CARBAMAZEPINE FLEX REAGENT CARTRIDGE

K962820 · Dade Intl., Inc. · KLT · Sep 13, 1996 · Clinical Toxicology

Device Facts

Record IDK962820
Device NameCARBAMAZEPINE FLEX REAGENT CARTRIDGE
ApplicantDade Intl., Inc.
Product CodeKLT · Clinical Toxicology
Decision DateSep 13, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3645
Device ClassClass 2

Indications for Use

The CRBM FLEX™ reagent cartridge used on the Dimension® clinical chemistry system is an in vitro diagnostic test intended to measure carbamazepine, an anti-convulsant drug, in plasma or serum. Measurements obtained by this device are used in the diagnosis and treatment of carbamazepine overdose and in monitoring levels of carbamazepine to ensure appropriate therapy.

Device Story

In vitro diagnostic reagent cartridge for Dimension® clinical chemistry system; measures carbamazepine in human serum/plasma. Uses Particle Enhanced Turbidimetric Inhibition Immunoassay (PETINIA) technique; latex particle-carbamazepine conjugate reacts with carbamazepine-specific monoclonal antibody. Operated by laboratory personnel on automated analyzer. Output provides quantitative carbamazepine concentration; assists clinicians in therapeutic drug monitoring and overdose management.

Clinical Evidence

Method comparison study using 82 samples tested on Dimension® system vs. aca® discrete clinical analyzer. Results: slope 0.99, intercept 0.06, correlation coefficient 0.979, range 0-18.6 µg/mL.

Technological Characteristics

Particle Enhanced Turbidimetric Inhibition Immunoassay (PETINIA); utilizes latex particle-carbamazepine conjugate and monoclonal antibody; prepackaged reagent cartridge format for automated clinical chemistry analyzer.

Indications for Use

Indicated for patients requiring measurement of carbamazepine levels in plasma or serum for diagnosis/treatment of overdose or therapeutic drug monitoring.

Regulatory Classification

Identification

A neuroleptic drugs radioceptor assay test system is a device intended to measure in serum or plasma the dopamine receptor blocking activity of neuroleptic drugs and their active metabolites. A neuroleptic drug has anti-psychotic action affecting principally psychomotor activity, is generally without hypnotic effects, and is a tranquilizer. Measurements obtained by this device are used to aid in determining whether a patient is taking the prescribed dosage level of such drugs.

Predicate Devices

Submission Summary (Full Text)

{1} DADE SEP 13 1996 K942820 DADE INTERNATIONAL Chemistry Systems P.O. Box 6101 Newark, DE 19714 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 ## Carbamazepine FLEX™ Reagent Cartridge ## Summary of Safety and Effectiveness The CRBM FLEX™ reagent cartridge used on the Dimension® clinical chemistry system is an *in vitro* diagnostic test intended to measure carbamazepine, an anti-convulsant drug, in plasma or serum. Measurements obtained by this device are used in the diagnosis and treatment of carbamazepine overdose and in monitoring levels of carbamazepine to ensure appropriate therapy. The CRBM method is based on Particle Enhanced Turbidimetric Inhibition Immunoassay (PETINIA) technique which uses a latex particle-carbamazepine conjugate and carbamazepine-specific monoclonal antibody. The CRBM FLEX™ reagent cartridge is substantially equivalent to the aca® carbamazepine analytical test pack, which was cleared by the FDA via the 510(k) process. Both tests use prepackaged reagents for the determination of carbamazepine in human serum and plasma. Eighty-two samples were tested with the CRBM FLEX™ reagent cartridge on the Dimension® system and the aca® CRBAM test pack on the aca® discrete clinical analyzer, with the following results: - slope = 0.99 - intercept = 0.06 - correlation coefficient = 0.979 - range of samples = 0-18.6 µg/mL Carolyn K. George Regulatory Affairs and Compliance Manager September 4, 1996 Date Printed on recycled paper with 25% post-consumer fiber
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...