PLEUR-EVAC SAHARA ADULT/PEDIATRIC CHEST DRAINAGE SYSTEM MODEL S-1100, S-1200, 2-2100, AND S-2200 WITH MODEL S-100 AUTOTR
K962856 · Deknatel, Inc. · KDQ · Aug 28, 1996 · General Hospital
Device Facts
Record ID
K962856
Device Name
PLEUR-EVAC SAHARA ADULT/PEDIATRIC CHEST DRAINAGE SYSTEM MODEL S-1100, S-1200, 2-2100, AND S-2200 WITH MODEL S-100 AUTOTR
Applicant
Deknatel, Inc.
Product Code
KDQ · General Hospital
Decision Date
Aug 28, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6740
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Models S-1100, S-1200, S-2100, and S-2200 covered by this submission are sterile, single use devices, that are intended for postoperative chest drainage. The Pleur-evac Sahara Model S-100 Autotransfusion Bag is a sterile, non-pyrogenic, single-use device, intended for post-surgical collection and reinfusion of autologous blood from the thoracic cavity when attached to a Pleur-evac Sahara Chest Drainage.
Device Story
Chest drainage system for postoperative thoracic fluid management; collects fluids via suction or gravity. System incorporates integral suction control, one-way check valve, and fluid collection chamber. Autotransfusion bag (Model S-100) collects autologous blood for reinfusion; features rigid top plate, flexible vinyl bag, and wire support frame. Clinician-operated in hospital settings; attaches to bed rails or stands via integrated fluorstand or base. Connects to patient via drainage tube; includes injection site for anticoagulants or sampling. System facilitates blood reinfusion by removing wire frame to collapse bag. Benefits include safe, sterile collection and reinfusion of patient's own blood, reducing need for donor blood.
Clinical Evidence
Bench testing only. Evaluated suction control accuracy, airflow capacity, response to patient air leak, system cracking pressure, high negative pressure relief valve performance, negative pressure indicator performance, and autotransfusion bag performance under simulated patient pressure. Biocompatibility testing for latex-free patient drainage tubes and injection sites conducted per ISO 10993-1.
Technological Characteristics
Three-bottle system; dry chest drainage unit. Materials: flexible vinyl bag, rigid top plate, latex-free tubing/injection sites. Features: integral suction control, one-way check valve, rotating fluorstand (S-2100/S-2200), Easy-Link adaptor. Sterile, single-use. Biocompatibility per ISO 10993.
Indications for Use
Indicated for postoperative chest drainage and post-surgical collection and reinfusion of autologous blood from the thoracic cavity in patients requiring such procedures.
Regulatory Classification
Identification
A vacuum-powered body fluid suction apparatus is a device used to aspirate, remove, or sample body fluids. The device is powered by an external source of vacuum. This generic type of device includes vacuum regulators, vacuum collection bottles, suction catheters and tips, connecting flexible aspirating tubes, rigid suction tips, specimen traps, noninvasive tubing, and suction regulators (with gauge).
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 880.9.
Predicate Devices
Thora-Klex Model 0077000 Chest Drainage System (K801043A, K830671)
Pleur-evac Model A-6000 Chest Drainage System (K881252, K905768A)
Pleur-evac Model A-1500 Autotransfusion Bag (K854301, K881252, K884844A)
Submission Summary (Full Text)
{0}
K962856
18. 510(k) Summary
AUG 28 1996
Submitter:
Deknatel DSP Worldwide Inc.
600 Airport Road
Fall River, MA 02720
Tel: (508) 677-6600
Fax: (508) 677-6667
Contact:
Dean E. Ciporkin, M.S.
Director of Regulatory Affairs
Date Prepared:
June 25, 1996
Trade Name:
Pleur-evac Sahara Adult/Pediatric
Chest Drainage System
Models S-1100, S-1200, S-2100, and S-2200
with Model S-100 Autotransfusion Bag
Common Name:
Chest Drainage Systems
Classification Name:
System, Drainage, Thoracic, Vacuum
Powered Body Fluid Suction Apparatus
Equivalent Device:
The Pleur-evac Sahara Adult/Pediatric Chest Drainage System Models S-1100, S-1200, S-2100, and S-2200 with Model S-100 Autotransfusion Bag are substantially equivalent in form, fit, function and intended use to the Thora-Klex Model 0077000 Chest Drainage System, cleared for marketing by FDA under 510(k)s #K801043A and #K830671, and the Pleur-evac Model A-6000 Chest Drainage System, cleared for marketing by FDA under 510(k)s #K881252 and #K905768A.
The Model S-100 Autotransfusion Bag, in specific, is substantially equivalent in form, fit, and function, to the currently marketed Pleur-evac Model A-1500 Autotransfusion Bag used with the Pleur-evac Model A-6000 Chest Drainage System, cleared for marketing by FDA under 510(k)s #K 854301, #K881252 and #K884844A.
Device Description:
Pleur-evac Sahara Adult/Pediatric Chest Drainage System Models S-1100, S-1200, S-2100, and S-2200
Deknatel DSP Worldwide, Inc. has recently acquired the Thora-Klex® Chest Drainage System product lines from Davol Inc., C.R. Bard, Inc. Deknatel DSP Worldwide, Inc. has incorporated certain features of the currently marketed Model 0077000 Thora-Klex
000020
{1}
Chest Drainage System with the features of the currently marketed Model A-6000 Pleur-evac® Chest Drainage System into a new chest drainage system, the Pleur-evac Sahara Adult/Pediatric Chest Drainage System. Combining the proven technologies from the Thora-Klex and the Pleur-evac into one unit, creates a new, completely dry Pleur-evac Chest Drainage Unit. Features from the Thora-Klex and the Pleur-evac were transferred to the Pleur-evac Sahara unit without design changes.
There are two basic Pleur-evac Sahara Chest Drainage models: the S-1100 Chest Drainage Unit which is shorter, wider and deeper than the standard Pleur-evac unit but more like the Thora-Klex Model 0077000 shape; and the S-2100 Chest Drainage Unit which carries the standard dimensions of the Pleur-evac Model A-6000 Chest Drainage unit. The S-1100 and S-2100 Chest Drainage Units are capable of autotransfusion option, by attaching a Pleur-evac Sahara Model S-100 Autotransfusion Bag. The Models S-1200 and the S-2200 are identical to the Models S-1100 and S-2100, but are provided with a Model S-100 Autotransfusion Bag attached. The Pleur-evac Sahara Chest Drainage are three bottle systems that include a means for integral suction control, a one way seal via a check valve, and a chamber for collecting fluids.
Hangers are located on the right and left sides of the chest drainage unit to provide support for the chest drainage unit when it is suspended from a hospital bed rail.
The Pleur-evac Sahara Models S-2100 and S-2200 are provided with an integral rotating fluorstand. These units free stand when placed on a leveled surface with the fluorstand in the open position. The fluorstand rotates freely from the closed position to the open position. Once open, the fluorstand requires the actuation of a lever to return to the closed position. The design of the Pleur-evac Sahara Models S-1100 and S-1200 does not require a rotating fluorstand, since the base of the unit acts as a fluorstand.
Pleur-evac Sahara Model S-100 Autotransfusion Bag
The Pleur-evac Sahara Autotransfusion Bag, includes an Autotransfusion Bag that contains a rigid top plate assembled onto a flexible vinyl bag. The Autotransfusion Bag is mounted over a wire support frame. The wire frame maintains the bag in an open position during the collection phase of operation. The frame is removed from the bag during the reinfusion phase to allow the bag to collapse and the reinfusion of the collected blood to occur.
The Pleur-evac Sahara Autotransfusion Bag operates with either suction drainage or gravity drainage. During operation with suction drainage, the Autotransfusion Bag is connected to an appropriate Pleur-evac Sahara Chest Drainage unit that is connected to a vacuum source. The patient vacuum is controlled by the Pleur-evac Sahara Chest Drainage System. During operation with gravity drainage, the Autotransfusion Bag is connected to an appropriate Pleur-evac Sahara Chest Drainage unit that is not attached to a suction source.
000021
{2}
The Pleur-evac Sahara Autotransfusion Bag is connected to the Pleur-evac Sahara S-1100 Chest Drainage Unit by means of the Easy-Link Adaptor. The adaptor, which is attached to the Autotransfusion Bag, locks into the Easy-Link receptor on the Pleur-evac Sahara Chest Drainage unit. For the Pleur-evac Sahara Model S-2100 Chest Drainage Unit, the Easy-Link Adaptor is removed from the Model S-100 Autotransfusion Bag by the clinician, and the Autotransfusion Bag is attached directly to the side of the Pleur-evac Chest Drainage Unit, using the metal frame on the bag and the hooks provided on the chest drainage unit. This means of attachment is employed by the currently marketed Pleur-evac Model A-6000 Chest Drainage System. Tubing connectors are provided for attaching the Pleur-evac Sahara Autotransfusion Bag to the appropriate Pleur-evac Sahara Chest Drainage Unit. The connectors are color coded for ease of proper connection.
Tubing clamps are located on each of the tubing ports on the Autotransfusion Bag. The tubing clamps must be closed in order to occlude the patient drainage tube prior to disconnecting the connectors. An injection site is located on one set of connectors through which anti-coagulants may be added to the Autotransfusion Bag or from which samples of the drainage fluid may be taken.
A hanger strap located on the top of the unit provides a means for suspending the Autotransfusion Bag from an I.V. pole during reinfusion.
The Models S-100 and the A-1500 Autotransfusion Bags have similar face graphics, box graphics, and instructions for use. The labeling for the Model S-100 Autotransfusion Bag also includes reference and use of the Easy-Link Adaptor.
Intended Use:
The Models S-1100, S-1200, S-2100, and S-2200 covered by this submission are sterile, single use devices, that are intended for postoperative chest drainage.
The Pleur-evac Sahara Model S-100 Autotransfusion Bag is a sterile, non-pyrogenic, single-use device, intended for post-surgical collection and reinfusion of autologous blood from the thoracic cavity when attached to a Pleur-evac Sahara Chest Drainage.
Summary of Technological Characteristic Equivalence:
All features and technology employed in the Pleur-evac Sahara Chest Drainage Systems are derived from either of the predicate devices, the Pleur-evac Model A-6000, or the Thora-Klex Model 0077000 Chest Drainage Systems. Each of these features from the Thora-Klex and the Pleur-evac Chest Drainage Systems were transferred to the Pleur-evac Sahara Chest drainage System unit without design changes.
000022
{3}
Summary of Performance Equivalence:
Testing was performed to compare the functional aspects of the proposed Pleur-evac Sahara Chest Drainage System Models S-1100/S-1200, and S-2100/S-2200 to the Pleur-evac Chest Drainage System Model A-6000 and the Thora-Klex Chest Drainage System Model 0077000. The testing also evaluated the effect of the Pleur-evac Sahara Model S-100 Autotransfusion bag connected to the Pleur-evac Sahara Chest Drainage System Models S-1100/S-1200 and the S-2100/S-2200 units when subjected to simulated patient pressure.
Suction control accuracy, airflow capacity, response to patient air leak, system cracking pressure, high negative pressure relief valve performance, negative pressure indicator performance, and Autotransfusion Bag performance were tested and compared.
The Pleur-evac Sahara Chest Drainage Systems met the performance specifications and were found to be comparable to the Pleur-evac Model A-6000 Chest Drainage System and the Thora-Klex Model 0077000 Chest Drainage System.
The components of a Pleur-evac Sahara Chest Drainage System which come into contact with blood to be reinfused, when the system is attached to a Pleur-evac Sahara Model S-100 Autotransfusion Bag, and connected to a patient, are: the inside of the Autotransfusion Bag, the internal components of the Autotransfusion Bag, the Patient Drainage Tube, the connectors with the Injection Site, and the universal connector
The internal components of the Pleur-evac Sahara Model S-100 Autotransfusion Bag are made of the identical materials as the Pleur-evac Model A-1500 Autotransfusion Bag. Therefore, the biocompatibility testing results provided in 510(k) #K854301 apply to the bag and internal components of the Pleur-evac Sahara Model S-100 Autotransfusion Bag.
The Patient Drainage Tube and the Injection Site materials were changed to be free of latex. The biocompatibility testing for the Patient Drainage Tubes and the Injection Site was performed according to the "Biological Evaluation of Medical Devices", ISO 10993 Part-1. Test results indicated that the requirements of ISO 10993 were met, and that the materials are suitable for use in the Patient Drainage Tube and the Injection Site for the Pleur-evac Sahara Model S-100 Autotransfusion Bag.
000023
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.