HEELBO QUILTED LIMB HOLDER WITH DOUBLE D-RING, QUICK-RELEASE SNAP, AND STRAP
K962834 · Heelbo, Inc. · FMQ · Oct 18, 1996 · General Hospital
Device Facts
Record ID
K962834
Device Name
HEELBO QUILTED LIMB HOLDER WITH DOUBLE D-RING, QUICK-RELEASE SNAP, AND STRAP
Applicant
Heelbo, Inc.
Product Code
FMQ · General Hospital
Decision Date
Oct 18, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6760
Device Class
Class 1
Indications for Use
Heelbo Quilted Limb Holder with Double D-Ring, Quick-Release Snap, and Strap is intended to be a protective limb restraint that limits the patient’s movement to the extent necessary for treatment, examination, or protection of the patient or others. The device can be used in the supine or sitting positions.
Device Story
Disposable quilted limb holder; secures to bed, gurney, chair, or wheelchair via straps, Velcro fastener, Double D-Ring, and quick-release snap. Used by caregivers to limit patient movement for treatment, examination, or protection. Allows wrist adjustment without cinching; supports supine and sitting positions. Machine washable. Provides physical restraint to prevent patient injury or interference with medical care.
Clinical Evidence
No clinical data. Biocompatibility assessment based on history of safe use of materials in clothing/garment industry and absence of reported skin irritation.
Technological Characteristics
Quilted limb holder; Velcro fastener; Double D-Ring; quick-release snap; straps. Available in two sizes (Small-Regular and Large-X-Large). Machine washable. Non-electronic, mechanical device.
Indications for Use
Indicated for patients requiring protective limb restraint to limit movement for treatment, examination, or safety of the patient or others; applicable for use in supine or sitting positions.
Regulatory Classification
Identification
A protective restraint is a device, including but not limited to a wristlet, anklet, vest, mitt, straight jacket, body/limb holder, or other type of strap, that is intended for medical purposes and that limits the patient's movements to the extent necessary for treatment, examination, or protection of the patient or others.
Predicate Devices
Heelbo Quilted Limb Holder with Double D-Ring, Quick-Release Snap and Strap
Related Devices
K962832 — HEELBO QUILTED LIMB HOLDER WITH DOUBLE D-RING AND STRAP · Heelbo, Inc. · Oct 18, 1996
K962839 — HEELBO QUILTED LIMB HOLDER WITH BUCKLE AND STRAP · Heelbo, Inc. · Oct 18, 1996
K962831 — THE INCH QUILTED LIMB HOLDER WITH QUICK RELEASE SNAP CLOSURE · Heelbo, Inc. · Oct 18, 1996
K963476 — DISPOSABLE LIMB HOLDER · Deroyal Industries, Inc. · Jan 22, 1997
K963372 — HEELBO ER WRIST RESTRAINT · Heelbo, Inc. · Nov 7, 1996
Submission Summary (Full Text)
{0}
Ke96 2834
Heelbo, Inc.
1134 N. Homan Ave. Chicago, IL 60651
312/489-7711
Telex RCA 297-248
800/323-5444
Fax 312/489-7744
OCT 18 1996
Heelbo, Inc.
Heelbo Quilted Limb Holder with Double D-Ring, Quick-Release Snap and Strap
## Safety and Effectiveness Summary
1. Submitter's name, Address and Contact Person
| Submitter | Contact Person |
| --- | --- |
| Heelbo, Inc.
1134 N. Homan Ave.
Chicago, IL 60651 | Joseph S. Tokarz; Manager, Regulatory Affairs
Hollister Incorporated
2000 Hollister Drive
Libertyville, IL 60048
Ph: (847)680-2849, Fax: (847)918-3860 |
Date Summary Prepared - July 17, 1996
2. Name of Device:
Heelbo Quilted Limb Holder with Double D-Ring, Quick-Release Snap and Strap
3. Name of Predicate Device
Heelbo Quilted Limb Holder with Double D-Ring, Quick-Release Snap and Strap
4. Description of Device
Heelbo Quilted Limb Holder with Double D-Ring, Quick-Release Snap and Strap is a disposable quilted limb holder with a Velcro® fastener, Double D-Ring, Quick-Release Snap and Straps that secure to a bed, gurney, chair or wheelchair. The Heelbo Quilted Limb Holder with Double D-Ring, Quick-Release Snap and Strap is intended to be a protective limb restraint that limits the patient’s movement to the extent necessary for treatment, examination, or protection of the patient or others. The Heelbo Quilted Limb Holder allows for wrist adjustment to provide a secure fit without cinching. The Double D-Ring is simple to use and allows the caregiver easy access to the patient. Also, the adjustable Double D-Ring and Quick-Release Snap allow for supine and sitting positions. The Heelbo Quilted Limb Holders are machine washable.
The Heelbo Quilted Limb Holder with Double D-Ring, Quick-Release Snap and Strap is available in the following sizes:
| Stock No. 9350 | Stock No. 9360 |
| --- | --- |
| Small-Regular / cuff size: 11 ½" long
strap: 3/4" wide, 60" long | Large-X-Large / cuff size: 13 ½" long
strap: 3/4" wide, 60" long |
Page 18
{1}
Heelbo, Inc.
1134 N. Homan Ave. Chicago, IL 60651
312/489-7711
Telex RCA 297-248
800/323-5444
Fax 312/489-7744
Heelbo, Inc.
Heelbo Quilted Limb Holder with Double D-Ring, Quick-Release Snap and Strap
## 5. Statement of Intended Use
Heelbo Quilted Limb Holder with Double D-Ring, Quick-Release Snap, and Strap is intended to be a protective limb restraint that limits the patient’s movement to the extent necessary for treatment, examination, or protection of the patient or others. The device can be used in the supine or sitting positions.
## 6. Statement of Technological Characteristics of the Device
Heelbo Quilted Limb Holder with Double D-Ring, Quick-Release Snap and Strap is a disposable quilted limb holder with a Velcro® fastener, Double D-Ring, Quick-Release Snap and Straps that secure to a bed, gurney, chair or wheelchair. The Heelbo Quilted Limb Holder with Double D-Ring, Quick-Release Snap and Strap is intended to be a protective limb restraint that limits the patient’s movement to the extent necessary for treatment, examination, or protection of the patient or others. The Heelbo Quilted Limb Holder allows for wrist adjustment to provide a secure fit without cinching. The Double D-Ring is simple to use and allows the caregiver easy access to the patient. Also, the adjustable Double D-Ring and Quick-Release Snap allow for supine and sitting positions.
The subject devices are identical in intended use, design, materials, manufacturing process, physical and mechanical specifications and issues of safety and effectiveness to the devices prior to the submission of this notification. The only difference is that the product labeling has been revised to comply with the Agency’s labeling requirements set forth in the draft “Guidance on the Content of Premarket Notification [501(k)] Submissions for Protective Restraints” dated December 1995.
## 7. Biocompatibility Assessment
The subject devices are identical in component materials to the predicate devices. The suppliers of the materials use in the fabrication of these devices have stated that there is a history of safe use of their materials in the clothing and garment industry. In addition, Heelbo, Inc., is not aware of any reports or complaints of skin irritation associated with these materials.
## Conclusion
Based upon the information presented above it is concluded that the proposed Heelbo Quilted Limb Holder with Double D-Ring, Quick-Release Snap, and Strap is safe and effective for its intended use and is substantially equivalent to the predicate device.
Page 19
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches