K962875 · Diagnostic Systems Laboratories, Inc. · CGI · Aug 23, 1996 · Clinical Chemistry

Device Facts

Record IDK962875
Device NameACTIVE ULTRASENSITIVE UNCONJUGATED ESTRIOL EIA
ApplicantDiagnostic Systems Laboratories, Inc.
Product CodeCGI · Clinical Chemistry
Decision DateAug 23, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1265
Device ClassClass 1

Indications for Use

The DSL ULTRA-SENSITIVE UNCONJUGATED ESTRIOL EIA assay is intended for the quantitative determination of Unconjugated Estriol in human serum. The measurement of Unconjugated Estriol is used as a diagnostic aid in the diagnosis and treatment of fetoplacental distress.

Device Story

Competitive binding enzyme immunoassay (EIA) for quantitative measurement of unconjugated estriol in human serum; utilizes horseradish peroxidase-labeled unconjugated estriol competing with sample estriol for antibody binding sites; substrate tetramethylbenzidine (TMB) added post-incubation/wash; enzymatic turnover measured via dual-wavelength absorbance; used in clinical laboratory settings by trained personnel; results aid clinicians in assessing fetoplacental status to guide management of pregnancy complications.

Clinical Evidence

Bench testing only. Comparative study of 106 patient serum samples using both the subject EIA kit and the predicate RIA method. Linear regression analysis showed r = 0.93, Y = 0.85(X) - 0.11.

Technological Characteristics

Competitive binding enzyme immunoassay (EIA); horseradish peroxidase label; tetramethylbenzidine (TMB) substrate; dual-wavelength absorbance detection; manual or automated plate reader format.

Indications for Use

Indicated for quantitative measurement of unconjugated estriol in human serum as a diagnostic aid in the diagnosis and treatment of fetoplacental distress.

Regulatory Classification

Identification

An estriol test system is a device intended to measure estriol, an estrogenic steroid, in plasma, serum, and urine of pregnant females. Estriol measurements are used in the diagnosis and treatment of fetoplacental distress in certain cases of high-risk pregnancy.

Predicate Devices

Submission Summary (Full Text)

{1} AUG 23 1996 K962875 Diagnostic Systems Laboratories, Inc. 445 Medical Center Boulevard Webster, Texas 77598-4217 USA Tel 713.332.9678 Fax 713.554.4220 Customer Assistance Center Tel 800.231.7970 Fax 713.338.1895 # SUMMARY OF SAFETY AND EFFECTIVENESS Name of Device: DSL 10-3700 ULTRA-SENSITIVE UNCONJUGATED ESTRIOL EIA Kit Classification Name: Enzyme Immunoassay, UNCONJUGATED ESTRIOL Analyte Code and Name: Unconjugated Estriol Regulatory Class: I Submitter: John Willis Diagnostic Systems Laboratories, Inc. 445 Medical Center Boulevard Webster, Texas 77598 Phone: 713-332-9678 Date: July 19, 1996 The DSL Ultra-Sensitive Unconjugated Estriol EIA kit was developed for the quantitative measurement of Unconjugated Estriol in human serum. The EIA format is a competitive binding protein assay. Horseradish peroxidase labelled unconjugated estriol competes with un-labeled Unconjugated Estriol in the serum sample for antibody binding sites. After incubation and washing the wells are incubated with the substrate tetramethylbenzidine (TMB). An acidic stopping solution is then added and the degree of enzymatic turnover of the substrate is determined by dual wavelength absorbance measurement. The DSL ULTRA-SENSITIVE UNCONJUGATED ESTRIOL EIA assay is intended for the quantitative determination of Unconjugated Estriol in human serum. The measurement of Unconjugated Estriol is used as a diagnostic aid in the diagnosis and treatment of fetoplacental distress. The DSL ULTRA-SENSITIVE UNCONJUGATED ESTRIOL EIA is substantially equivalent to the DSL ULTRA-SENSITIVE UNCONJUGATED ESTRIOL RIA. To demonstrate substantial equivalence between the two assays, patient samples (n = 106) were collected and assayed using both methods. Samples were chosen based on expected Unconjugated Estriol levels so that samples with low, intermediate and high levels would be evaluated. Linear regression analysis of the results obtained for the comparison gave the equation $Y = 0.85(X) - 0.11$ with a correlation coefficient of (r) = 0.93.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...