-secreting adrenal carcinomas.

Regulatory Classification

Identification

A dehydroepiandrosterone (free and sulfate) test system is a device intended to measure dehydroepiandrosterone (DHEA) and its sulfate in urine, serum, plasma, and amniotic fluid. Dehydroepiandrosterone measurements are used in the diagnosis and treatment of DHEA-secreting adrenal carcinomas.

Predicate Devices

Submission Summary (Full Text)

{1} SEP 10 1996 K962895 Diagnostic Systems Laboratories, Inc. 445 Medical Center Boulevard Webster Texas 77598-4217 USA Tel 713.332.9678 Fax 713.554.4220 Customer Assistance Center Tel 800.231.7970 Fax 713.338.1895 # SUMMARY OF SAFETY AND EFFECTIVENESS Name of Device: DSL C2700 DHEA-S-7 RIA Kit Classification Name: Radioimmunoassay, DHEA-S Analyte Code and Name: DHEA-S-7 Regulatory Class: I Submitter: John Willis Diagnostic Systems Laboratories, Inc. 445 Medical Center Boulevard Webster, Texas 77598 Phone: 713-332-9678 Date: July 24, 1996 The DSL C2700 DHEA-S-7 RIA kit was developed for the quantitative measurement of DHEA-S in human serum. The RIA format is a competitive binding protein assay. Radio-labeled DHEA-S competes with un-labeled DHEA-S in the serum sample for binding sites to the DHEA-S antiserum provided with the kit. Separation of free from bound DHEA-S is achieved by the centrifugation and decantation of the tubes. The resultant is analyzed in a gamma counter for bound counts per minute. The amount of radio-labeled DHEA-S bound to the antibody is inversely proportional to the concentration of the DHEA-S present in the sample. The DSL C2700 DHEA-S-7 RIA assay is intended for the quantitative determination of DHEA-S in human serum. The measurement of DHEA-S is used for in vitro diagnostic use in the diagnosis and treatment of DHEA-secreting adrenal carcinomas. The DSL C2700 DHEA-S-7 RIA is substantially equivalent to the DSL 2700 DHEA-S RIA. Both kits have the same intended use. To demonstrate substantial equivalence between the two assays, patient samples (n=418) were collected and assayed using both methods. Samples were chosen based on expected DHEA-S levels so that samples with low, intermediate and high levels would be evaluated. Linear regression analysis of the results obtained for the comparison gave the equation $Y = 0.76(X) + 36.2$ with a correlation coefficient of $(r) = 0.98$.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...