ODS1 - HOOD DRIVER

K962871 · Reimers Systems, Inc. · CBF · Oct 23, 1996 · Anesthesiology

Device Facts

Record IDK962871
Device NameODS1 - HOOD DRIVER
ApplicantReimers Systems, Inc.
Product CodeCBF · Anesthesiology
Decision DateOct 23, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5470
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Model ODS1 - Hood Driver is intended for use with monoplace hyperbaric chambers. It is normally used to permit the patient to breath oxygen while the chamber is pressurized on air. However, it can be used to supply the patient with any physiologically appropriate gas which may or may not be the same as the gas used to pressurize the chamber.

Device Story

ODS1 Hood Driver; hyperbaric chamber breathing station. Supplies steady flow of breathing gas (oxygen or air) to patient via 'Sea-Long' hood or free-flow mask. Controls exhaust flow to chamber exterior. Used in monoplace hyperbaric chambers; operated by clinician (inside or outside chamber). Protects patient from suction injury via flow regulation. Benefits patient by enabling controlled gas delivery during hyperbaric therapy.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Materials: Copper, brass, stainless steel; 100% oxygen compatible softgoods. Connectivity: Standard 22 mm respiratory fittings. Flow control: 20 - 200 scfh. Mechanical gas flow regulation system.

Indications for Use

Indicated for patients undergoing hyperbaric oxygen therapy for conditions listed by the Undersea Hyperbaric & Medical Association.

Regulatory Classification

Identification

A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).

Predicate Devices

Submission Summary (Full Text)

{0} OCT 23 1996 K962871 # REIMERS SYSTEMS, INC. 7668 FULLERTON ROAD SPRINGFIELD, VIRGINIA 22153 S. D. REIMERS PRESIDENT # 510(k) SUMMARY - ODS1 HOOD DRIVER TELEPHONE (703) 866-5490 TELECOPIER (703) 866-9066 DATE: July 18, 1996 SUBMITTER: Stephen D. Reimers, PE President CONTACT: Carlyle H. Smith, PE Staff Engineer # DEVICE INFORMATION: Trade name: ODS1 - Hood Driver Common name: Hyperbaric Chamber Breathing Station Classification name: Chambers, Hyperbaric 73CBF # DESCRIPTION: The Model ODS1 - Hood Driver is intended for use with monoplace hyperbaric chambers. It is normally used to permit the patient to breath oxygen while the chamber is pressurized on air. However, it can be used to supply the patient with any physiologically appropriate gas which may or may not be the same as the gas used to pressurize the chamber. The ODS1 circuit is designed to supply a steady flow of breathing gas to the breathing apparatus used in the chamber. The breathing apparatus can be either a "Sea-Long" hood or a free-flow mask arrangement. # INDICATIONS: Hyperbaric chambers are used for treating the indications as currently listed by the Undersea Hyperbaric & Medical Association for hyperbaric oxygen therapy. # COMPARISON TO EXISTING EQUIPMENT: ## Similarities: The ODS1 is substantially equivalent to the breathing stations found in larger hyperbaric chamber systems, such as the Lifeforce Hyperbaric Chamber System (K951125) built by Reimers Engineering. They are similar in the following ways: 1. Supplies a steady flow of breathing gas to the breathing apparatus. 2. Controls a steady exhaust flow from the breathing apparatus to the chamber exterior. Page 1 of 2 {1} 3. Can be used with both hoods and constant flow masks. 4. Designed around standard 22 mm respiratory fittings. 5. Patient is adequately protected from suction injury caused by excessive exhaust flow or inadequate supply flow. 6. Requires minimal flow adjustments after initial setup. 7. Same range of supply flow control: ODS1: 20 - 200 scfh Lifeforce: 21 - 212 scfh 8. Works with oxygen or air. 9. Same wetted materials (Copper, Brass, Stainless-Steel) 10. 100% oxygen compatible softgoods ## Differences The significant differences between the ODS1 - Hood Driver and the Lifeforce breathing stations are as follows: 1. The Lifeforce breathing stations are located inside the chamber. The ODS1 controls can be placed inside or outside of the chamber. With the controls outside of the chamber, an outside attendant sets the controls instead of an inside attendant. 2. The Lifeforce breathing stations contain an additional supply flowmeter and throttle valve for a separate nebulizer flow. The ODS1 has only one supply flowmeter and throttle valve. The ODS1 can still be used with certain types of nebulizers if desired. However the additional dedicated nebulizer circuit has been found unnecessary for most monoplace chamber applications. ## DISCUSSION: Neither of these differences affect the safety or effectiveness of the ODS1 which is inherent in the circuit design. The differences are only related to the packaging of the circuit. ## CONCLUSION: The ODS1 - Hood Driver is substantially equivalent to existing, legally marketed devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...